Triathlon Knee System Posterior Stabilized Components

by Stryker Orthopaedics

510(k) K141056

Active United States

510(k)

Valid until

No expiry

510(k) clearance is not time-limited

FDA recalls

1 ongoing

Recall on shared 510(k) K141056; other models may be affected

Documents

Trust

Unverified

Registry data, not yet checked by deviCERTS. Source FDA 510(k), synced 22 Oct 2026. What this means

Regulatory status

as of registry sync 22 Oct 2026

RegionAuthorizationNumberDateStatus
United StatesFDA 510(k)SESE Substantially Equivalent · Traditional NumberK141056 Date5 Aug 2014ClearedStatusActive

Recalls

11 recalls · 1 ongoing · FDA

    • Ongoing Z-0023-2025 FDA Class II initiated 4 Sep 2024 on shared 510(k) K141056

    Anterior locking tabs on the Triathlon Universal Baseplate are out of specification which may prevent engagement of either of the two locking tabs with the locking wire of the tibial insert.

    • Terminated Z-2329-2021 FDA Class II initiated 1 Jul 2021 on shared 510(k) K141056

    The internal profile of the Triathlon Prim (Baseplate) is potentially oversized; there may be a witness mark (taper) at the bottom of the internal profile of the baseplate, which prevents the insert from locking on the baseplate.

    • Terminated Z-1186-2021 FDA Class II initiated 10 Feb 2021

    Stryker has discovered that the internal profile of the Triathlon Tritanium Tibial Component (Baseplate), size 4, is potentially oversized. Due to a manufacturing issue in Lot Number CTD44301, there may be a witness mark (taper) at the bottom of the internal profile of the baseplate. The taper prevents the insert from locking on the baseplate.

  • 4 recalls, same reason

    • Terminated Z-2757-2019 FDA Class II initiated 19 Aug 2019
    • Terminated Z-2760-2019 FDA Class II initiated 19 Aug 2019
    • Terminated Z-2759-2019 FDA Class II initiated 19 Aug 2019
    • Terminated Z-2758-2019 FDA Class II initiated 19 Aug 2019

    Due to device overhang issue that protrudes beyond the medial periphery of the mating femoral component

    • Terminated Z-1462-2018 FDA Class II initiated 30 Jan 2018

    Customers may receive the incorrect size implant from what is labeled on the box.

    • Terminated Z-1512-2018 FDA Class II initiated 12 Jan 2018

    Firm became aware that the specific product/lot combinations referenced above may contain the incorrect size implant from what is labeled on the box.

    • Terminated Z-2389-2015 FDA Class II initiated 29 May 2015

    Stryker has received a report which identified that a Triathlon Tibial Bearing Insert CS X3 Size #2 9mm (5531-G-209) was reported to be in a pack that was labelled as a CR insert (5530-G-209).

    • Terminated Z-2360-2015 FDA Class II initiated 29 May 2015 on shared 510(k) K141056

    Stryker Orthopaedics received a report that 5537-G-319 No 3. Triathlon TS Plus Tibial Insert X3 Poly 19 mm i insert was in a package labeled as a 5532-G-511 Triathlon PS Stibialism Insert.

Recalls on a shared submission name another device cleared under the same FDA submission as this one.

Coverage

1 jurisdiction

United States of America
  • Active
  • Expiring
  • Withdrawn
  • Listed
  • Not covered

United States — FDA 510(k) Active

Specification

510(k)

510(k) Number
FDA Product Code
MBH
Device Class
Class II
Regulation
21 CFR Part 888.3565
Advisory Committee
Orthopedic
FDA Decision
SESE Substantially Equivalent