RegionAuthorizationNumberDateStatus
United StatesFDA 510(k)SESE Substantially Equivalent · Traditional NumberK111473 Date16 Feb 2012ClearedStatusActive

Recalls

2 recalls · 1 ongoing · FDA

    • Ongoing Z-2130-2024 FDA Class I initiated 30 Apr 2024

    Operator's Guide & Quick Reference Guide (QRG) ZOLL 731 Ventilator for MRI Compatible Devices. Patient Safety Ventilator MRI information omitted from the manuals. Update to the instructions ensures safe distance from the MRI to ensure proper function of the ventilator and reduce the risk of delay in ventilator therapy

    • Terminated Z-0812-2018 FDA Class II initiated 30 Jun 2017

    A software anomaly in the 731 software version 05/20/00, was identified, which can lead to a user inadvertently changing device settings.

Coverage

1 jurisdiction

United States of America
  • Active
  • Expiring
  • Withdrawn
  • Listed
  • Not covered

United States — FDA 510(k) Active

Specification

510(k)

510(k) Number
FDA Product Code
CBK
Device Class
Class II
Regulation
21 CFR Part 868.5895
Advisory Committee
Anesthesiology
FDA Decision
SESE Substantially Equivalent

Other devices from Impact Instrumentation, Inc.