UniCel DxH 800 Coulter Cellular Analysis System
Ref B24802
Active
Valid until
Not listed
FDA recalls
2 ongoingTrust
Unverified
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| European Union | CE/IVDD EUDAMED | NumberB-B24802UJ D-B24802UJ | Date7 Jun 2022 | StatusActive |
| United States | FDA 510(k) 510(k) | NumberK193124 | Date16 Apr 2020 | StatusActive |
| United States | GUDID listing GUDID | NumberD-B24802UJ | Date23 Sep 2016 | StatusListed |
Certificates
EUDAMED lists no certificate for this device's Basic UDI-DI.
Recalls
-
- Ongoing Z-2574-2025
- Ongoing Z-2575-2025
Hematology analyzers configured with HGB photometers may result in erroneously high hemoglobin (HGB) results on samples with elevated white blood cell (WBC) counts.
-
- Terminated Z-0516-2020
When the UniCel DxH Slidemaker Stainer and UniCel DxH Slidemaker Stainer II dispense probe is slightly bent, there is a potential for cell carryover from the previous sample, and erroneous results could be reported from the slide review. Erroneous results may be reported from the laboratory which may trigger unnecessary or inappropriate patient treatment.
-
- Terminated Z-1383-2019
- Terminated Z-1384-2019
- Terminated Z-1382-2019
Complaints received of sporadic erroneously elevated platelet results without flags or system messages. Thrombocytopenia may go unrecognized.
-
- Terminated Z-0236-2018
- Terminated Z-0238-2018
- Terminated Z-0237-2018
The shield for the Single-Tube Station on the DxH instrument may interfere with sample aspiration in the open vial position.
-
- Terminated Z-3101-2017
The firm has identified that a fitting in some Slidemaker Stainer units may not completely seal after installation.
-
- Terminated Z-2158-2017
- Terminated Z-2157-2017
There is a possibility of a data acquisition disruption that may cause an unusual light scatter event pattern for the white blood cell differential that may not have a system flag and/or message. The events associated with the unusual light scatter pattern may be incorrectly removed from analysis which can result in an erroneous differential result.
-
- Terminated Z-1013-2017
- Terminated Z-1014-2017
Beckman Coulter has determined that additional clarification for the Blast Suspect messages is necessary. In rare situations, the UniCel DxH 800 and DxH 600 Coulter Cellular Analysis System may not flag or detect blasts in some blood samples. This is due to limitations in the available technology as well as sample limitations.
-
- Terminated Z-2732-2016
There is a possibility of specimen misidentification for the stand-alone DxH 800 and DxH 600 where the Primary Identifier is configured to Tube Position ID.
-
- Terminated Z-2692-2016
Beckman Coulter is recalling the UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System to reduce the potential for flammable liquids coming into contact with electrical components in the Stainer module possibly causing a fire.
-
- Terminated Z-0432-2016
- Terminated Z-0433-2016
- Terminated Z-0434-2016
Beckman Coulter is recalling the DxH 800 Slidemaker Stainer Software version 3.0.2.0 and DxH 600 software version 1.1.1.0 because the software for the DxH systems allows the creation of multiple orders with the same Specimen identification (ID) but different Patient identification when manually editing pending orders at the System Manager.
-
- Terminated Z-1616-2015
Beckman Coulter has received an increased number of complaints related to Hemoglobin (HGB) on UniCel DxH 800 and DxH 600 Coulter Cellular Analysis Systems. The use of the DxH Diluent lots listed in the recall notice may eventually result in a compromised HGB chamber.
-
- Terminated Z-1829-2014
- Terminated Z-1828-2014
Beckman Coulter is recalling the UniCel DxH800 and UniCel DxH600 Coulter Analysis Systems because the optical degradation may potentially cause a delay in reporting results for nucleated red blood cells (nRBC), automated white blood cell differential and the reticulocyte.
-
- Terminated Z-0147-2014
Beckman Coulter is initiating a recall for UniCel DxH 800 Coulter Cellular Analysis System (PN 629029) and UniCel DxH 600 Coulter Cellular Analysis System (PN B23858) because the company's internal investigation indicates that specific lots of check valves used in several locations within the DxH 800 and DxH 600 systems may fail. A failure may result in an air or liquid leak at the VCS Waste Chamb
-
- Terminated Z-0021-2013
Beckman Coulter is recalling the UniCel DxH 800 Coulter Cellular Analysis System because the NRBC, Differential or Reticulocyte mix chamber drain ports may become plugged by rubber pieces from specimen tube caps, and result in biohazard or chemical exposure and/or the potential for generating incorrect results.
-
- Terminated Z-3125-2011
The recall was initiated because Beckman Coulter has confirmed that the probe wash collar tubing may become stretched or trapped at the Sample Aspiration Module (SAM) during routine use of single-tube presentation station. When the tension on the tubing is severe, the tubing may not be able to return to its home state and may pull the wash collar out of alignment with the aspirate probe. The misal
-
- Terminated Z-2227-2011
The recall was initiated because Beckman Coulter has confirmed the auto prune will not remove patient results if the patient record includes demographics where the results are less than 30 days old.
-
- Terminated Z-2172-2011
The recall was initiated because: Specimen tubes may fall out of a Type A cassette (PN A53551, kit of 20) when the analyzer inverts the cassette in the mixing station, or when an operator manually inverts a cassette The impact: - Specimen tubes may fall out of cassettes and be misplaced inside the instrument - Specimen tubes may break upon impact, causing possible biohazard exposure - Patien
-
- Terminated Z-2145-2011
The recall was initiated because the DxH 800 may fail to meet the carryover specification for the WBC Differential (DIFF) when the WBC is greater than 70 X 10 to the 3rd power cells/micro L (70 X 10 to the 9th power cells/L). Beckman Coulter indicated the DIFF% carryover is unlikely to have clinical impact.
-
- Terminated Z-2148-2011
The recall was initiated because: The DxH 800 may fail to meet the carryover specification for NRBC% when the WBC is greater than 50 X 10 to the 3rd power cells/micro L (50 X 10 to the 9th power cells/L) Beckman Coulter indicated the NRBC% carryover is unlikely to have clinical impact.
-
- Terminated Z-2157-2011
The recall was initiated because: The handheld barcode scanner does not read barcode labels using NW 7 symbology. Impact: Specimens with NW7 labels cannot be read by the handheld scanner.
-
- Terminated Z-1980-2011
Sharp edges on drawer slides within the DxH800 Floor Stand (Cabinet) may expose operators to physical injury while replacing reagents and waste containers. IMPACT: There is potential for exposure to blood borne pathogens such as HIV, HBV, and HCV.
-
- Terminated Z-2114-2011
For Panels containing CBC: The RBC and HCT are improperly corrected and not flagged when there is a WBC Pump Failure and the WBC ~ 140x103/uL. For Body Fluid Panels: The RBC is improperly corrected and not flagged when there is a WBC Pump Failure. IMPACT: Erroneous RBC and/or HCT could affect the interpretation and management of patients at risk with various types of anemias.
-
- Terminated Z-2128-2011
Body Fluids results may be incorrectly interpreted by the Laboratory Information System (LIS). Body Fluid results ~ 1,000 celis/uL are reported using the US numeric separators to demark thousandths places (comma, ",") and partial units (decimal, "."). An LIS that uses non-US formats could interpret the comma as a partial units separator, which could lead to reporting erroneously low body fluid r
-
- Terminated Z-1977-2011
- Terminated Z-1978-2011
There is a possibility that a specimen could be diluted if a user requests Diluent Dispense while specimens are being automatically processed in cassette presentation. The dilution may cause erroneous results. IMPACT: Erroneous results from a diluted specimen could impact clinical decision making in cases such as blood transfusion, chemotherapy or antibiotic administration- particularly for popu
-
- Terminated Z-2121-2011
RETIC results may be invalid when the instrument user interface allows reflex processing of a pre-diluted specimen in the RETIC test panel. Beckman Coulter has validated pre-diluted specimen processing in CBC panels. RETIC test panel using pre-diluted samples has not been validated. IMPACT: The reticulocyte value is an indicator for assessing red cell production. Erroneous results could impact c
-
- Terminated Z-1656-2010
Issue 1: The Diagnostic procedure, Simulate-CDNR Cycle, is not functioning properly. Impact: There is a potential for inadequate sample preparation of samples following the use of this Diagnostic function impacting WBC, Differential, Reticulocyte and/or NRBC results. Issue 2: Use of the Auto Prune function causes the deletion of Quality Control files. Impact: If the last run for a control lot
-
- Terminated Z-2338-2010
The recall was initiated after Beckman Coulter identified four (4) issues with the COULTER LH 500 Series System and the Unicel DxH 800 Coulter Cellular Analysis System: Issue 1: There is a potential for misidentification to occur when the system is configured in languages other than English or Chinese. The characters #,@,[,\,],`,{,|,}, or ~ are substituted or omitted in other languages. The issu
-
- Terminated Z-1658-2010
Issue 1: The system will substitute or omit the characters # @ 1 \ ] ' ( I ) - when the instrument is configured in a language other than English, Chinese or Japanese. Example: Using the character @, a substitution in the string '123@456' would result in '123#456' and an omission would result in '123456'. Issue2: The characters 7 " and * (question mark, quotation mark, and asterisk) used as part
-
- Terminated Z-1680-2010
The Corrective Action was initiated after Beckman Coulter confirmed three issues related to host transmissions with the UniCel DxH 800. This information only affects instruments interfaced to a host system.
-
- Terminated Z-1905-2010
The recall was initiated after Beckman Coulter confirmed an issue with the UniCel DxH 800 during in-house testing. It was determined that there is a potential for a false increase in platelet results, due to micro-bubbles. This situation could present a risk of falsely increased platelet results in patient samples as diluent containers are depleted.
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
- United States
- European Union
Specification
- Trade Name
- UniCel DxH 800 Coulter Cellular Analysis System EUDAMEDUniCel DxH GUDID
- Generic Name
- Haematological cell analyser IVD
- Model / Ref / Catalog
- B24802, 800
- Primary DI / UDI-DI
- D-B24802UJ EUDAMED15099590672423 GUDID
- Basic UDI-DI
- B-B24802UJ
- FDA Product Code
- GKZ
- EMDN Code
- W0202010101 AUTOMATED WHOLE BLOOD HEMATOLOGY ANALYSERS - LOW ROUTINE (Throughput <= 100 samples/hr)
- GMDN Term
- Haematological cell analyser IVD
- Device Class
- Class IVD General EUDAMEDClass II GUDID
- Regulation Number
- 864.5220
- EU market of registration
- Ireland
- GUDID distribution status
- In Commercial Distribution
Truth per registry
| Field | EUDAMED | GUDID |
|---|---|---|
| Model / Reference | B24802 | 800 |
| Primary DI / UDI-DI | D-B24802UJ | 15099590672423 |
| Device Class | Class IVD General | Class II |
Manufacturer details
- EUDAMED SRN
- US-MF-000010288
- Authorised representative
- Beckman Coulter Ireland Inc. IE-AR-000000886
Other devices from Beckman Coulter, Inc.
Properties
Other properties (18)
- Sterile: No
- Single Use: No
- Implantable: No
- Contains Latex: No
- Human Tissues: No
- Animal Tissues: No
- Microbial Substances: No
- Reagent: No
- Self Testing: No
- Near-Patient Testing: No
- Companion Diagnostics: No
- Over-the-Counter: No
- Convenience Kit: No
- Combination Product: No
- HCT/P: No
- Premarket Exempt: No
- Lot / Batch Number: No
- Expiration Date: No