# Trilogy System with Rapidarc

> Active: United States. Medical device record on deviCERTS, compiled from public registries.

- Canonical page: <https://devicerts.com/devices/trilogy-system-with-rapidarc-iye>
- Manufacturer: [Varian Medical Systems, Inc.](https://devicerts.com/manufacturers/varian-medical-systems-inc) (United States)
- Name in the registry: TRILOGY SYSTEM WITH RAPIDARC
- Device class: Class II (US FDA)
- FDA product code: IYE

## Registrations and authorizations

- **510(k)** · FDA 510(k) · United States · K072916 · decision 2007-11-09 · decision SESE · status: Active

## Recalls

- Z-0821-2011 ·  · terminated · initiated 2010-07-20: The system may deliver an inaccurate dose that does not match the treatment plan. As a result of a system configuration problem, the VMAT (Volumetric Modulated Arc Therapy) plans may be delivered as Dynamic Arc therapy plans. If this occurs, the VMAT plan will not be recognized. This issue affects Trilogy machines with the Rapid Arc Treatment delivery option.
- Z-1001-2011 ·  · terminated · initiated 2009-12-01: A circuit design anomaly whereby a faulty balance potentiometer can lead to an undetected electron beam asymmetry.

## Identifiers

- 510(k) Number: K072916 [FDA 510(k)]
- FDA product code: IYE [FDA 510(k)]

## Classification

- Device class: Class II [FDA 510(k)]
- Regulation: 21 CFR Part 892.5050 [FDA 510(k)]
- Advisory committee: Radiology [FDA 510(k)]
- FDA decision: SESE [FDA 510(k)]

## Source and citation

- Cite as: Trilogy System with Rapidarc, Varian Medical Systems, Inc. deviCERTS medical device catalogue. https://devicerts.com/devices/trilogy-system-with-rapidarc-iye (data as of 2026-11-22).
- Data sources: [FDA 510(k)](https://open.fda.gov/apis/device/510k/), [FDA recalls](https://open.fda.gov/apis/device/recall/)
- Registry records last synced: 2026-10-22
- Statuses computed as of: 2026-11-22
- JSON: <https://devicerts.com/devices/trilogy-system-with-rapidarc-iye.json>
- Licence: [CC BY 4.0](https://creativecommons.org/licenses/by/4.0/). The deviCERTS layout and text are licensed under CC BY 4.0. Registry data remain under their source terms: U.S. FDA data (openFDA, AccessGUDID) are U.S. public domain; EUDAMED data are subject to the European Commission reuse terms.
- Attribution: DevicErts, https://devicerts.com/devices/trilogy-system-with-rapidarc-iye
- Compiled by deviCERTS (devicerts.com) from public medical device registries (EUDAMED, European Commission; U.S. FDA via openFDA; AccessGUDID, U.S. National Library of Medicine). deviCERTS is not an official register and gives no regulatory advice: the record in the registry itself prevails.