TRUE METRIX AIR Self-Monitoring Blood Glucose System, TRUE METRIX AIR PRO Professional Blood Gulcose Monitoring System
510(k) K150052
Active
Valid until
No expiry
FDA recalls
1 ongoingDocuments
Trust
Unverified
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK150052 | Date26 Mar 2015 | StatusActive |
Recalls
-
- Ongoing Z-1547-2026
The system labeling (user manual and online labeling) did not provide adequate directions for lay users to act on the E-5 error code. The error code description and recommended actions for the E-5 error code in the labeling do not: (a) clearly emphasize that an E-5 error code could represent a very high blood glucose level, and (b) prominently convey the appropriate immediate actions, such as contacting a healthcare professional. This could lead to a delay in treatment if the user does not seek medical attention immediately when receiving an E-5 error code and experiencing symptoms of high glucose.
-
- Terminated Z-2201-2020
One (1) TRUE METRIX AIR blood glucose meter distributed in the United States was packaged into a TRUE METRIX blood glucose meter kit and has an incorrect factory-set unit of measure; the meter displays glucose results in mmol/L rather than mg/dL.
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA Product Code
- NBW
- Device Class
- Class II
- Regulation
- 21 CFR Part 862.1345
- Advisory Committee
- Clinical Chemistry
- FDA Decision
- SESE Substantially Equivalent