Skip to content

Trauma Internal Fixation Systems

by Depuy Orthopaedics, Inc.

510(k) K111663

Trauma Internal Fixation Systems is a Class II device made by Depuy Orthopaedics, Inc., cleared under FDA 510(k) K111663 on 12 Mar 2012, active as of 22 Oct 2026. 5 FDA recalls are linked to it.

Active United States

510(k)

Valid until

No expiry

510(k) clearance is not time-limited

FDA recalls

None ongoing

17 closed

Documents

Trust

Unverified

Source FDA 510(k), synced 22 Oct 2026.

Regulatory status

as of registry sync 22 Oct 2026

RegionAuthorizationNumberDateStatus
United StatesFDA 510(k)SESE Substantially Equivalent · Traditional NumberK111663 Date12 Mar 2012ClearedStatusActive

Official records

Recalls

17 recalls · FDA

    • Terminated Z-0811-2022 FDA Class II initiated 14 Feb 2022 on shared 510(k) K111663

    Packaging of various implants may not have sufficient adhesion on all sides, leading to loss of sterile barrier integrity. Use of product poses risk of infection, leading to surgical intervention, and/or extension of surgery to find another part.

  • 11 recalls, same reason

    • Terminated Z-1234-2018 FDA Class II initiated 20 Dec 2017 on shared 510(k) K111663
    • Terminated Z-1231-2018 FDA Class II initiated 20 Dec 2017 on shared 510(k) K111663
    • Terminated Z-1227-2018 FDA Class II initiated 20 Dec 2017 on shared 510(k) K111663
    • Terminated Z-1229-2018 FDA Class II initiated 20 Dec 2017 on shared 510(k) K111663
    • Terminated Z-1233-2018 FDA Class II initiated 20 Dec 2017 on shared 510(k) K111663
    • Terminated Z-1225-2018 FDA Class II initiated 20 Dec 2017 on shared 510(k) K111663
    • Terminated Z-1228-2018 FDA Class II initiated 20 Dec 2017 on shared 510(k) K111663
    • Terminated Z-1226-2018 FDA Class II initiated 20 Dec 2017 on shared 510(k) K111663
    • Terminated Z-1232-2018 FDA Class II initiated 20 Dec 2017 on shared 510(k) K111663
    • Terminated Z-1235-2018 FDA Class II initiated 20 Dec 2017 on shared 510(k) K111663
    • Terminated Z-1230-2018 FDA Class II initiated 20 Dec 2017 on shared 510(k) K111663

    Zimmer Biomet is conducting a medical device recall for the ACE Trochanteric Nail (ATN) system after a Device Performance Review (DPR) failed clinical analysis for survivorship pursuant to the acceptance criteria for the DPR protocol for the ATN system.

    • Terminated Z-1306-2018 FDA Class II initiated 25 Oct 2017

    The drill guides were manufactured with the incorrect color anodization.

    • Terminated Z-1817-2016 FDA Class II initiated 1 Apr 2016

    Insufficient seal strength on sterile barrier packaging. The seal could be jeopardized causing the sterile barrier to be compromised and the product could become nonsterile. If the product becomes non-sterile and is used or implanted, it may lead to an infection.

    • Terminated Z-0450-2016 FDA Class II initiated 20 Nov 2015

    These screws have a single lead (instead of a triple lead) thread. The screw may sit proud in the plate, If the screw does not lock into the plate; the screw may back out. If the screw sits proud, soft tissue irritation may occur. If the screw does not lock into the plate and backs out, a revision surgery may be necessary.

    • Terminated Z-2058-2015 FDA Class II initiated 4 Jun 2015

    Sterile packaging was not sealed. The implantation of an unsterile screw may lead to infection.

    • Terminated Z-2046-2015 FDA Class II initiated 4 Jun 2015

    The thread location on one of the thread holes is offset from the axis of the predrill hole. The threads are too deep on one side and too shallow on the other. The locking screw may back out of the plate if it does not achieve proper purchase. This may require a revision surgery to replace or remove the screw and/or plate. The threads, not uniformly loaded, may shear off, causing metal slivers.

Recalls on a shared submission name another device cleared under the same FDA submission as this one.

Timeline

18 dated events

  1. Recall initiated FDA recalls Z-1227-2018
  2. Recall initiated FDA recalls Z-1229-2018
  3. Recall initiated FDA recalls Z-1233-2018
  4. Recall initiated FDA recalls Z-1225-2018
  5. Recall initiated FDA recalls Z-1228-2018
  6. Recall initiated FDA recalls Z-1226-2018
  7. Recall initiated FDA recalls Z-1232-2018
  8. Recall initiated FDA recalls Z-1235-2018
  9. Recall initiated FDA recalls Z-1230-2018
  10. Recall initiated FDA recalls Z-0811-2022

8 earlier events are in the sections above.

Coverage

1 jurisdiction

United States of America
  • Active
  • Expiring
  • Withdrawn
  • Listed
  • Not covered

United States — FDA 510(k) Active

Specification

510(k)

510(k) Number
FDA product code
HRS
Device class
Class II
Regulation
21 CFR Part 888.3030
Advisory committee
Orthopedic
FDA decision
SESE Substantially Equivalent

Other devices from Depuy Orthopaedics, Inc.

See all devices from Depuy Orthopaedics, Inc. →