Skip to content

Trilogy 200 Ventilator, Germany-Recert

by Respironics Inc.

Ref U1032802

Trilogy 200 Ventilator, Germany-Recert (U1032802), a portable electric ventilator (EU MDD Class IIb; FDA Class II) made by Respironics Inc., was registered in EUDAMED on 26 Feb 2025 (Basic UDI-DI B-00606959033924), cleared under FDA 510(k) K093416 on 29 Jan 2010 and listed in FDA GUDID on 23 Sep 2016 (UDI-DI 00606959033924); the registry lists it as withdrawn (EUDAMED) as of 23 Oct 2026. FDA recalls on record: 5.

Active Active in United States. Withdrawn in European Union. Also listed in GUDID (US).

CE MDD, withdrawn 510(k)

Valid until

No expiry

510(k) clearance is not time-limited

FDA recalls

2 ongoing

This device is named in the recall

Documents

None linked in EUDAMED

A group certificate may exist

510(k) K093416

FDA submission, from GUDID

Trust

Unverified

Source EUDAMED, GUDID, synced 23 Oct 2026.

Regulatory status

as of registry sync 23 Oct 2026

RegionAuthorizationNumberDateStatus
European UnionCE/MDD EUDAMED NumberB-00606959033924 Basic UDI-DI 00606959033924UDI-DIDate26 Feb 2025RegisteredStatusWithdrawn No longer placed on the EU market
United StatesGUDID listing GUDID Number00606959033924 UDI-DI Date23 Sep 2016ListedStatusListed GUDID: in commercial distribution
United StatesFDA 510(k)SESE GUDID from GUDID Substantially Equivalent · Traditional NumberK093416 Date29 Jan 2010DecidedStatusActive

Official records

From the FDA record

MRI safety
Labeling does not contain MRI Safety Information
Natural rubber latex
No latex statement on the label
Sterility
Supplied non-sterile

From the GUDID record (AccessGUDID), as read on 30 May 2026, as the labeller entered it.

Across registries

3 registries list this device

  • EUDAMED Basic UDI-DI B-00606959033924 registered 26 Feb 2025 Withdrawn
  • FDA 510(k) K093416 decided 29 Jan 2010 Active
  • FDA GUDID UDI-DI 00606959033924 listed 23 Sep 2016 Listed

Recalls

5 recalls · 2 ongoing · FDA

    • Ongoing Z-0883-2023 FDA Class I initiated 7 Dec 2022

    The replacement silicone sound abatement foam installed into some Trilogy 100 and Trilogy 200 devices may separate from the plastic backing to which it is attached. If this were to happen, the foam could potentially block air inlet, which could result in a reduction in delivered therapy volume or pressure and could cause the device to alarm.

    • Ongoing Z-1956-2021 FDA Class I initiated 14 Jun 2021

    The polyester-based polyurethane (PE-PUR) foam may degrade into particles which may enter the devices air pathway and be ingested or inhaled by the user, and the PE-PUR foam may off-gas certain chemicals.

    • Terminated Z-1065-2016 FDA Class II initiated 18 Jan 2016

    Software Issue

    • Terminated Z-1143-2014 FDA Class I initiated 11 Feb 2014

    Internal testing revealed a potentially defective component on the power management board that could affect the function of certain Trilogy Ventilators resulting in loss of therapy to patients.

    • Terminated Z-2024-2012 FDA Class I initiated 27 Apr 2012

    The specified Trilogy units are being recalled because of the potential for a faulty component in the power supply.

Timeline

8 dated events

  1. Cleared FDA 510(k) K093416
  2. Recall initiated FDA recalls Z-2024-2012
  3. Recall initiated FDA recalls Z-1143-2014
  4. Recall initiated FDA recalls Z-1065-2016
  5. Listed FDA GUDID 00606959033924
  6. Recall initiated FDA recalls Z-1956-2021
  7. Recall initiated FDA recalls Z-0883-2023
  8. Registered EUDAMED B-00606959033924

Coverage

2 jurisdictions

United States of AmericaSwedenFinlandLatviaEstoniaDenmarkLithuaniaIrelandNetherlandsGermanyPolandFranceBelgiumLuxembourgCzech RepublicSlovakiaAustriaHungaryRomaniaSloveniaBulgariaPortugalSpainItalyCroatiaGreeceMaltaCyprus
  • Active
  • Expiring
  • Withdrawn
  • Listed
  • Not covered
  • European Union CE/MDD Withdrawn
  • United States FDA 510(k) Active

Specification

EUDAMED · GUDID

Trade name
Trilogy 200 Ventilator, Germany-Recert EUDAMED
Trilogy 200 GUDID
Generic name
Portable electric ventilator
Model / Ref / Catalog
U1032802, Trilogy 200 Ventilator, Germany-Recert
Primary DI / UDI-DI
00606959033924
Basic UDI-DI
B-00606959033924
510(k) Number
FDA product code
CBK, OUG, DQA
EMDN code
Z12030103 Pulmonary ventilators for non–hospital use
GMDN term
Portable electric ventilator
Device class
Class IIb EUDAMED
Class II GUDID
Regulation number
868.5895, 880.6310, 870.2700
EU registration type
Legacy device (under a directive)
EU market of registration
Germany
GUDID distribution status
In Commercial Distribution

Truth per registry

2 sources · 2 fields differ

FieldEUDAMEDGUDID
Model / referenceEUDAMED
U1032802
GUDID
Trilogy 200 Ventilator, Germany-Recert
Device classEUDAMED
Class IIb
GUDID
Class II

Device class is stated on each registry's own scale: Class IIb in the EU (EUDAMED), Class II in the US (FDA GUDID). EU and US classes are different schemes, so this is not a disagreement.

Manufacturer details

EUDAMED SRN
US-MF-000002301
Authorised representative
Respironics Deutschland GmbH & Co.KG DE-AR-000006164

Similar devices in this category

Other devices filed under ANAESTHESIA AND PULMONARY VENTILATION SUPPORT INSTRUMENTS.

Other devices from Respironics Inc.

See all devices from Respironics Inc. →

Properties

  • Prescription use
  • Serial number
Other properties (23, all answered No)