> Active: European Union and United States. Medical device record on deviCERTS, compiled from public registries.

- Canonical page: <https://devicerts.com/devices/vantage-galan-3t-mrt-3020>
- Manufacturer: [Canon Inc.](https://devicerts.com/manufacturers/canon-inc) (Japan)
- Device class: Class IIa (EU MDR); Class II (US FDA)
- FDA product code: LNH
- EMDN: Z11050105 CLOSED MAGNET MRI SCANNERS WITH MAGNETIC FIELD GREATER THAN 2T AND LESS THAN OR EQUAL TO 4T
- GMDN: Full-body MRI system, superconducting magnet (37654)

## Registrations and authorizations

- **EUDAMED** · CE/MDR · European Union · Basic UDI-DI 4987670A0601RR · registered 2026-06-18 · status: Active
- **GUDID** · FDA 510(k) · United States · K243335 · decision 2025-01-07 · decision SESE · status: Active · from GUDID
- **GUDID** · FDA 510(k) · United States · K241496 · decision 2024-08-20 · decision SESE · status: Active · from GUDID
- **GUDID** · FDA 510(k) · United States · K230355 · decision 2023-08-30 · decision SESE · status: Active · from GUDID
- **GUDID** · FDA 510(k) · United States · K222387 · decision 2022-08-31 · decision SESE · status: Active · from GUDID
- **GUDID** · FDA 510(k) · United States · K220192 · decision 2022-04-08 · decision SESE · status: Active · from GUDID
- **GUDID** · FDA 510(k) · United States · K212056 · decision 2021-08-04 · decision SESE · status: Active · from GUDID
- **GUDID** · FDA 510(k) · United States · K211037 · decision 2021-05-17 · decision SESE · status: Active · from GUDID
- **GUDID** · FDA 510(k) · United States · K203323 · decision 2021-03-04 · decision SESE · status: Active · from GUDID
- **GUDID** · FDA 510(k) · United States · K203553 · decision 2021-02-02 · decision SESE · status: Active · from GUDID
- **GUDID** · FDA 510(k) · United States · K192574 · decision 2020-03-09 · decision SESE · status: Active · from GUDID
- **GUDID** · FDA 510(k) · United States · K192506 · decision 2020-03-02 · decision SESE · status: Active · from GUDID
- **GUDID** · FDA 510(k) · United States · K183657 · decision 2019-03-15 · decision SESE · status: Active · from GUDID
- **GUDID** · FDA 510(k) · United States · K181593 · decision 2018-08-13 · decision SESE · status: Active · from GUDID
- **GUDID** · FDA 510(k) · United States · K173382 · decision 2018-01-26 · decision SESE · status: Active · from GUDID
- **GUDID** · GUDID listing · United States · UDI-DI 04987670102615 · listed 2017-03-08 · status: Listed (GUDID: in commercial distribution)
- **GUDID** · FDA 510(k) · United States · K162183 · decision 2016-11-25 · decision SESE · status: Active · from GUDID

## Recalls

- Z-1017-2026 · Class II · ongoing · initiated 2025-12-19: There is a potential for formation of ice occurring in the venting system of the superconducting magnet of the MRI systems. if a quench occurs in this condition, helium gas may not be able to escape via the designed vent paths, and the pressure in the helium vessel may rise. This pressure build-up could eventually rupture the helium vessel, causing helium gas to be released into the MR scanning room.
- Z-2041-2024 · Class II · ongoing · initiated 2024-05-10: The terminal block used to secure the power supply cable for the gradient coil of the MRI system was not secured as designed, which could cause cable disconnection, which could lead to arcing, followed by melting of the cable, ignition, smoke, and fire.
- Z-0327-2026 · Class II · ongoing · initiated 2025-10-16: Affected system contains components that do not meet the required distance between electrodes for compliance with the IEC 60601-1 standard for medical electrical equipment.
- Z-2047-2021 · Class II · ongoing · initiated 2021-05-21: The MRI system contains a component that does not meet the appropriate flame retardancy-rating which could result in smoke or fire.

## Identity

- Trade name: Vantage Galan 3T [FDA GUDID (AccessGUDID)]
- Generic name: Full-body MRI system, superconducting magnet [FDA GUDID (AccessGUDID)]
- Device name: Magnetic Resonance Imaging System [EUDAMED]
- Model / reference: MRT-3020 [EUDAMED, FDA GUDID (AccessGUDID)]; MRT-1550 [EUDAMED]; etc. [EUDAMED]

## Identifiers

- Primary DI / UDI-DI: 04987670102615 [EUDAMED, FDA GUDID (AccessGUDID)]; 04987670104879 [EUDAMED]
- Basic UDI-DI: 4987670A0601RR [EUDAMED]
- 510(k) Number: K162183 [FDA GUDID (AccessGUDID)]; K173382 [FDA GUDID (AccessGUDID)]; K181593 [FDA GUDID (AccessGUDID)]; K183657 [FDA GUDID (AccessGUDID)]; K192506 [FDA GUDID (AccessGUDID)]; K192574 [FDA GUDID (AccessGUDID)]; K203323 [FDA GUDID (AccessGUDID)]; K203553 [FDA GUDID (AccessGUDID)]; K211037 [FDA GUDID (AccessGUDID)]; K212056 [FDA GUDID (AccessGUDID)]; K220192 [FDA GUDID (AccessGUDID)]; K222387 [FDA GUDID (AccessGUDID)]; K230355 [FDA GUDID (AccessGUDID)]; K241496 [FDA GUDID (AccessGUDID)]; K243335 [FDA GUDID (AccessGUDID)]
- FDA product code: LNH [FDA GUDID (AccessGUDID)]
- EMDN code: Z11050105 [EUDAMED]
- GMDN term: Full-body MRI system, superconducting magnet (37654) [FDA GUDID (AccessGUDID)]

## Classification

- Device class: Class IIa [EUDAMED]; Class II [FDA GUDID (AccessGUDID)]
- Regulation number: 892.1000 [FDA GUDID (AccessGUDID)]
- EU registration type: Standard device [EUDAMED]
- EU market of registration: Netherlands [EUDAMED]
- GUDID distribution status: In Commercial Distribution [FDA GUDID (AccessGUDID)]

## Properties

- Serial number: Yes [FDA GUDID (AccessGUDID)]
- Supplied sterile: No [EUDAMED, FDA GUDID (AccessGUDID)]
- Sterilise before use: No [EUDAMED, FDA GUDID (AccessGUDID)]
- Single use: No [FDA GUDID (AccessGUDID)]
- Implantable: No [EUDAMED, FDA GUDID (AccessGUDID)]
- Reusable: No [EUDAMED]
- Contains latex: No [EUDAMED, FDA GUDID (AccessGUDID)]
- Reprocessed single-use device: No [EUDAMED]
- Contains CMR substances (carcinogenic, mutagenic or toxic to reproduction): No [EUDAMED]
- Incorporates a substance derived from human blood or plasma: No [EUDAMED]
- Contains human tissues or cells: No [EUDAMED]
- Contains animal tissues or cells: No [EUDAMED]
- Contains endocrine-disrupting substances: No [EUDAMED]
- Administering medicine: No [EUDAMED]
- Measuring function: No [EUDAMED]
- Incorporates a medicinal substance: No [EUDAMED]
- Prescription use: No [FDA GUDID (AccessGUDID)]
- Over-the-counter: No [FDA GUDID (AccessGUDID)]
- Convenience kit: No [FDA GUDID (AccessGUDID)]
- Combination product: No [FDA GUDID (AccessGUDID)]
- HCT/P: No [FDA GUDID (AccessGUDID)]
- Premarket exempt: No [FDA GUDID (AccessGUDID)]
- Lot / batch number: No [FDA GUDID (AccessGUDID)]
- Expiration date: No [FDA GUDID (AccessGUDID)]
- Manufacturing date: No [FDA GUDID (AccessGUDID)]

## Source and citation

- Cite as: Vantage Galan 3T MRT-3020, Canon Inc. deviCERTS medical device catalogue. https://devicerts.com/devices/vantage-galan-3t-mrt-3020 (data as of 2026-11-29).
- Data sources: [FDA GUDID (AccessGUDID)](https://accessgudid.nlm.nih.gov/), [EUDAMED](https://ec.europa.eu/tools/eudamed), [FDA recalls](https://open.fda.gov/apis/device/recall/)
- Registry records last synced: 2026-11-07
- Statuses computed as of: 2026-11-29
- JSON: <https://devicerts.com/devices/vantage-galan-3t-mrt-3020.json>
- Licence: [CC BY 4.0](https://creativecommons.org/licenses/by/4.0/). The deviCERTS layout and text are licensed under CC BY 4.0. Registry data remain under their source terms: U.S. FDA data (openFDA, AccessGUDID) are U.S. public domain; EUDAMED data are subject to the European Commission reuse terms.
- Attribution: DevicErts, https://devicerts.com/devices/vantage-galan-3t-mrt-3020
- Compiled by deviCERTS (devicerts.com) from public medical device registries (EUDAMED, European Commission; U.S. FDA via openFDA; AccessGUDID, U.S. National Library of Medicine). deviCERTS is not an official register and gives no regulatory advice: the record in the registry itself prevails.