# Trilogy R1119867

> Active: United States. Medical device record on deviCERTS, compiled from public registries.

- Canonical page: <https://devicerts.com/devices/trilogy-r1119867>
- Manufacturer: [Respironics Inc.](https://devicerts.com/manufacturers/respironicsinc) (United States)
- Device class: Class I (US FDA)
- FDA product code: OUG, DQA, CBK
- GMDN: Portable electric ventilator (47083)

## Registrations and authorizations

- **GUDID** · GUDID listing · United States · UDI-DI 00606959038851 · listed 2020-02-26 · status: Listed (GUDID: in commercial distribution)
- **GUDID** · FDA 510(k) · United States · K083526 · decision 2009-03-13 · decision SESE · status: Active · from GUDID

## Recalls

- Z-0882-2023 · Class I · ongoing · initiated 2022-12-07: The replacement silicone sound abatement foam installed into some Trilogy 100 and Trilogy 200 devices may separate from the plastic backing to which it is attached. If this were to happen, the foam could potentially block air inlet, which could result in a reduction in delivered therapy volume or pressure and could cause the device to alarm.
- Z-0884-2023 · Class I · ongoing · initiated 2022-12-07: The replacement silicone sound abatement foam installed into some Trilogy 100 and Trilogy 200 devices may separate from the plastic backing to which it is attached. If this were to happen, the foam could potentially block air inlet, which could result in a reduction in delivered therapy volume or pressure and could cause the device to alarm.
- Z-1956-2021 · Class I · ongoing · initiated 2021-06-14: The polyester-based polyurethane (PE-PUR) foam may degrade into particles which may enter the devices air pathway and be ingested or inhaled by the user, and the PE-PUR foam may off-gas certain chemicals.
- Z-1065-2016 · Class II · terminated · initiated 2016-01-18: Software Issue
- Z-1143-2014 · Class I · terminated · initiated 2014-02-11: Internal testing revealed a potentially defective component on the power management board that could affect the function of certain Trilogy Ventilators resulting in loss of therapy to patients.
- Z-2024-2012 · Class I · terminated · initiated 2012-04-27: The specified Trilogy units are being recalled because of the potential for a faulty component in the power supply.
- Z-0296-2012 ·  · terminated · initiated 2011-10-11: Trilogy 100, 200 and 202 units are being recalled due to low impeller press-force used during the manufacturing process. If the press force were low enough, theoretically the impeller could move up the shaft and come into contact with the inside of the blower housing/impeller enclosure. This may impact therapy.
- Z-1896-2010 ·  · terminated · initiated 2010-06-10: It is possible that the battery could electrically short, resulting in a potential for fire or a thermal event.

## Identity

- Trade name: Trilogy [FDA GUDID (AccessGUDID)]
- Generic name: Portable electric ventilator [FDA GUDID (AccessGUDID)]
- Model / Ref / Catalog: R1119867 [FDA GUDID (AccessGUDID)]

## Identifiers

- Primary DI / UDI-DI: 00606959038851 [FDA GUDID (AccessGUDID)]
- 510(k) Number: K083526 [FDA GUDID (AccessGUDID)]
- FDA product code: OUG [FDA GUDID (AccessGUDID)]; DQA [FDA GUDID (AccessGUDID)]; CBK [FDA GUDID (AccessGUDID)]
- GMDN term: Portable electric ventilator (47083) [FDA GUDID (AccessGUDID)]

## Classification

- Device class: Class I [FDA GUDID (AccessGUDID)]
- Regulation number: 880.6310 [FDA GUDID (AccessGUDID)]; 870.2700 [FDA GUDID (AccessGUDID)]; 868.5895 [FDA GUDID (AccessGUDID)]
- GUDID distribution status: In Commercial Distribution [FDA GUDID (AccessGUDID)]

## Properties

- Prescription use: Yes [FDA GUDID (AccessGUDID)]
- Serial number: Yes [FDA GUDID (AccessGUDID)]
- Manufacturing date: Yes [FDA GUDID (AccessGUDID)]
- Supplied sterile: No [FDA GUDID (AccessGUDID)]
- Sterilise before use: No [FDA GUDID (AccessGUDID)]
- Single use: No [FDA GUDID (AccessGUDID)]
- Implantable: No [FDA GUDID (AccessGUDID)]
- Contains latex: No [FDA GUDID (AccessGUDID)]
- Over-the-counter: No [FDA GUDID (AccessGUDID)]
- Convenience kit: No [FDA GUDID (AccessGUDID)]
- Combination product: No [FDA GUDID (AccessGUDID)]
- HCT/P: No [FDA GUDID (AccessGUDID)]
- Premarket exempt: No [FDA GUDID (AccessGUDID)]
- Lot / batch number: No [FDA GUDID (AccessGUDID)]
- Expiration date: No [FDA GUDID (AccessGUDID)]

## Source and citation

- Cite as: Trilogy R1119867, Respironics Inc. deviCERTS medical device catalogue. https://devicerts.com/devices/trilogy-r1119867 (data as of 2026-11-17).
- Data sources: [FDA GUDID (AccessGUDID)](https://accessgudid.nlm.nih.gov/), [FDA recalls](https://open.fda.gov/apis/device/recall/)
- Registry records last synced: 2026-10-23
- Statuses computed as of: 2026-11-17
- JSON: <https://devicerts.com/devices/trilogy-r1119867.json>
- Licence: [CC BY 4.0](https://creativecommons.org/licenses/by/4.0/). The deviCERTS layout and text are licensed under CC BY 4.0. Registry data remain under their source terms: U.S. FDA data (openFDA, AccessGUDID) are U.S. public domain; EUDAMED data are subject to the European Commission reuse terms.
- Attribution: DevicErts, https://devicerts.com/devices/trilogy-r1119867
- Compiled by deviCERTS (devicerts.com) from public medical device registries (EUDAMED, European Commission; U.S. FDA via openFDA; AccessGUDID, U.S. National Library of Medicine). deviCERTS is not an official register and gives no regulatory advice: the record in the registry itself prevails.