  "format": "devicerts.device.v1",
  "cite_as": "V2 Injection Site, B. Braun Medical, Inc. deviCERTS medical device catalogue. https://devicerts.com/devices/v2-injection-site-fpa (data as of 2026-11-24).",
  "source": {
    "canonical_url": "https://devicerts.com/devices/v2-injection-site-fpa",
    "json_url": "https://devicerts.com/devices/v2-injection-site-fpa.json",
    "markdown_url": "https://devicerts.com/devices/v2-injection-site-fpa.md",
    "publisher": "deviCERTS (devicerts.com)",
    "data_sources": [
      {
        "registry": "FDA 510(k)",
        "url": "https://open.fda.gov/apis/device/510k/"
      },
      {
        "registry": "FDA recalls",
        "url": "https://open.fda.gov/apis/device/recall/"
      }
    ],
    "registry_records_last_synced": "2026-11-08",
    "status_as_of": "2026-11-24",
    "license": "CC BY 4.0",
    "license_url": "https://creativecommons.org/licenses/by/4.0/",
    "license_scope": "The deviCERTS layout and text are licensed under CC BY 4.0. Registry data remain under their source terms: U.S. FDA data (openFDA, AccessGUDID) are U.S. public domain; EUDAMED data are subject to the European Commission reuse terms.",
    "attribution": "DevicErts, https://devicerts.com/devices/v2-injection-site-fpa",
    "disclaimer": "Compiled by deviCERTS (devicerts.com) from public medical device registries (EUDAMED, European Commission; U.S. FDA via openFDA; AccessGUDID, U.S. National Library of Medicine). deviCERTS is not an official register and gives no regulatory advice: the record in the registry itself prevails."
  },
  "device": {
    "name": "V2 Injection Site",
    "registry_name": "V2 INJECTION SITE",
    "url": "https://devicerts.com/devices/v2-injection-site-fpa",
    "manufacturer": {
      "name": "B. Braun Medical, Inc.",
      "url": "https://devicerts.com/manufacturers/b-braun-medical-inc",
      "country": "United States",
      "country_basis": "FDA 510(k) applicant",
      "fda_registration_country": null
    },
    "status": {
      "state": "active",
      "label": "Active",
      "detail": "United States"
    },
    "classification": [
      {
        "class": "Class II",
        "system": "US FDA"
      }
    ],
    "identifiers": [
      {
        "label": "510(k) Number",
        "values": [
          {
            "value": "K955585",
            "registries": [
              "FDA 510(k)"
            ]
          }
        ]
      }
    ],
    "product_codes": [
      "FPA"
    ],
    "emdn": [],
    "gmdn": [],
    "description": null,
    "noindex": false
  },
  "authorizations": [
    {
      "registry": "510(k)",
      "region": "United States",
      "type": "FDA 510(k)",
      "number": "K955585",
      "number_label": null,
      "device_udi_di": null,
      "date": "1996-05-15",
      "date_kind": "decision",
      "decision_code": "SESE",
      "expires": null,
      "withdrawn": null,
      "status": "Active",
      "status_note": null,
      "derived_from_gudid": false
    }
  ],
  "no_longer_listed_by": [],
  "certificates": [],
  "recalls": [
    {
      "number": "Z-0714-2026",
      "classification": "Class II",
      "status": "ongoing",
      "initiated": "2025-10-29",
      "reason": "Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion).",
      "applies": "this device",
      "via": "K955585"
    },
    {
      "number": "Z-0090-2023",
      "classification": "Class II",
      "status": "ongoing",
      "initiated": "2022-09-15",
      "reason": "The extension set has a high probability of  failure which could lead to possible leakage or detachment at the device's bonded joint during use. The issue may result in delay of therapy, increased risk of bloodstream infections (BSI), and may expose both patients and healthcare workers to hazardous medications.  Depending on the nature of the leak and type of infusion, there is also the potential for life-threatening situations to occur including air embolization or hypovolemia.",
      "applies": "this device",
      "via": "K955585"
    },
    {
      "number": "Z-0987-2014",
      "classification": "Class II",
      "status": "terminated",
      "initiated": "2014-01-08",
      "reason": "Complaints were received reporting leakage on certain lots of SAFSITE Injection Sites and IV Sets manufactured with SAFSITE Injection Sites.The SAFSITE Injection Site is intended to provide needle free access on IV and Extension Sets.",
      "applies": "this device",
      "via": "K955585"
    },
    {
      "number": "Z-2166-2013",
      "classification": "Class II",
      "status": "terminated",
      "initiated": "2013-08-05",
      "reason": "Reports were received of an incorrectly assembled Outlook Safety Infusion System Burette Set (Product Code 375038), and the investigation determined that one of the ULTRASITE injection sites and corresponding low pressure backcheck valve was assembled upside down.",
      "applies": "shared submission",
      "via": "K955585"
    },
    {
      "number": "Z-0678-2010",
      "classification": "",
      "status": "terminated",
      "initiated": "2009-11-24",
      "reason": "System error 9 alarms. A System Error 9 alarm indicates a potential pressure loss in the pump cassette delivery chamber (chamber closest to the patient). This alarm is generated during the pumps initiation sequence and will not allow the set to be used. This requires a new set to be obtained, thus creating a potential for delay in therapy.",
      "applies": "this device",
      "via": "K955585"
    },
    {
      "number": "Z-0684-2010",
      "classification": "",
      "status": "terminated",
      "initiated": "2009-11-24",
      "reason": "System error 9 alarms. A System Error 9 alarm indicates a potential pressure loss in the pump cassette delivery chamber (chamber closest to the patient). This alarm is generated during the pumps initiation sequence and will not allow the set to be used. This requires a new set to be obtained, thus creating a potential for delay in therapy.",
      "applies": "this device",
      "via": "K955585"
    },
    {
      "number": "Z-0685-2010",
      "classification": "",
      "status": "terminated",
      "initiated": "2009-11-24",
      "reason": "System error 9 alarms. A System Error 9 alarm indicates a potential pressure loss in the pump cassette delivery chamber (chamber closest to the patient). This alarm is generated during the pumps initiation sequence and will not allow the set to be used. This requires a new set to be obtained, thus creating a potential for delay in therapy.",
      "applies": "this device",
      "via": "K955585"
    },
    {
      "number": "Z-0686-2010",
      "classification": "",
      "status": "terminated",
      "initiated": "2009-11-24",
      "reason": "System error 9 alarms. A System Error 9 alarm indicates a potential pressure loss in the pump cassette delivery chamber (chamber closest to the patient). This alarm is generated during the pumps initiation sequence and will not allow the set to be used. This requires a new set to be obtained, thus creating a potential for delay in therapy.",
      "applies": "this device",
      "via": "K955585"
    },
    {
      "number": "Z-0706-2010",
      "classification": "",
      "status": "terminated",
      "initiated": "2009-11-24",
      "reason": "System error 9 alarms. A System Error 9 alarm indicates a potential pressure loss in the pump cassette delivery chamber (chamber closest to the patient). This alarm is generated during the pumps initiation sequence and will not allow the set to be used. This requires a new set to be obtained, thus creating a potential for delay in therapy.",
      "applies": "this device",
      "via": "K955585"
    },
    {
      "number": "Z-0707-2010",
      "classification": "",
      "status": "terminated",
      "initiated": "2009-11-24",
      "reason": "System error 9 alarms. A System Error 9 alarm indicates a potential pressure loss in the pump cassette delivery chamber (chamber closest to the patient). This alarm is generated during the pumps initiation sequence and will not allow the set to be used. This requires a new set to be obtained, thus creating a potential for delay in therapy.",
      "applies": "this device",
      "via": "K955585"
    },
    {
      "number": "Z-0708-2010",
      "classification": "",
      "status": "terminated",
      "initiated": "2009-11-24",
      "reason": "System error 9 alarms. A System Error 9 alarm indicates a potential pressure loss in the pump cassette delivery chamber (chamber closest to the patient). This alarm is generated during the pumps initiation sequence and will not allow the set to be used. This requires a new set to be obtained, thus creating a potential for delay in therapy.",
      "applies": "this device",
      "via": "K955585"
    },
    {
      "number": "Z-1427-2009",
      "classification": "",
      "status": "terminated",
      "initiated": "2009-03-27",
      "reason": "Device defect is occlusion/no flow.",
      "applies": "this device",
      "via": "K955585"
    },
    {
      "number": "Z-1428-2009",
      "classification": "",
      "status": "terminated",
      "initiated": "2009-03-27",
      "reason": "Device defect is occlusion/no flow.",
      "applies": "this device",
      "via": "K955585"
    },
    {
      "number": "Z-1429-2009",
      "classification": "",
      "status": "terminated",
      "initiated": "2009-03-27",
      "reason": "Device defect is occlusion/no flow.",
      "applies": "this device",
      "via": "K955585"
    },
    {
      "number": "Z-1430-2009",
      "classification": "",
      "status": "terminated",
      "initiated": "2009-03-27",
      "reason": "Device defect is occlusion/no flow.",
      "applies": "this device",
      "via": "K955585"
    },
    {
      "number": "Z-1431-2009",
      "classification": "",
      "status": "terminated",
      "initiated": "2009-03-27",
      "reason": "Device defect is occlusion/no flow.",
      "applies": "this device",
      "via": "K955585"
    },
    {
      "number": "Z-1432-2009",
      "classification": "",
      "status": "terminated",
      "initiated": "2009-03-27",
      "reason": "Device defect is occlusion/no flow.",
      "applies": "this device",
      "via": "K955585"
    },
    {
      "number": "Z-1433-2009",
      "classification": "",
      "status": "terminated",
      "initiated": "2009-03-27",
      "reason": "Device defect is occlusion/no flow.",
      "applies": "this device",
      "via": "K955585"
    },
    {
      "number": "Z-0199-2009",
      "classification": "",
      "status": "terminated",
      "initiated": "2008-09-17",
      "reason": "Product was assembled incorrectly and connected to the wrong part.",
      "applies": "this device",
      "via": "K955585"
    },
    {
      "number": "Z-2407-2008",
      "classification": "",
      "status": "terminated",
      "initiated": "2008-06-30",
      "reason": "incorrect burette was packaged with the product",
      "applies": "shared submission",
      "via": "K955585"
    },
    {
      "number": "Z-2410-2008",
      "classification": "",
      "status": "terminated",
      "initiated": "2008-06-30",
      "reason": "incorrect burette was packaged with the product",
      "applies": "this device",
      "via": "K955585"
    },
    {
      "number": "Z-2411-2008",
      "classification": "",
      "status": "terminated",
      "initiated": "2008-06-30",
      "reason": "incorrect burette was packaged with the product",
      "applies": "this device",
      "via": "K955585"
    },
    {
      "number": "Z-2412-2008",
      "classification": "",
      "status": "terminated",
      "initiated": "2008-06-30",
      "reason": "incorrect burette was packaged with the product",
      "applies": "this device",
      "via": "K955585"
    },
    {
      "number": "Z-2413-2008",
      "classification": "",
      "status": "terminated",
      "initiated": "2008-06-30",
      "reason": "incorrect burette was packaged with the product",
      "applies": "this device",
      "via": "K955585"
    },
    {
      "number": "Z-0818-2007",
      "classification": "",
      "status": "terminated",
      "initiated": "2007-04-18",
      "reason": "Secondary infusion backs up into primary container.",
      "applies": "this device",
      "via": "K955585"
    },
    {
      "number": "Z-0819-2007",
      "classification": "",
      "status": "terminated",
      "initiated": "2007-04-18",
      "reason": "Secondary infusion backs up into primary container.",
      "applies": "this device",
      "via": "K955585"
    },
    {
      "number": "Z-0933-03",
      "classification": "",
      "status": "terminated",
      "initiated": "2003-04-14",
      "reason": "possible leakage",
      "applies": "this device",
      "via": "K955585"
    }
  ],
  "eu_market_availability": null,
  "properties": [
    {
      "group": "Identifiers",
      "fields": [
        {
          "label": "510(k) Number",
          "values": [
            {
              "value": "K955585",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        },
        {
          "label": "FDA product code",
          "values": [
            {
              "value": "FPA",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        }
      ]
    },
    {
      "group": "Classification",
      "fields": [
        {
          "label": "Device class",
          "values": [
            {
              "value": "Class II",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        },
        {
          "label": "Regulation",
          "values": [
            {
              "value": "21 CFR Part 880.5440",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        },
        {
          "label": "Advisory committee",
          "values": [
            {
              "value": "General Hospital",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        },
        {
          "label": "FDA decision",
          "values": [
            {
              "value": "SESE",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        }
      ]
    }
  ]
}