Vantage Titan with Helios Gradient Model Mrt-1504/Us
by Toshiba America Medical Systems, In.C
510(k) K122613
Active
Valid until
No expiry
FDA recalls
3 ongoingDocuments
Trust
Unverified
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK122613 | Date13 Dec 2012 | StatusActive |
Recalls
-
- Ongoing Z-2041-2024
The terminal block used to secure the power supply cable for the gradient coil of the MRI system was not secured as designed, which could cause cable disconnection, which could lead to arcing, followed by melting of the cable, ignition, smoke, and fire.
-
- Ongoing Z-1250-2024
For some MRI systems, it has been found that some of the maximum Spatial Field Gradient (SFG) values displayed in "System Information" of the operation window and some of the maximum SFG values described in the safety manual are lower than the correct values, which may cause a patient with an MR conditional device to feel some discomfort during the MR scan.
-
- Ongoing Z-2044-2021
The MRI system contains a component that does not meet the appropriate flame retardancy-rating which could result in smoke or fire.
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA Product Code
- LNH
- Device Class
- Class II
- Regulation
- 21 CFR Part 892.1000
- Advisory Committee
- Radiology
- FDA Decision
- SESE Substantially Equivalent
Other devices from Toshiba America Medical Systems, In.C
See all devices from Toshiba America Medical Systems, In.C →