Triage Cardioprofiler and Profiler S.O.B. Panels
510(k) K080269
Active
Valid until
No expiry
FDA recalls
None ongoingDocuments
Trust
Unverified
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK080269 | Date6 Jun 2008 | StatusActive |
Recalls
-
- Terminated Z-1928-2012
- Terminated Z-1930-2012
Certain lots of the affected products may have significantly decreased precision relative to the package insert, which could result in an increased frequency of false positive or false negative results. These false positive and false negative results are unpredictable within lots and may not be detected by Quality Control testing.
-
- Terminated Z-1466-2012
- Terminated Z-1467-2012
Alere San Diego, Inc. is initiating a voluntary recall of the Alere Triage CardioProfiler Panel PN 9700CP, Lot W49569 and the Alere Triage Profiler SOB Panel PN 97300, Lot W48990. Alere San Diego, Inc. has determined that these lots have an increased frequency of Troponin I results >0.05ng/mL for samples which are found to be below 0.05ng/mL upon additional testing.
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA Product Code
- NBC
- Device Class
- Class II
- Regulation
- 21 CFR Part 862.1117
- Advisory Committee
- Clinical Chemistry
- FDA Decision
- SESE Substantially Equivalent