Varicam

by Elscint, Inc.

510(k) K953801

Active United States

510(k)

Valid until

No expiry

510(k) clearance is not time-limited

FDA recalls

2 ongoing

This device is named in the recall

Documents

Trust

Unverified

Registry data, not yet checked by deviCERTS. Source FDA 510(k), synced 22 Oct 2026. What this means

Regulatory status

as of registry sync 22 Oct 2026

RegionAuthorizationNumberDateStatus
United StatesFDA 510(k)SESE Substantially Equivalent · Traditional NumberK953801 Date22 Nov 1995ClearedStatusActive

Recalls

7 recalls · 2 ongoing · FDA

    • Ongoing Z-2098-2025 FDA Class II initiated 20 Jun 2025

    Unintended radial detector motion may occur during patient setup or during patient scan if system does not have correct version of gantry software installed. Unintended detector motion may result in life-threatening injury.

    • Ongoing Z-0900-2025 FDA Class II initiated 18 Dec 2024

    GE HealthCare has become aware that certain Infinia, VariCam/VG/VH, Brivo NM 615, Discovery NM 630, Discovery NM/CT 670, and Optima NM/CT 640 Nuclear Medicine systems (see Affected Products Details and Appendices) could have been transported or relocated without proper detector support, which can result in excessive load being applied to the detector mounting mechanisms. If this occurs, it can potentially result in a detector fall and lead to life-threatening bodily injury.

  • 2 recalls, same reason

    • Completed Z-0704-2019 FDA Class II initiated 30 Oct 2017
    • Completed Z-0703-2019 FDA Class II initiated 30 Oct 2017

    A potential hazardous situation may occur if the collimator locking handle is not properly placed in the lock position and the warning circuit does not detect that the collimator is unlocked because of the patient proximity to the collimator during a clinical scan.

    • Terminated Z-1279-2015 FDA Class II initiated 5 Mar 2015

    GE Healthcare has recently become aware of a potential safety issue due to a collimator falling off a VG system during a Quality Control check, while the detectors were rotating. No injuries have been reported to date

    • Terminated Z-1788-2013 FDA Class I initiated 13 Jun 2013

    GE became aware of an incident at a VA Medical Center facility in the US in which a patient died due to injuries sustained while being scanned on an Infinia Hawkeye 4 Nuclear Medicine System.

    • Terminated Z-0012-05 FDA Unclassified (pre-2012) initiated 3 Sep 2004

    GE Healthcare received reports of intermittent low ejection fraction (EF) results for MUGA (multi-gated acquisitions) studies on Millennium VG systems. This issue can occasionally lead to distortions in the volume curve in some MUGA studies, and result in calculated EF values lower (never higher) than actual values. A low EF value could lead to a physician decision to stop needed treatment.

Coverage

1 jurisdiction

United States of America
  • Active
  • Expiring
  • Withdrawn
  • Listed
  • Not covered

United States — FDA 510(k) Active

Specification

510(k)

510(k) Number
FDA Product Code
KPS
Device Class
Class II
Regulation
21 CFR Part 892.1200
Advisory Committee
Radiology
FDA Decision
SESE Substantially Equivalent