  "format": "devicerts.device.v1",
  "cite_as": "88000057, Shanghai United Imaging Healthcare Co., Ltd. deviCERTS medical device catalogue. https://devicerts.com/devices/umi-550-88000057 (data as of 2026-11-27).",
  "source": {
    "canonical_url": "https://devicerts.com/devices/umi-550-88000057",
    "json_url": "https://devicerts.com/devices/umi-550-88000057.json",
    "markdown_url": "https://devicerts.com/devices/umi-550-88000057.md",
    "publisher": "deviCERTS (devicerts.com)",
    "data_sources": [
      {
        "registry": "EUDAMED",
        "url": "https://ec.europa.eu/tools/eudamed"
      },
      {
        "registry": "FDA GUDID (AccessGUDID)",
        "url": "https://accessgudid.nlm.nih.gov/"
      },
      {
        "registry": "FDA recalls",
        "url": "https://open.fda.gov/apis/device/recall/"
      }
    ],
    "registry_records_last_synced": "2026-10-22",
    "status_as_of": "2026-11-27",
    "license": "CC BY 4.0",
    "license_url": "https://creativecommons.org/licenses/by/4.0/",
    "license_scope": "The deviCERTS layout and text are licensed under CC BY 4.0. Registry data remain under their source terms: U.S. FDA data (openFDA, AccessGUDID) are U.S. public domain; EUDAMED data are subject to the European Commission reuse terms.",
    "attribution": "DevicErts, https://devicerts.com/devices/umi-550-88000057",
    "disclaimer": "Compiled by deviCERTS (devicerts.com) from public medical device registries (EUDAMED, European Commission; U.S. FDA via openFDA; AccessGUDID, U.S. National Library of Medicine). deviCERTS is not an official register and gives no regulatory advice: the record in the registry itself prevails."
  },
  "device": {
    "name": "88000057",
    "registry_name": "88000057",
    "url": "https://devicerts.com/devices/umi-550-88000057",
    "manufacturer": {
      "name": "Shanghai United Imaging Healthcare Co., Ltd.",
      "url": "https://devicerts.com/manufacturers/shanghai-united-imaging-healthcare-co-ltd",
      "country": "China",
      "country_basis": null,
      "fda_registration_country": null
    },
    "status": {
      "state": "active",
      "label": "Active",
      "detail": "European Union and United States"
    },
    "classification": [
      {
        "class": "Class IIb",
        "system": "EU MDR / EU MDD"
      },
      {
        "class": "Class II",
        "system": "US FDA"
      }
    ],
    "identifiers": [
      {
        "label": "Primary DI / UDI-DI",
        "values": [
          {
            "value": "06971576832026",
            "registries": [
              "EUDAMED",
              "FDA GUDID (AccessGUDID)"
            ]
          }
        ]
      },
      {
        "label": "Basic UDI-DI",
        "values": [
          {
            "value": "B-06971576832026",
            "registries": [
              "EUDAMED"
            ]
          },
          {
            "value": "697157683202N2",
            "registries": [
              "EUDAMED"
            ]
          }
        ]
      },
      {
        "label": "510(k) Number",
        "values": [
          {
            "value": "K182237",
            "registries": [
              "FDA GUDID (AccessGUDID)"
            ]
          },
          {
            "value": "K193210",
            "registries": [
              "FDA GUDID (AccessGUDID)"
            ]
          },
          {
            "value": "K193241",
            "registries": [
              "FDA GUDID (AccessGUDID)"
            ]
          },
          {
            "value": "K210001",
            "registries": [
              "FDA GUDID (AccessGUDID)"
            ]
          },
          {
            "value": "K210418",
            "registries": [
              "FDA GUDID (AccessGUDID)"
            ]
          }
        ]
      }
    ],
    "product_codes": [
      "KPS",
      "JAK"
    ],
    "emdn": [
      {
        "code": "Z11020301",
        "term": "CT/PET SYSTEMS"
      }
    ],
    "gmdn": [
      {
        "code": "45143",
        "term": "PET/CT system"
      }
    ],
    "description": null,
    "noindex": false
  },
  "authorizations": [
    {
      "registry": "EUDAMED",
      "region": "European Union",
      "type": "CE/MDR",
      "number": "B-06971576832026",
      "number_label": "Basic UDI-DI",
      "device_udi_di": "06971576832026",
      "date": "2024-02-22",
      "date_kind": "registered",
      "decision_code": null,
      "expires": null,
      "withdrawn": null,
      "status": "Active",
      "status_note": null,
      "derived_from_gudid": false
    },
    {
      "registry": "EUDAMED",
      "region": "European Union",
      "type": "CE/MDD",
      "number": "B-06971576832026",
      "number_label": "Basic UDI-DI",
      "device_udi_di": "06971576832026",
      "date": "2024-02-22",
      "date_kind": "registered",
      "decision_code": null,
      "expires": null,
      "withdrawn": null,
      "status": "Active",
      "status_note": null,
      "derived_from_gudid": false
    },
    {
      "registry": "GUDID",
      "region": "United States",
      "type": "FDA 510(k)",
      "number": "K210001",
      "number_label": null,
      "device_udi_di": null,
      "date": "2021-04-30",
      "date_kind": "decision",
      "decision_code": "SESE",
      "expires": null,
      "withdrawn": null,
      "status": "Active",
      "status_note": null,
      "derived_from_gudid": true
    },
    {
      "registry": "GUDID",
      "region": "United States",
      "type": "FDA 510(k)",
      "number": "K210418",
      "number_label": null,
      "device_udi_di": null,
      "date": "2021-04-09",
      "date_kind": "decision",
      "decision_code": "SESE",
      "expires": null,
      "withdrawn": null,
      "status": "Active",
      "status_note": null,
      "derived_from_gudid": true
    },
    {
      "registry": "GUDID",
      "region": "United States",
      "type": "FDA 510(k)",
      "number": "K193210",
      "number_label": null,
      "device_udi_di": null,
      "date": "2020-08-04",
      "date_kind": "decision",
      "decision_code": "SESE",
      "expires": null,
      "withdrawn": null,
      "status": "Active",
      "status_note": null,
      "derived_from_gudid": true
    },
    {
      "registry": "GUDID",
      "region": "United States",
      "type": "FDA 510(k)",
      "number": "K193241",
      "number_label": null,
      "device_udi_di": null,
      "date": "2020-01-26",
      "date_kind": "issued",
      "decision_code": null,
      "expires": null,
      "withdrawn": null,
      "status": "Active",
      "status_note": null,
      "derived_from_gudid": true
    },
    {
      "registry": "GUDID",
      "region": "United States",
      "type": "GUDID listing",
      "number": "06971576832026",
      "number_label": "UDI-DI",
      "device_udi_di": null,
      "date": "2019-04-03",
      "date_kind": "listed",
      "decision_code": null,
      "expires": null,
      "withdrawn": null,
      "status": "Listed",
      "status_note": "GUDID: in commercial distribution",
      "derived_from_gudid": false
    },
    {
      "registry": "GUDID",
      "region": "United States",
      "type": "FDA 510(k)",
      "number": "K182237",
      "number_label": null,
      "device_udi_di": null,
      "date": "2018-10-12",
      "date_kind": "decision",
      "decision_code": "SESE",
      "expires": null,
      "withdrawn": null,
      "status": "Active",
      "status_note": null,
      "derived_from_gudid": true
    }
  ],
  "no_longer_listed_by": [],
  "certificates": [],
  "recalls": [
    {
      "number": "Z-3283-2026",
      "classification": "Class II",
      "status": "ongoing",
      "initiated": "2026-08-11",
      "reason": "Positron emission tomography and computed tomography x-ray systems have a software problem when Split or Local function is enabled, and multiple RIS exams are merged into a single study, that prevents patient images from being visible on the Patient Administration interface, which may result in the need for rescans.",
      "applies": "this device",
      "via": "K182237"
    },
    {
      "number": "Z-0223-2026",
      "classification": "Class II",
      "status": "ongoing",
      "initiated": "2025-09-12",
      "reason": "Third-party outer covers can become trapped between patient table and gantry, causing mechanical interference during table movement.",
      "applies": "this device",
      "via": "K182237"
    },
    {
      "number": "Z-1353-2023",
      "classification": "Class II",
      "status": "ongoing",
      "initiated": "2023-03-01",
      "reason": "Due to a software issue where the process of patient scanning, the scatter correction may occasionally fail with will potentially cause a failure of the PET image reconstruction and generation of PET images.",
      "applies": "this device",
      "via": "K182237"
    },
    {
      "number": "Z-0087-2023",
      "classification": "Class II",
      "status": "ongoing",
      "initiated": "2022-05-30",
      "reason": "The wireless VSM module of a mobile PET/CT system, operating in an environment with strong Wi-Fi signals, may experience ECG signal and respiratory signal loss due to Wi-Fi interference. ECG and respiratory signal loss during acquisition can result in the failure of ECG and respiratory-gated reconstruction of the PET scan, which may require rescanning of the patient.",
      "applies": "this device",
      "via": "K182237"
    },
    {
      "number": "Z-2593-2021",
      "classification": "Class II",
      "status": "ongoing",
      "initiated": "2021-08-02",
      "reason": "The EXAM acquisition workflow could occasionally freeze after canceling the PET scan.",
      "applies": "this device",
      "via": "K182237"
    },
    {
      "number": "Z-0448-2021",
      "classification": "Class II",
      "status": "ongoing",
      "initiated": "2020-08-05",
      "reason": "A potential issue of an incorrect CT scan delay timer to be displayed on the gantry Digital Display Panel (DDP) during\nmulti-phase contrast scanning. This could lead the operator to receive unnecessary exposure to radiation.",
      "applies": "this device",
      "via": "K182237"
    },
    {
      "number": "Z-2819-2020",
      "classification": "Class II",
      "status": "terminated",
      "initiated": "2020-03-26",
      "reason": "In affected software version, of Positron emission tomography and computed tomography system, the CT Hounsfield unit curve may not display in real time when using the bolus tracking protocol with the tracker scan function. If this occurs, the subsequent clinical scanning protocols will need to be started manually, which may effect image enhancement, and may result in patient rescan.",
      "applies": "this device",
      "via": "K182237"
    },
    {
      "number": "Z-1708-2020",
      "classification": "Class II",
      "status": "terminated",
      "initiated": "2019-08-26",
      "reason": "Potential sporadic software bugs in R001.3.0.0.750505 software version may cause an issue with the PET acquisition raw data causing reconstruction problems during PET/CT scanning and could possibly cause rescans of patients with additional dose.",
      "applies": "this device",
      "via": "K182237"
    },
    {
      "number": "Z-2594-2021",
      "classification": "Class II",
      "status": "ongoing",
      "initiated": "2021-08-02",
      "reason": "The EXAM acquisition workflow could occasionally freeze after canceling the PET scan.",
      "applies": "this device",
      "via": "K182237"
    },
    {
      "number": "Z-2800-2020",
      "classification": "Class II",
      "status": "terminated",
      "initiated": "2020-07-16",
      "reason": "1) Service function problem may cause false marking of a bad channel resulting in ring artifacts; 2) Potential intermittent scout scanning interruption due to angle signal drift caused by external interference; 3) Metal edge overlap of mylar strip in gantry may create sharp edge. May lead to repeat scanning and abrasions (N/A for Serial No. 200036)",
      "applies": "this device",
      "via": "K182237"
    }
  ],
  "eu_market_availability": null,
  "properties": [
    {
      "group": "Identity",
      "fields": [
        {
          "label": "Trade name",
          "values": [
            {
              "value": "uMI",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "Generic name",
          "values": [
            {
              "value": "PET/CT system",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "Model / reference",
          "values": [
            {
              "value": "uMI 550",
              "registries": [
                "EUDAMED",
                "FDA GUDID (AccessGUDID)"
              ]
            },
            {
              "value": "88000057",
              "registries": [
                "EUDAMED"
              ]
            }
          ]
        }
      ]
    },
    {
      "group": "Identifiers",
      "fields": [
        {
          "label": "Primary DI / UDI-DI",
          "values": [
            {
              "value": "06971576832026",
              "registries": [
                "EUDAMED",
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "Basic UDI-DI",
          "values": [
            {
              "value": "B-06971576832026",
              "registries": [
                "EUDAMED"
              ]
            },
            {
              "value": "697157683202N2",
              "registries": [
                "EUDAMED"
              ]
            }
          ]
        },
        {
          "label": "510(k) Number",
          "values": [
            {
              "value": "K182237",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            },
            {
              "value": "K193210",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            },
            {
              "value": "K193241",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            },
            {
              "value": "K210001",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            },
            {
              "value": "K210418",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "FDA product code",
          "values": [
            {
              "value": "KPS",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            },
            {
              "value": "JAK",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "EMDN code",
          "values": [
            {
              "value": "Z11020301",
              "registries": [
                "EUDAMED"
              ]
            }
          ]
        },
        {
          "label": "GMDN term",
          "values": [
            {
              "value": "PET/CT system",
              "code": "45143",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        }
      ]
    },
    {
      "group": "Classification",
      "fields": [
        {
          "label": "Device class",
          "values": [
            {
              "value": "Class IIb",
              "registries": [
                "EUDAMED"
              ]
            },
            {
              "value": "Class II",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "Regulation number",
          "values": [
            {
              "value": "892.1200",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            },
            {
              "value": "892.1750",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "EU registration type",
          "values": [
            {
              "value": "Legacy device (under a directive)",
              "registries": [
                "EUDAMED"
              ]
            },
            {
              "value": "Standard device",
              "registries": [
                "EUDAMED"
              ]
            }
          ]
        },
        {
          "label": "EU market of registration",
          "values": [
            {
              "value": "Poland",
              "registries": [
                "EUDAMED"
              ]
            }
          ]
        },
        {
          "label": "GUDID distribution status",
          "values": [
            {
              "value": "In Commercial Distribution",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        }
      ]
    },
    {
      "group": "Properties",
      "fields": [
        {
          "label": "Prescription use",
          "values": [
            {
              "value": "Yes",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "Serial number",
          "values": [
            {
              "value": "Yes",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "Manufacturing date",
          "values": [
            {
              "value": "Yes",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "Supplied sterile",
          "values": [
            {
              "value": "No",
              "registries": [
                "EUDAMED",
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "Sterilise before use",
          "values": [
            {
              "value": "No",
              "registries": [
                "EUDAMED",
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "Single use",
          "values": [
            {
              "value": "No",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "Implantable",
          "values": [
            {
              "value": "No",
              "registries": [
                "EUDAMED",
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "Reusable",
          "values": [
            {
              "value": "No",
              "registries": [
                "EUDAMED"
              ]
            }
          ]
        },
        {
          "label": "Contains latex",
          "values": [
            {
              "value": "No",
              "registries": [
                "EUDAMED",
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "Reprocessed single-use device",
          "values": [
            {
              "value": "No",
              "registries": [
                "EUDAMED"
              ]
            }
          ]
        },
        {
          "label": "Contains CMR substances (carcinogenic, mutagenic or toxic to reproduction)",
          "values": [
            {
              "value": "No",
              "registries": [
                "EUDAMED"
              ]
            }
          ]
        },
        {
          "label": "Incorporates a substance derived from human blood or plasma",
          "values": [
            {
              "value": "No",
              "registries": [
                "EUDAMED"
              ]
            }
          ]
        },
        {
          "label": "Contains human tissues or cells",
          "values": [
            {
              "value": "No",
              "registries": [
                "EUDAMED"
              ]
            }
          ]
        },
        {
          "label": "Contains animal tissues or cells",
          "values": [
            {
              "value": "No",
              "registries": [
                "EUDAMED"
              ]
            }
          ]
        },
        {
          "label": "Contains endocrine-disrupting substances",
          "values": [
            {
              "value": "No",
              "registries": [
                "EUDAMED"
              ]
            }
          ]
        },
        {
          "label": "Administering medicine",
          "values": [
            {
              "value": "No",
              "registries": [
                "EUDAMED"
              ]
            }
          ]
        },
        {
          "label": "Measuring function",
          "values": [
            {
              "value": "No",
              "registries": [
                "EUDAMED"
              ]
            }
          ]
        },
        {
          "label": "Incorporates a medicinal substance",
          "values": [
            {
              "value": "No",
              "registries": [
                "EUDAMED"
              ]
            }
          ]
        },
        {
          "label": "Over-the-counter",
          "values": [
            {
              "value": "No",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "Convenience kit",
          "values": [
            {
              "value": "No",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "Combination product",
          "values": [
            {
              "value": "No",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "HCT/P",
          "values": [
            {
              "value": "No",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "Premarket exempt",
          "values": [
            {
              "value": "No",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "Lot / batch number",
          "values": [
            {
              "value": "No",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "Expiration date",
          "values": [
            {
              "value": "No",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        }
      ]
    }
  ]
}