Variant II Turbo Link Hemoglobin A1C Program, and Variant II Turbo Link Hemoglobin Testing with Model, 270-2716,
by Bio-Rad Laboratories Inc., Clinical Systems Divisi 510(k)
Active
510(k) K070819
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK070819 | Date25 Jun 2007 | StatusActive |
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Recalls
-
- Terminated Z-0399-2012
It was discovered that the Variant II Turbo Link Calibrator Values card contains incorrect values for IFCC units. The numbers are missing a decimal point.
-
- Terminated Z-0027-2012
- Terminated Z-0026-2012
A previous recall related to reports of calibration failure and late retention times using Elution Buffer led to an examination of other buffers manufactured during the same time period and a decision was made to field correct these products until a definitive root cause and corrective action have been found and implemented.
-
- Terminated Z-3159-2011
The Elution Buffer B, lot no. GA 10430, was shipped in a 2.0 L bottle instead of the standard 2.5 L bottle.
-
- Terminated Z-1259-2011
The results from the blood device had the occurrence of some ramping baselines on their chromatograms. The ramping baseline can affect quantitation of the blood results.
Specification
- 510(k) Number
- FDA Product Code
- LCP
- Device Class
- Class II
- Regulation
- 21 CFR Part 864.7470
- Advisory Committee
- Hematology
- FDA Decision
- SESE Substantially Equivalent