Trilogy Trilogy200 Ventilator, Australia
Trilogy Trilogy200 Ventilator, Australia (AU1032800) is a portable electric ventilator (FDA Class II) made by Respironics Inc., cleared under FDA 510(k) K093416 on 29 Jan 2010 and listed in FDA GUDID since 22 Oct 2015 (UDI-DI 00606959022843), active as of 23 Oct 2026. 5 FDA recalls are linked to it.
Active
Valid until
No expiry
FDA recalls
2 ongoingDocuments
Trust
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | GUDID listing | Number00606959022843 | Date22 Oct 2015 | StatusListed |
| United States | FDA 510(k)SESE | NumberK093416 | Date29 Jan 2010 | StatusActive |
Official records
- AccessGUDID record 00606959022843
- FDA 510(k) database K093416
From the FDA record
- Device description
- Trilogy 200 Ventilator, Australia
- MRI safety
- Labeling does not contain MRI Safety Information
- Natural rubber latex
- No latex statement on the label
- Sterility
- Supplied non-sterile
Across registries
- FDA 510(k) K093416 decided 29 Jan 2010 Active
- FDA GUDID UDI-DI 00606959022843 listed 22 Oct 2015 Listed
Recalls
-
- Ongoing Z-0883-2023
The replacement silicone sound abatement foam installed into some Trilogy 100 and Trilogy 200 devices may separate from the plastic backing to which it is attached. If this were to happen, the foam could potentially block air inlet, which could result in a reduction in delivered therapy volume or pressure and could cause the device to alarm.
-
- Ongoing Z-1956-2021
The polyester-based polyurethane (PE-PUR) foam may degrade into particles which may enter the devices air pathway and be ingested or inhaled by the user, and the PE-PUR foam may off-gas certain chemicals.
-
- Terminated Z-1065-2016
Software Issue
-
- Terminated Z-1143-2014
Internal testing revealed a potentially defective component on the power management board that could affect the function of certain Trilogy Ventilators resulting in loss of therapy to patients.
-
- Terminated Z-2024-2012
The specified Trilogy units are being recalled because of the potential for a faulty component in the power supply.
Timeline
- Cleared FDA 510(k) K093416
- Recall initiated FDA recalls Z-2024-2012
- Recall initiated FDA recalls Z-1143-2014
- Listed FDA GUDID 00606959022843
- Recall initiated FDA recalls Z-1065-2016
- Recall initiated FDA recalls Z-1956-2021
- Recall initiated FDA recalls Z-0883-2023
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- Trade name
- Trilogy
- Generic name
- Portable electric ventilator
- Model / Ref / Catalog
- Trilogy200 Ventilator, Australia, AU1032800
- Primary DI / UDI-DI
- 00606959022843
- 510(k) Number
- FDA product code
- DQA, CBK, NOU
- GMDN term
- Portable electric ventilator
- Device class
- Class II
- Regulation number
- 870.2700, 868.5895
- GUDID distribution status
- In Commercial Distribution
Similar devices
Other devices filed under FDA product code DQA (Oximeter).
- SpO2.COM® RS-1 adult SpO2 reusable sensor 1778 by MASIMO CORPORATION
- LNOP Blue Oximetry Sensor by MASIMO CORPORATION
- Fingertip Pulse Oximeter by SHENZHEN BSX TECHNOLOGY ELECTRONICS CO.,LTD.
- Oximax Softcare Sensors Models SC-A, SC-PR, SC-NEO, SC-N by Nellcor Puritan Bennett, Inc.
- Bci WW1021 Pc Application by Smiths Medical Pm, Inc.
- FINGERTIP PULSE OXIMETER, MODEL CMS-50C by Contec Medical System Co., Ltd.
- Sempulse Halo Vital Signs Monitor by Sempulse Corporation
- BURDICK 200 PULSE OXIMETER by Burdick Corp.
- OXICARE FINGERTIP PULSE OXIMETER by Delbio Incorporated
- FORE-SIGHT, MC-2000 SERIES by CAS MEDICAL SYSTEMS, INC.