# Triage 97022HS

> Active: United States. Medical device record on deviCERTS, compiled from public registries.

- Canonical page: <https://devicerts.com/devices/triage-97022hs>
- Manufacturer: [Quidel Cardiovascular Inc.](https://devicerts.com/manufacturers/fda-fei-3013982035) (United States)
- Device class: Class II (US FDA)
- FDA product code: MMI, JHX, NBC
- GMDN: Multiple cardiac marker IVD, kit, fluorescent immunoassay (47384)

## Registrations and authorizations

- **GUDID** · GUDID listing · United States · UDI-DI 30014613337942 · listed 2018-11-01 · status: Listed (GUDID: in commercial distribution)
- **GUDID** · FDA 510(k) · United States · K030286 · decision 2003-02-21 · decision SESE · status: Active · from GUDID

## Recalls

- Z-1974-2023 · Class I · ongoing · initiated 2023-05-25: Raw material issue may cause cardiac panel products to have an average of 30% troponin signal reduction across measurement range 0.05 ng/mL to 30 ng/mL, QC testing won't detect anomaly, so an alternate method is recommended, if unavailable, use recommendations to mitigate potential patient impact, which could be missed/delayed myocardial infarction diagnosis.
- Z-1929-2012 · Class I · terminated · initiated 2012-05-22 · matched through shared submission K030286, may concern other models: Certain lots of the affected products may have significantly decreased precision relative to the package insert, which could result in an increased frequency of false positive or false negative results. These false positive and false negative results are unpredictable within lots and may not be detected by Quality Control testing.
- Z-1391-2009 ·  · terminated · initiated 2009-04-27: The recall was initiated because the use of the affected lot may lead to reporting falsely low values for troponin I and myoglobin, possibly resulting in missed or incorrect diagnosis.

## Identity

- Trade name: Triage [FDA GUDID (AccessGUDID)]
- Generic name: Multiple cardiac marker IVD, kit, fluorescent immunoassay [FDA GUDID (AccessGUDID)]
- Model / Ref / Catalog: 97022HS [FDA GUDID (AccessGUDID)]

## Identifiers

- Primary DI / UDI-DI: 30014613337942 [FDA GUDID (AccessGUDID)]
- 510(k) Number: K030286 [FDA GUDID (AccessGUDID)]
- FDA product code: MMI [FDA GUDID (AccessGUDID)]; JHX [FDA GUDID (AccessGUDID)]
- GMDN term: Multiple cardiac marker IVD, kit, fluorescent immunoassay (47384) [FDA GUDID (AccessGUDID)]

## Classification

- Device class: Class II [FDA GUDID (AccessGUDID)]
- Regulation number: 862.1215 [FDA GUDID (AccessGUDID)]
- GUDID distribution status: In Commercial Distribution [FDA GUDID (AccessGUDID)]

## Properties

- Single use: Yes [FDA GUDID (AccessGUDID)]
- Prescription use: Yes [FDA GUDID (AccessGUDID)]
- Convenience kit: Yes [FDA GUDID (AccessGUDID)]
- Lot / batch number: Yes [FDA GUDID (AccessGUDID)]
- Serial number: Yes [FDA GUDID (AccessGUDID)]
- Expiration date: Yes [FDA GUDID (AccessGUDID)]
- Manufacturing date: Yes [FDA GUDID (AccessGUDID)]
- Supplied sterile: No [FDA GUDID (AccessGUDID)]
- Sterilise before use: No [FDA GUDID (AccessGUDID)]
- Implantable: No [FDA GUDID (AccessGUDID)]
- Contains latex: No [FDA GUDID (AccessGUDID)]
- Over-the-counter: No [FDA GUDID (AccessGUDID)]
- Combination product: No [FDA GUDID (AccessGUDID)]
- HCT/P: No [FDA GUDID (AccessGUDID)]
- Premarket exempt: No [FDA GUDID (AccessGUDID)]

## Source and citation

- Cite as: Triage 97022HS, Quidel Cardiovascular Inc. deviCERTS medical device catalogue. https://devicerts.com/devices/triage-97022hs (data as of 2026-11-23).
- Data sources: [FDA GUDID (AccessGUDID)](https://accessgudid.nlm.nih.gov/), [FDA recalls](https://open.fda.gov/apis/device/recall/)
- Registry records last synced: 2026-10-23
- Statuses computed as of: 2026-11-23
- JSON: <https://devicerts.com/devices/triage-97022hs.json>
- Licence: [CC BY 4.0](https://creativecommons.org/licenses/by/4.0/). The deviCERTS layout and text are licensed under CC BY 4.0. Registry data remain under their source terms: U.S. FDA data (openFDA, AccessGUDID) are U.S. public domain; EUDAMED data are subject to the European Commission reuse terms.
- Attribution: DevicErts, https://devicerts.com/devices/triage-97022hs
- Compiled by deviCERTS (devicerts.com) from public medical device registries (EUDAMED, European Commission; U.S. FDA via openFDA; AccessGUDID, U.S. National Library of Medicine). deviCERTS is not an official register and gives no regulatory advice: the record in the registry itself prevails.