Trilogy 200 Ventilator, Germany
Ref 1032802Model Trilogy 200
Trilogy 200 Ventilator, Germany (1032802), a portable electric ventilator (EU MDD Class IIb; FDA Class II) made by Respironics Inc., was registered in EUDAMED on 23 Jul 2025 (Basic UDI-DI B-00606959022690), cleared under FDA 510(k) K093416 on 29 Jan 2010 and listed in FDA GUDID on 22 Oct 2015 (UDI-DI 00606959022690); the registry lists it as withdrawn (EUDAMED) as of 24 Oct 2026. FDA recalls on record: 5.
Active
Valid until
No expiry
FDA recalls
2 ongoingTrust
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| European Union | CE/MDD EUDAMED | NumberB-00606959022690 00606959022690 | Date23 Jul 2025 | StatusWithdrawn |
| United States | GUDID listing GUDID | Number00606959022690 | Date22 Oct 2015 | StatusListed |
| United States | FDA 510(k)SESE GUDID | NumberK093416 | Date29 Jan 2010 | StatusActive |
Official records
- EUDAMED public site B-00606959022690
- AccessGUDID record 00606959022690
- FDA 510(k) database K093416
From the FDA record
- MRI safety
- Labeling does not contain MRI Safety Information
- Natural rubber latex
- No latex statement on the label
- Sterility
- Supplied non-sterile
Across registries
- EUDAMED Basic UDI-DI B-00606959022690 registered 23 Jul 2025 Withdrawn
- FDA 510(k) K093416 decided 29 Jan 2010 Active
- FDA GUDID UDI-DI 00606959022690 listed 22 Oct 2015 Listed
Recalls
-
- Ongoing Z-0883-2023
The replacement silicone sound abatement foam installed into some Trilogy 100 and Trilogy 200 devices may separate from the plastic backing to which it is attached. If this were to happen, the foam could potentially block air inlet, which could result in a reduction in delivered therapy volume or pressure and could cause the device to alarm.
-
- Ongoing Z-1956-2021
The polyester-based polyurethane (PE-PUR) foam may degrade into particles which may enter the devices air pathway and be ingested or inhaled by the user, and the PE-PUR foam may off-gas certain chemicals.
-
- Terminated Z-1065-2016
Software Issue
-
- Terminated Z-1143-2014
Internal testing revealed a potentially defective component on the power management board that could affect the function of certain Trilogy Ventilators resulting in loss of therapy to patients.
-
- Terminated Z-2024-2012
The specified Trilogy units are being recalled because of the potential for a faulty component in the power supply.
Timeline
- Cleared FDA 510(k) K093416
- Recall initiated FDA recalls Z-2024-2012
- Recall initiated FDA recalls Z-1143-2014
- Listed FDA GUDID 00606959022690
- Recall initiated FDA recalls Z-1065-2016
- Recall initiated FDA recalls Z-1956-2021
- Recall initiated FDA recalls Z-0883-2023
- Registered EUDAMED B-00606959022690
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
- European Union
- United States
Specification
- Trade name
- Trilogy 200 Ventilator, Germany EUDAMEDTrilogy GUDID
- Generic name
- Portable electric ventilator
- Model / Ref / Catalog
- Trilogy 200, 1032802, Trilogy200 Ventilator, Germany
- Primary DI / UDI-DI
- 00606959022690
- Basic UDI-DI
- B-00606959022690
- 510(k) Number
- FDA product code
- DQA, NOU, CBK
- EMDN code
- Z12030103 Pulmonary ventilators for non–hospital use
- GMDN term
- Portable electric ventilator
- Device class
- Class IIb EUDAMEDClass II GUDID
- Regulation number
- 870.2700, 868.5895
- EU registration type
- Legacy device (under a directive)
- EU market of registration
- Germany
- GUDID distribution status
- In Commercial Distribution
Truth per registry
| Field | EUDAMED | GUDID |
|---|---|---|
| Model / reference | Trilogy 200 1032802 | Trilogy200 Ventilator, Germany |
| Device class | Class IIb | Class II |
Manufacturer details
- EUDAMED SRN
- US-MF-000002301
- Authorised representative
- Respironics Deutschland GmbH & Co.KG DE-AR-000006164
Similar devices in this category
Other devices filed under ANAESTHESIA AND PULMONARY VENTILATION SUPPORT INSTRUMENTS.
- Vivo 2 228000 by Breas Medical AB
- Vivo 3 229000 by Breas Medical AB
- Microvent 670-0411-00 by B.N.O.S. Meditech Limited
- Accutron Inc. 27076 by ACCUTRON
- Carestation 620 A1 1012-9620-000 by Datex-Ohmeda Inc.
- Sigma Delta Vaporizer / Sevoflurane 410502S by Penlon Limited
- Accutron Inc. 35350 by ACCUTRON
- Aladin2 Sevoflurane Easy-Fil 1100-9030-000 by Datex-Ohmeda Inc.
- Anesthesia Machine A9C by Shenzhen Mindray Bio-Medical Electronics Co., Ltd.
- Perseus A500 MK06000 by Drägerwerk AG & Co. KGaA