USTAR II Hip System
by United Orthopedic Corporation Taiwan EUDAMEDGUDID
Ref 1115-1417
Product code KROEMDN P09080404
Regulatory status
Recall ongoing. 4 ongoing FDA recalls. See recalls
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| European Union | CE/MDD EUDAMED | NumberB-04719872153924 | Date30 Dec 2024 | StatusActive |
| United States | GUDID listing GUDID | Number04719872153924 | Date24 Feb 2020 | StatusIn distribution |
Coverage
- Expiring (legacy directive)
- Listed
- Not covered
- European Union
- United States
Recalls
-
- Ongoing Z-2884-2026
Due to an incomplete process of the component screw by a subcontractor which prevents the properly assembling of stem trail.
-
- Ongoing Z-1216-2024
- Ongoing Z-1215-2024
- Ongoing Z-1217-2024
Their is a potential that the implant curved stem may be oriented incorrectly.
-
- Terminated Z-0110-2022
Due to product not meeting specification for concentricity and failing to meet requirements of dynamic fatigue testing.
Specification
- Trade Name
- USTAR II Hip System EUDAMED"UNITED" USTAR II Hip System GUDID
- Generic Name
- Metal-on-polyethylene total hip prosthesis
- Device Name
- USTAR II Hip System Press-Fit Curved Stem, RHS, Ti Plasma Spray, 17mm Dia., 200mm Stem Length
- Model / Reference
- 1115-1417
- Primary DI / UDI-DI
- 04719872153924
- Basic UDI-DI
- B-04719872153924
- 510(k) Number
- FDA Product Code
- KRO, LPH, LWJ, KWL
- EMDN Code
- P09080404 Large resections femoral stems
- GMDN Term
- Metal-on-polyethylene total hip prosthesis
- Device Class
- Class III EUDAMEDClass II GUDID
- Regulation Number
- 888.3510, 888.3358, 888.3360
- Market of Registration
- France
- Commercial Distribution
- In Commercial Distribution
Truth per registry
| Field | EUDAMED | GUDID |
|---|---|---|
| Device Class | Class III | Class II |
Manufacturer details
- EUDAMED SRN
- TW-MF-000009523
- Authorised representative
- mdi Europa GmbH DE-AR-000006218
Other devices from United Orthopedic Corp.
Properties & raw data
Show properties, registry sources and raw records
Other properties (18)
- Reusable: No
- Contains Latex: No
- Reprocessed: No
- CMR Substance: No
- Human-derived: No
- Human Tissues: No
- Animal Tissues: No
- Endocrine Disruptor: No
- Administering Medicine: No
- Measuring Function: No
- Incorporates Medicinal Product: No
- Over-the-Counter: No
- Convenience Kit: No
- Combination Product: No
- HCT/P: No
- Premarket Exempt: No
- Serial Number: No
- Manufacturing Date: No
Sources (2)
- EUDAMED aa650100-b444-49f1-8ce3-3cdb14a81a68
- FDA UDI c281038e-811d-4065-bb60-ab4754a3b4cc