Unidos Universal Dosimeter Model 10002-With Tnc Connector, 10005-With Bnt Connector
510(k) K951764
Active
510(k)
Valid until
No expiry
FDA recalls
None ongoingDocuments
Trust
Unverified
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK951764 | Date7 Oct 1995 | StatusActive |
Recalls
-
- Terminated Z-2439-2015
In some cases, especially with high dose-per-pulse-rates typical for particle (specifically proton) therapy, measuring errors can result. Individual measurement intervals can be lost that are not visible to the user when there is a high dose-per-pulse rate resulting in a too-low dose being displayed on the monitor. This could result in a radiation over-exposure.
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA Product Code
- IYE
- Device Class
- Class II
- Regulation
- 21 CFR Part 892.5050
- Advisory Committee
- Radiology
- FDA Decision
- SESE Substantially Equivalent