Varian C-Series Clinacs Clinac 600C, 2100C, 2500C
510(k) K904364
Active
Valid until
No expiry
FDA recalls
None ongoingDocuments
Trust
Unverified
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK904364 | Date27 Feb 1991 | StatusActive |
Recalls
-
- Terminated Z-1249-2011
Position errors may result in misdelivery of radiation.
-
- Terminated Z-0886-2011
The Auto goto or Auto setup functions in C-Series version 7 software ignore the couch angle and exceed motion zone (unless a tolerance is defined by the user); it has the potential for collision with the patient on the couch.
-
- Terminated Z-1003-2011
The product has been shipped since 2005 with the wrong label regarding max weight to be used on accessory rails.
-
- Terminated Z-1729-2009
Unexpected Movement: if the stereotactic motion disable function is turned on, the couch can be moved via float mode unexpectedly.
-
- Terminated Z-0117-2009
During couch movement, a patient's fingers may get pinched at certain points which could result in broken, disengaged or pinched fingers.
-
- Terminated Z-1530-05
Emergency Off Switch located on the device may not function correctly resulting in the device remaining powered on after the switch is activated.
-
- Terminated Z-0777-05
Treatment beam may not completely pause while device changes field shape; Potential for mistreatment could occur.
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA Product Code
- IYE
- Device Class
- Class II
- Regulation
- 21 CFR Part 892.5050
- Advisory Committee
- Radiology
- FDA Decision
- SESE Substantially Equivalent