Varicam
510(k) K953801
Active
Valid until
No expiry
FDA recalls
2 ongoingDocuments
Trust
Unverified
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK953801 | Date22 Nov 1995 | StatusActive |
Recalls
-
- Ongoing Z-2098-2025
Unintended radial detector motion may occur during patient setup or during patient scan if system does not have correct version of gantry software installed. Unintended detector motion may result in life-threatening injury.
-
- Ongoing Z-0900-2025
GE HealthCare has become aware that certain Infinia, VariCam/VG/VH, Brivo NM 615, Discovery NM 630, Discovery NM/CT 670, and Optima NM/CT 640 Nuclear Medicine systems (see Affected Products Details and Appendices) could have been transported or relocated without proper detector support, which can result in excessive load being applied to the detector mounting mechanisms. If this occurs, it can potentially result in a detector fall and lead to life-threatening bodily injury.
-
- Completed Z-0704-2019
- Completed Z-0703-2019
A potential hazardous situation may occur if the collimator locking handle is not properly placed in the lock position and the warning circuit does not detect that the collimator is unlocked because of the patient proximity to the collimator during a clinical scan.
-
- Terminated Z-1279-2015
GE Healthcare has recently become aware of a potential safety issue due to a collimator falling off a VG system during a Quality Control check, while the detectors were rotating. No injuries have been reported to date
-
- Terminated Z-1788-2013
GE became aware of an incident at a VA Medical Center facility in the US in which a patient died due to injuries sustained while being scanned on an Infinia Hawkeye 4 Nuclear Medicine System.
-
- Terminated Z-0012-05
GE Healthcare received reports of intermittent low ejection fraction (EF) results for MUGA (multi-gated acquisitions) studies on Millennium VG systems. This issue can occasionally lead to distortions in the volume curve in some MUGA studies, and result in calculated EF values lower (never higher) than actual values. A low EF value could lead to a physician decision to stop needed treatment.
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA Product Code
- KPS
- Device Class
- Class II
- Regulation
- 21 CFR Part 892.1200
- Advisory Committee
- Radiology
- FDA Decision
- SESE Substantially Equivalent