Trinidad CH system, TD-LYTE Integrated Multisensor (Na, K, CI), TD-LYTE IMT Standard A, TD-LYTE IMT Standard B + Salt Bridge, Trinidad CH Albumin BCP reagent (Alb_P), Trinidad CH Albumin BCP Calibrator

by Siemens Healthcare Diagnostics, Inc.

510(k) K151767

Active United States

510(k)

Valid until

No expiry

510(k) clearance is not time-limited

FDA recalls

5 ongoing

This device is named in the recall

Documents

Trust

Unverified

Registry data, not yet checked by deviCERTS. Source FDA 510(k), synced 22 Oct 2026. What this means

Regulatory status

as of registry sync 22 Oct 2026

RegionAuthorizationNumberDateStatus
United StatesFDA 510(k)SESE Substantially Equivalent · Traditional NumberK151767 Date11 Dec 2015ClearedStatusActive

Recalls

12 recalls · 5 ongoing · FDA

    • Ongoing Z-2027-2025 FDA Class II initiated 22 May 2025 on shared 510(k) K151767

    Contamination of the diluent by sodium hypochlorite (NaOCl) which may impact calibrator, quality control (QC), and patient results for all assays except for sodium, potassium, and chloride.

    • Ongoing Z-1003-2024 FDA Class II initiated 15 Dec 2023 on shared 510(k) K151767

    Potential for a negative bias with quality control (QC) and patient sample results when using the Atellica CH Immunoglobulin M_2 (IgM_2) reagent.

    • Ongoing Z-0449-2022 FDA Class II initiated 18 Nov 2021

    (1)Software (SW) versions V1.25.1 and lower may result in TDef (Test Definition) Parameters for Open Channel Assays reverting to Default Value would shift test results by the magnitude of the implemented Assay Correlation factor; (2) On Board Stability (OBS) Not Updating with Manual Change for Open Channel Assays may continue to use the reagent past its expiration date leading to potential erroneous patient sample results

    • Ongoing Z-2026-2021 FDA Class II initiated 13 May 2021

    Valves used on the Atellica CH 930, Atellica IM 1300 and Atellica IM 1600 analyzers may develop a malfunction due to a manufacturing defect and may result in the valve wearing and leaking over time. The leaking is a risk in valves subject to high pressure or high wear and may impact the result accuracy of any assay.

    • Ongoing Z-1870-2021 FDA Class II initiated 4 May 2021

    Photometer Lamp May Reach Saturation Without Flagging Results, may generate erroneously elevated or depressed patient results

    • Terminated Z-0861-2021 FDA Class II initiated 2 Dec 2020

    Multiple issues in Software v1.23.2 and lower-may lead to the reporting of discrepant patient results

    • Terminated Z-1797-2020 FDA Class II initiated 27 Feb 2020

    Editing an assay Test Definition and switching to another assay s Test Definition screen without saving the changes using the save button on the screen, may corrupt the settings of the assay s test definition, and Auto-Rerun or Auto-Dilution results may generate a result of Zero (0) for quantitative assays or Negative for qualitative assays if a well of the reagent pack is not calibrated.

    • Terminated Z-0862-2020 FDA Class II initiated 11 Oct 2019

    On the Atellica CH 930 Analyzer, when scanning barcodes at the module console for Integrated Multisensor Technology (IMT) system fluids (Standard A (Std A) and Standard B (Std B)) for Sodium (Na), Potassium (K), and Chloride (Cl), the system will translate the decimal point separators to commas. This will result in invalid concentration values for the fluids, therefore the system will utilize default nominal values instead of lot specific IMT fluid concentration values for calculation of the calibrations and results for the Na, K, and Cl methods.

    • Terminated Z-0581-2020 FDA Class II initiated 23 Sep 2019 on shared 510(k) K151767

    Potential for calibration error resulting in QC failures after the calibration run with QC and patient results being 5 times higher.

    • Terminated Z-0148-2020 FDA Class II initiated 30 Jul 2019

    Atellica CH 930 Analyzer - Erroneous Indices for Hemolysis and Lipemia in Software versions 1.19.2 and below

    • Terminated Z-0782-2019 FDA Class II initiated 26 Sep 2018 on shared 510(k) K151767

    Multiple Issues Identified in Atellica Solution System Software in V 1.14.2 and lower in: Atellica IM 1300 Analyzer Atellica IM 1600 Analyzer;Atellica CH 930 Analyzer; Atellica Sample Handler Prime

    • Terminated Z-0194-2019 FDA Class II initiated 29 Jun 2018 on shared 510(k) K151767

    Atellica Solution system software (SW) version 1.13 and lower, five (5) behavioral issues were identified: 1. Reagent Lock may not work as intended with Primary Reagent ReadyPacks and Ancillary Packs when more than one (1) lot of specific Reagent kits are loaded on the analyzer at the same time. 2. Sample barcodes may be incorrectly read 3. The Atellica IM 1300 Analyzer and Atellica IM 1600 Analyzer may not run CEA dilutions at 50x and 100x. The Analyzer may post a Sample Integrity Error when 50x or 100x dilutions are ordered for Carcinoembryonic Antigen (CEA). Dilution factors lower than 50x and neat samples are not affected by this behavior. 4. The Atellica CH 930 Analyzer may continue to display the default reporting units after the reporting units have been changed. 5. The Atellica CH 930 Analyzer may not display results for sample if certain assays are configured to auto-dilute results above or below the assay measuring interval.

Recalls on a shared submission name another device cleared under the same FDA submission as this one.

Coverage

1 jurisdiction

United States of America
  • Active
  • Expiring
  • Withdrawn
  • Listed
  • Not covered

United States — FDA 510(k) Active

Specification

510(k)

510(k) Number
FDA Product Code
JGS
Device Class
Class II
Regulation
21 CFR Part 862.1665
Advisory Committee
Clinical Chemistry
FDA Decision
SESE Substantially Equivalent

Other devices from Siemens Healthcare Diagnostics

See all devices from Siemens Healthcare Diagnostics →