USA Elite 190-NRS
USA Elite 190-NRS is a bladder evacuator (FDA Class II; FDA product code HIH, Hysteroscope (And Accessories)) made by Gyrus Acmi, Inc., cleared under FDA 510(k) K890328 on 6 Feb 1990 and listed in FDA GUDID since 11 Oct 2016 (UDI-DI 00821925003521), active as of 23 Oct 2026. 4 FDA recalls are linked to it.
Active
Valid until
No expiry
FDA recalls
6 ongoingDocuments
Trust
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | GUDID listing | Number00821925003521 | Date11 Oct 2016 | StatusListed |
| United States | FDA 510(k)SESE | NumberK890328 | Date6 Feb 1990 | StatusActive |
Official records
- AccessGUDID record 00821925003521
- FDA 510(k) database K890328
From the FDA record
- Device description
- USA Elite System and USA Series Ellik Evacuator Adapter, Locking
- MRI safety
- Labeling does not contain MRI Safety Information
- Natural rubber latex
- No latex statement on the label
- Sterility
- Supplied non-sterile; sterilise before use (Moist Heat or Steam Sterilization)
Across registries
- FDA 510(k) K890328 decided 6 Feb 1990 Active
- FDA GUDID UDI-DI 00821925003521 listed 11 Oct 2016 Listed
Recalls
-
- Ongoing Z-3300-2026
- Ongoing Z-3299-2026
There are identified discrepancies between the sterilization parameters listed in the Instructions for Use and the sterilization parameters listed in the Japanese package insert.
-
- Ongoing Z-2720-2026
Some sterilization parameters included in the IFU may not consistently render the devices adequately sterile and therefore should no longer be used to sterilize the devices.
-
- Ongoing Z-2140-2025
- Ongoing Z-2139-2025
Discrepancies between sterilization instructions in the Japanese language product package inserts and those published in the manufacturer product Instructions For Use (IFU).
-
- Ongoing Z-0568-2024
Incorrect pouch labels, the contents within the mislabeled pouch were Roller Bar electrodes.
-
- Terminated Z-1952-2018
There is a potential for the cord to spark and cause a fire.
Timeline
- Cleared FDA 510(k) K890328
- Listed FDA GUDID 00821925003521
- Recall initiated FDA recalls Z-1952-2018
- Recall initiated FDA recalls Z-0568-2024
- Recall initiated FDA recalls Z-2140-2025
- Recall initiated FDA recalls Z-2139-2025
- Recall initiated FDA recalls Z-2720-2026
- Recall initiated FDA recalls Z-3300-2026
- Recall initiated FDA recalls Z-3299-2026
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- Trade name
- USA Elite
- Generic name
- Bladder evacuator, reusable
- Model / reference
- 190-NRS
- Primary DI / UDI-DI
- 00821925003521
- 510(k) Number
- FDA product code
- HIH
- GMDN term
- Bladder evacuator, reusable
- Device class
- Class II
- Regulation number
- 884.1690
- GUDID distribution status
- In Commercial Distribution
- Sterilisation method
- Moist Heat or Steam Sterilization
Similar devices
Other devices filed under FDA product code HIH (Hysteroscope (And Accessories)).
- PANOVIEW PLUS 8986401 8986.401 by Richard Wolf GmbH
- Olympus WA22639S by Olympus Winter & Ibe GmbH
- MyoSure XL Rod Lens Hysteroscope Outflow Channel 50-201XL by Hologic
- SCOPE LIGHT by Zinnanti Surgical Instruments, Inc.
- KARL STORZ FIXED MAGNIFICATION TELESCOPE/KS VARIABLE MAGNIFICATION TELESCOPE by Karl Storz Endoscopy-America, Inc.
- Olympus WA47740S by Olympus Winter & Ibe GmbH
- Aveta Pearl Disposable Hysteroscope 214-228 by Meditrina, Inc.
- SCOPE MASTER by Synemed, Inc.
- Luminelle DTx Hysteroscopy System, Luminelle 360 Rotatable Sterile Disposable Sheath Rigid by Uvision360, Inc.
- MyoSure Lite Tissue Removal Device (1 Pack), MyoSure Lite Tissue Removal Device (3 Pack) by Hologic, Inc.