# Triathlon Ts Knee System

> Active: United States. Medical device record on deviCERTS, compiled from public registries.

- Canonical page: <https://devicerts.com/devices/triathlon-ts-knee-system-jwh>
- Manufacturer: [Howmedica Osteonics Corp.](https://devicerts.com/manufacturers/howmedica-osteonics-corp) (United States)
- Name in the registry: TRIATHLON TS KNEE SYSTEM
- Device class: Class II (US FDA)
- FDA product code: JWH

## Registrations and authorizations

- **510(k)** · FDA 510(k) · United States · K070095 · decision 2007-06-01 · decision SESE · status: Active

## Recalls

- Z-1635-2015 · Class II · terminated · initiated 2014-08-15: The peel strength of the inner blister may have been below internal validated requirements. The strength of the seal is directly related to long term shelf life of the sterile barrier.
- Z-1370-2013 · Class II · terminated · initiated 2013-03-25: Stryker Orthopaedics received a report that a label on the external packaging (box) of a Triathalon Femoral Distal Augment 10mm- Size 5 Left depicted a glyph of an acetabular shell rather than a femoral distal augment. In addition the label was missing a reference to the "10mm" augment thickness which per the label print is required to appear below the size reference.
- Z-0757-2010 ·  · terminated · initiated 2009-12-10: Triathlon TS femoral components may have been damaged during manufacturing, possibly preventing the assembly of a stem extension.
- Z-0758-2010 ·  · terminated · initiated 2009-12-10: Triathlon TS femoral components may have been damaged during manufacturing, possibly preventing the assembly of a stem extension.
- Z-0759-2010 ·  · terminated · initiated 2009-12-10: Triathlon TS femoral components may have been damaged during manufacturing, possibly preventing the assembly of a stem extension.
- Z-0760-2010 ·  · terminated · initiated 2009-12-10: Triathlon TS femoral components may have been damaged during manufacturing, possibly preventing the assembly of a stem extension.
- Z-0761-2010 ·  · terminated · initiated 2009-12-10: Triathlon TS femoral components may have been damaged during manufacturing, possibly preventing the assembly of a stem extension.
- Z-1607-2009 ·  · terminated · initiated 2009-04-16: Triathlon Offset Adaptors may seize during surgery and the OR staff may be unable to disassemble the instrument using the removal tool.
- Z-1608-2009 ·  · terminated · initiated 2009-04-16: Triathlon Offset Adaptors may seize during surgery and the OR staff may be unable to disassemble the instrument using the removal tool.
- Z-1609-2009 ·  · terminated · initiated 2009-04-16: Triathlon Offset Adaptors may seize during surgery and the OR staff may be unable to disassemble the instrument using the removal tool.
- Z-1610-2009 ·  · terminated · initiated 2009-04-16: Triathlon Offset Adaptors may seize during surgery and the OR staff may be unable to disassemble the instrument using the removal tool.
- Z-0890-2009 ·  · terminated · initiated 2008-11-13: Stryker Orthopaedics became aware of a visual defect in certain blister packages of certain devices.
- Z-1231-2007 ·  · terminated · initiated 2007-07-31: Stabilizer posts, packaged with the Triathlon TS Plus Tibial inserts have a specification anomaly related to length, resulting in a potential for the post to slightly protrude above the top surface of the plastic post.

## Identifiers

- 510(k) Number: K070095 [FDA 510(k)]
- FDA product code: JWH [FDA 510(k)]

## Classification

- Device class: Class II [FDA 510(k)]
- Regulation: 21 CFR Part 888.3560 [FDA 510(k)]
- Advisory committee: Orthopedic [FDA 510(k)]
- FDA decision: SESE [FDA 510(k)]

## Source and citation

- Cite as: Triathlon Ts Knee System, Howmedica Osteonics Corp. deviCERTS medical device catalogue. https://devicerts.com/devices/triathlon-ts-knee-system-jwh (data as of 2026-11-29).
- Data sources: [FDA 510(k)](https://open.fda.gov/apis/device/510k/), [FDA recalls](https://open.fda.gov/apis/device/recall/)
- Registry records last synced: 2026-10-22
- Statuses computed as of: 2026-11-29
- JSON: <https://devicerts.com/devices/triathlon-ts-knee-system-jwh.json>
- Licence: [CC BY 4.0](https://creativecommons.org/licenses/by/4.0/). The deviCERTS layout and text are licensed under CC BY 4.0. Registry data remain under their source terms: U.S. FDA data (openFDA, AccessGUDID) are U.S. public domain; EUDAMED data are subject to the European Commission reuse terms.