Skip to content

Triage Tox Drug Screen Control Level 1 Model 94413 and Level 2 Model 94414

by Biosite Incorporated

510(k) K060788

Active United States

510(k)

Valid until

No expiry

510(k) clearance is not time-limited

FDA recalls

None ongoing

2 closed

Documents

Trust

Unverified

Registry data, not yet checked by deviCERTS. Source FDA 510(k), synced 12 Oct 2026. What this means

Regulatory status

as of registry sync 12 Oct 2026

RegionAuthorizationNumberDateStatus
United StatesFDA 510(k)SESE Substantially Equivalent · Traditional NumberK060788 Date10 Jul 2006ClearedStatusActive

Recalls

2 recalls · FDA

    • Terminated Z-1552-2020 FDA Class II initiated 9 Apr 2019

    Due to an error in the programming of the Control CODE CHIP module for this lot, a failing result is inappropriately displayed for AMP and mAMP. This issue results in the Triage Meter displaying a failing control result for the AMP and mAMP tests when the control did not fail.

    • Terminated Z-0053-2014 FDA Class III initiated 23 Sep 2013

    Alere San Diego, Inc. initiated a voluntary recall of Alere Triage TOX Drug Screen Control 1 (PN 94413, Lot C2681A) because the CODE CHIP that was provided with this lot causes the control to fail when run in the QC Sample Mode using Alere Triage TOX Drug Screeen products. The Alere Triage TOX Drug Screen test devices and patient results are not affected.

Coverage

1 jurisdiction

United States of America
  • Active
  • Expiring
  • Withdrawn
  • Listed
  • Not covered

United States — FDA 510(k) Active

Specification

510(k)

510(k) Number
FDA Product Code
DIF
Device Class
Class I
Regulation
21 CFR Part 862.3280
Advisory Committee
Clinical Toxicology
FDA Decision
SESE Substantially Equivalent

Other devices from Biosite Incorporated