Triage 97000HS
Triage 97000HS is a multiple cardiac marker IVD (FDA Class II) made by Quidel Cardiovascular Inc., cleared under FDA 510(k) K030286 on 21 Feb 2003 and listed in FDA GUDID since 1 Nov 2018 (UDI-DI 30014613337911), active as of 23 Oct 2026. 2 FDA recalls are linked to it.
Active
Valid until
No expiry
FDA recalls
1 ongoingDocuments
Trust
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | GUDID listing | Number30014613337911 | Date1 Nov 2018 | StatusListed |
| United States | FDA 510(k)SESE | NumberK030286 | Date21 Feb 2003 | StatusActive |
Official records
- AccessGUDID record 30014613337911
- FDA 510(k) database K030286
From the FDA record
- Device description
- Triage Cardiac Panel
- MRI safety
- Labeling does not contain MRI Safety Information
- Natural rubber latex
- No latex statement on the label
- Sterility
- Supplied non-sterile
Across registries
- FDA 510(k) K030286 decided 21 Feb 2003 Active
- FDA GUDID UDI-DI 30014613337911 listed 1 Nov 2018 Listed
Recalls
-
- Ongoing Z-1974-2023
Raw material issue may cause cardiac panel products to have an average of 30% troponin signal reduction across measurement range 0.05 ng/mL to 30 ng/mL, QC testing won't detect anomaly, so an alternate method is recommended, if unavailable, use recommendations to mitigate potential patient impact, which could be missed/delayed myocardial infarction diagnosis.
-
- Terminated Z-1929-2012
Certain lots of the affected products may have significantly decreased precision relative to the package insert, which could result in an increased frequency of false positive or false negative results. These false positive and false negative results are unpredictable within lots and may not be detected by Quality Control testing.
-
- Terminated Z-1391-2009
The recall was initiated because the use of the affected lot may lead to reporting falsely low values for troponin I and myoglobin, possibly resulting in missed or incorrect diagnosis.
Timeline
- Cleared FDA 510(k) K030286
- Recall initiated FDA recalls Z-1391-2009
- Recall initiated FDA recalls Z-1929-2012
- Listed FDA GUDID 30014613337911
- Recall initiated FDA recalls Z-1974-2023
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- Trade name
- Triage
- Generic name
- Multiple cardiac marker IVD, kit, fluorescent immunoassay
- Model / Ref / Catalog
- 97000HS
- Primary DI / UDI-DI
- 30014613337911
- 510(k) Number
- FDA product code
- MMI, DDR, JXH
- GMDN term
- Multiple cardiac marker IVD, kit, fluorescent immunoassay
- Device class
- Class II
- Regulation number
- 862.1215, 866.5680, 882.5030
- GUDID distribution status
- In Commercial Distribution
Similar devices
Other devices filed under FDA product code MMI (IMMUNOASSAY METHOD, TROPONIN SUBUNIT).
- STRATUS (R) CS CARDIAC TROPONIN I (TROP) TESTPAK by Dade Behring, Inc.
- ABBOTT AXSYM TROPONIN-I ADV by Abbott Laboratories
- AIA-PACK CTNI 2ND-GEN ASSAY by Tosoh Medics, Inc.
- THE CARDIAC READER SYSTEM by Roche Diagnostics Corp.
- ALPHA DX SYSTEM by First Medical, Inc.
- ELECSYS TROPONIN T by Boehringer Mannheim Corp.
- NANO-CHECK AMI CTNI CARDIAC MARKER TEST by Nano-Ditech Corporation
- ELECSYS TROPONIN T STAT TEST by Roche Diagnostics Corp.
- VANCOUVER BIOTECH HUMAN CARDIAC TROPONIN I IMMUNOASSAY TEST KIT by Vancouver Biotech, Ltd.
- Atellica® IM High-Sensitivity Troponin I (TnIH) by Siemens Healthcare Diagnostics