Triage D-Dimer Test, Model 98100

by Biosite Incorporated

510(k) K042890

Active United States

510(k)

Valid until

No expiry

510(k) clearance is not time-limited

FDA recalls

None ongoing

2 closed

Documents

Trust

Unverified

Registry data, not yet checked by deviCERTS. Source FDA 510(k), synced 10 Oct 2026. What this means

Regulatory status

as of registry sync 10 Oct 2026

RegionAuthorizationNumberDateStatus
United StatesFDA 510(k)SESE Substantially Equivalent · Traditional NumberK042890 Date29 Nov 2004ClearedStatusActive

Recalls

2 recalls · FDA

    • Terminated Z-1727-2014 FDA Class II initiated 2 May 2014

    Alere initiated this recall because a limited number of Alere Triage D-Dimer devices from PN 98100 Lot W53884B were incorrectly manufactured. As a result, when testing patient sample these affected devices will either generate an error code (e.g. E4 error), or they may generate an incorrect result.

    • Terminated Z-1932-2012 FDA Class I initiated 22 May 2012

    Certain lots of the affected products may have significantly decreased precision relative to the package insert, which could result in an increased frequency of false positive or false negative results. These false positive and false negative results are unpredictable within lots and may not be detected by Quality Control testing.

Coverage

1 jurisdiction

United States of America
  • Active
  • Expiring
  • Withdrawn
  • Listed
  • Not covered

United States — FDA 510(k) Active

Specification

510(k)

510(k) Number
FDA Product Code
GHH
Device Class
Class II
Regulation
21 CFR Part 864.7320
Advisory Committee
Hematology
FDA Decision
SESE Substantially Equivalent

Other devices from Biosite Incorporated

See all devices from Biosite Incorporated →