  "format": "devicerts.device.v1",
  "cite_as": "Trilogy Evo Universal, Respironics, Inc. deviCERTS medical device catalogue. https://devicerts.com/devices/trilogy-evo-universal-cbk (data as of 2026-11-21).",
  "source": {
    "canonical_url": "https://devicerts.com/devices/trilogy-evo-universal-cbk",
    "json_url": "https://devicerts.com/devices/trilogy-evo-universal-cbk.json",
    "markdown_url": "https://devicerts.com/devices/trilogy-evo-universal-cbk.md",
    "publisher": "deviCERTS (devicerts.com)",
    "data_sources": [
      {
        "registry": "FDA 510(k)",
        "url": "https://open.fda.gov/apis/device/510k/"
      },
      {
        "registry": "FDA recalls",
        "url": "https://open.fda.gov/apis/device/recall/"
      }
    ],
    "registry_records_last_synced": "2026-10-22",
    "status_as_of": "2026-11-21",
    "license": "CC BY 4.0",
    "license_url": "https://creativecommons.org/licenses/by/4.0/",
    "license_scope": "The deviCERTS layout and text are licensed under CC BY 4.0. Registry data remain under their source terms: U.S. FDA data (openFDA, AccessGUDID) are U.S. public domain; EUDAMED data are subject to the European Commission reuse terms.",
    "attribution": "DevicErts, https://devicerts.com/devices/trilogy-evo-universal-cbk",
    "disclaimer": "Compiled by deviCERTS (devicerts.com) from public medical device registries (EUDAMED, European Commission; U.S. FDA via openFDA; AccessGUDID, U.S. National Library of Medicine). deviCERTS is not an official register and gives no regulatory advice: the record in the registry itself prevails."
  },
  "device": {
    "name": "Trilogy Evo Universal",
    "registry_name": "Trilogy Evo Universal",
    "url": "https://devicerts.com/devices/trilogy-evo-universal-cbk",
    "manufacturer": {
      "name": "Respironics, Inc.",
      "url": "https://devicerts.com/manufacturers/respironics-inc",
      "country": "United States",
      "country_basis": null,
      "fda_registration_country": null
    },
    "status": {
      "state": "active",
      "label": "Active",
      "detail": "United States"
    },
    "classification": [
      {
        "class": "Class II",
        "system": "US FDA"
      }
    ],
    "identifiers": [
      {
        "label": "510(k) Number",
        "values": [
          {
            "value": "K181170",
            "registries": [
              "FDA 510(k)"
            ]
          }
        ]
      }
    ],
    "product_codes": [
      "CBK"
    ],
    "emdn": [],
    "gmdn": [],
    "description": null,
    "noindex": false
  },
  "authorizations": [
    {
      "registry": "510(k)",
      "region": "United States",
      "type": "FDA 510(k)",
      "number": "K181170",
      "number_label": null,
      "device_udi_di": null,
      "date": "2019-07-18",
      "date_kind": "decision",
      "decision_code": "SESE",
      "expires": null,
      "withdrawn": null,
      "status": "Active",
      "status_note": null,
      "derived_from_gudid": false
    }
  ],
  "no_longer_listed_by": [],
  "certificates": [],
  "recalls": [
    {
      "number": "Z-1617-2026",
      "classification": "Class I",
      "status": "ongoing",
      "initiated": "2026-03-02",
      "reason": "Using non-pneumatic nebulizers with Trilogy Evo Platform ventilators may result in a discrepancy between the set tidal volume and the tidal volume received by the patient.",
      "applies": "this device",
      "via": "K181170"
    },
    {
      "number": "Z-1649-2026",
      "classification": "Class I",
      "status": "ongoing",
      "initiated": "2026-03-02",
      "reason": "In some situations, the Obstruction Alarm does not trigger within the timeframe dictated by the relevant standards and may take up to four breaths.",
      "applies": "this device",
      "via": "K181170"
    },
    {
      "number": "Z-0292-2025",
      "classification": "Class I",
      "status": "ongoing",
      "initiated": "2024-10-07",
      "reason": "Potential for in-line nebulizer configuration in certain locations to result in aerosol deposit accumulation on the device's internal flow sensor, which could lead to inaccurate flow measurements.",
      "applies": "this device",
      "via": "K181170"
    },
    {
      "number": "Z-2943-2024",
      "classification": "Class I",
      "status": "ongoing",
      "initiated": "2024-07-16",
      "reason": "Firm has released a mandatory software update Version 1.05.10.00 and a User Manual addendum for affected ventilators. These updates resolve safety issues identified in earlier recalls.",
      "applies": "this device",
      "via": "K181170"
    },
    {
      "number": "Z-2944-2024",
      "classification": "Class I",
      "status": "ongoing",
      "initiated": "2024-07-16",
      "reason": "Firm has released a mandatory software update Version 1.05.10.00 and a User Manual addendum for affected ventilators. These updates resolve safety issues identified in earlier recalls.",
      "applies": "shared submission",
      "via": "K181170"
    },
    {
      "number": "Z-2942-2024",
      "classification": "Class I",
      "status": "ongoing",
      "initiated": "2024-07-16",
      "reason": "Firm has released a mandatory software update Version 1.05.10.00 and a User Manual addendum for affected ventilators. These updates resolve safety issues identified in earlier recalls.",
      "applies": "shared submission",
      "via": "K181170"
    },
    {
      "number": "Z-1504-2024",
      "classification": "Class I",
      "status": "ongoing",
      "initiated": "2024-03-19",
      "reason": "Potential for sudden loss of ventilation due to Battery Depleted or Loss of Power alarm while unit has sufficient power.",
      "applies": "this device",
      "via": "K181170"
    },
    {
      "number": "Z-1505-2024",
      "classification": "Class I",
      "status": "ongoing",
      "initiated": "2024-03-19",
      "reason": "Potential for sudden loss of ventilation due to Battery Depleted or Loss of Power alarm while unit has sufficient power.",
      "applies": "this device",
      "via": "K181170"
    },
    {
      "number": "Z-1652-2023",
      "classification": "Class I",
      "status": "ongoing",
      "initiated": "2023-05-01",
      "reason": "The Trilogy Evo Universal ventilator provides continuous or intermittent positive pressure ventilation for the care of individuals who require mechanical ventilation. Trilogy Evo Universal is intended for pediatric through adult patients weighing at least 2.5 kg. The ventilator can measure, display, record, and alarm SpO2, FiO2, CO2, and pulse rate data when integrated with the appropriate accessories. The ventilator is suitable for use in institutional and hospital settings and non-emergency transport settings; for example, wheelchair. It may be used for both invasive and non-invasive ventilation.",
      "applies": "this device",
      "via": "K181170"
    },
    {
      "number": "Z-1289-2023",
      "classification": "Class I",
      "status": "ongoing",
      "initiated": "2023-03-01",
      "reason": "The accuracy of delivered oxygen may deviate below the required tolerance of 5% from setpoint when providing high concentration oxygen therapy. Additionally, if equipped, the internal FiO2 sensor may indicate a value higher than the device is actually delivering. This may lead to under delivery of oxygen.",
      "applies": "this device",
      "via": "K181170"
    },
    {
      "number": "Z-0493-2022",
      "classification": "Class I",
      "status": "ongoing",
      "initiated": "2021-12-22",
      "reason": "Philips has identified a specific lot of non-conforming material manufactured by one of its suppliers.  A Philips supplier incorrectly used polyester-based polyurethane (PE-PUR) sound abatement foam, a non-conforming material in the devices and repair kits. PE-PUR foam may degrade into particles which may enter the device s air pathway and be ingested or inhaled by the user. The foam degradation may be exacerbated by use of unapproved cleaning methods such as unapproved ozone cleaners, and the PE-PUR foam may off-gas certain chemicals during initial use of the device.",
      "applies": "this device",
      "via": "K181170"
    },
    {
      "number": "Z-0494-2022",
      "classification": "Class I",
      "status": "ongoing",
      "initiated": "2021-12-22",
      "reason": "Philips has identified a specific lot of non-conforming material manufactured by one of its suppliers.  A Philips supplier incorrectly used polyester-based polyurethane (PE-PUR) sound abatement foam, a non-conforming material in the devices and repair kits. PE-PUR foam may degrade into particles which may enter the device s air pathway and be ingested or inhaled by the user. The foam degradation may be exacerbated by use of unapproved cleaning methods such as unapproved ozone cleaners, and the PE-PUR foam may off-gas certain chemicals during initial use of the device.",
      "applies": "this device",
      "via": "K181170"
    },
    {
      "number": "Z-2290-2021",
      "classification": "Class II",
      "status": "ongoing",
      "initiated": "2021-07-26",
      "reason": "Two software issues have been identified related to pressure increase in the device: The first issue is described as \"Infant/Pediatric EFS Calibration Pressure Increase\" and the second issue is described as \"Pressure Drift (Continuous Usage).",
      "applies": "this device",
      "via": "K181170"
    }
  ],
  "eu_market_availability": null,
  "properties": [
    {
      "group": "Identifiers",
      "fields": [
        {
          "label": "510(k) Number",
          "values": [
            {
              "value": "K181170",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        },
        {
          "label": "FDA product code",
          "values": [
            {
              "value": "CBK",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        }
      ]
    },
    {
      "group": "Classification",
      "fields": [
        {
          "label": "Device class",
          "values": [
            {
              "value": "Class II",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        },
        {
          "label": "Regulation",
          "values": [
            {
              "value": "21 CFR Part 868.5895",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        },
        {
          "label": "Advisory committee",
          "values": [
            {
              "value": "Anesthesiology",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        },
        {
          "label": "FDA decision",
          "values": [
            {
              "value": "SESE",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        }
      ]
    }
  ]
}