TruDi NAV Cable TDNC001Z
TruDi NAV Cable TDNC001Z is a medical device (FDA Class II; FDA product code PGW, Ear, Nose, and Throat Stereotaxic Instrument) made by Acclarent, Inc., cleared under FDA 510(k) K180948 on 20 Jul 2018 and listed in FDA GUDID since 31 Oct 2018 (UDI-DI 20705031245867), active as of 23 Oct 2026. 5 FDA recalls are linked to it.
Active
Valid until
No expiry
FDA recalls
2 ongoingDocuments
Trust
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | GUDID listing | Number20705031245867 | Date31 Oct 2018 | StatusListed |
| United States | FDA 510(k)SESE | NumberK180948 | Date20 Jul 2018 | StatusActive |
Official records
- AccessGUDID record 20705031245867
- FDA 510(k) database K180948
From the FDA record
- Device description
- TruDi NAV Cable - 1PK
- MRI safety
- MR Unsafe
- Natural rubber latex
- No latex statement on the label
- Sterility
- Supplied sterile
Across registries
- FDA 510(k) K180948 decided 20 Jul 2018 Active
- FDA GUDID UDI-DI 20705031245867 listed 31 Oct 2018 Listed
Recalls
-
- Ongoing Z-2507-2024
Specific lots of TruDi NAV Suction Instruments 0 Degrees were incorrectly calibrated which when used with the TruDi Navigation System, may result in a discrepancy between the actual tip location and the visualized location displayed on the system.
-
- Ongoing Z-0415-2022
Suction Instruments were incorrectly calibrated so they may not meet the system accuracy specification for visual verification of device location in the patient anatomy. If the surgeon does not detect the accuracy error, this could lead to an intraprocedural delay, a potential for reversible tissue damage or damage to the carotid artery, CSF leak, or damage to optic nerve.
-
- Terminated Z-0824-2020
- Terminated Z-0825-2020
- Terminated Z-0826-2020
The firm became aware that three (3) lots of the NAV Suction Instruments contain calibration issues causing incorrect tip orientation display in the system.
Timeline
- Cleared FDA 510(k) K180948
- Listed FDA GUDID 20705031245867
- Recall initiated FDA recalls Z-0824-2020
- Recall initiated FDA recalls Z-0825-2020
- Recall initiated FDA recalls Z-0826-2020
- Recall initiated FDA recalls Z-0415-2022
- Recall initiated FDA recalls Z-2507-2024
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- Trade name
- TruDi NAV Cable
- Generic name
- Electromagnetic/optical surgical navigation/tracking system
- Model / Ref / Catalog
- TDNC001Z
- Primary DI / UDI-DI
- 20705031245867
- 510(k) Number
- FDA product code
- PGW
- GMDN term
- Electromagnetic/optical surgical navigation/tracking system
- Device class
- Class II
- Regulation number
- 882.4560
- GUDID distribution status
- In Commercial Distribution
- MRI safety
- MR Unsafe
Similar devices
Other devices filed under FDA product code PGW (Ear, Nose, and Throat Stereotaxic Instrument).
- Fusion Compact Navigation System by Medtronic Navigation Inc.
- FlexPointer 1.5 mm by Fiagon GmbH
- PointerShell Universal, PointerShell LS by Fiagon GmbH
- TGS Universal Headrest with Mounting Arm by Stryker Ent
- StealthStationTM S8 ENT Software by Medtronic Navigation Inc.
- Scopis Extended Instrument Set EM by Scopis GmbH
- ENT EM by Brainlab AG
- N / A 8000060020 8000-060-020 by Stryker Leibinger
- Instrument Clamps Electromagnetic by Stryker Corp.
- PROFESS SOFTWARE, PROFESS CURVED 70 (DEGREES) SUCTION, PROFESS CURVED 70 (DEGREES) SUCTION, PROFESS CURVED 90 (DEGREES) by Stryker Corporate