# VarioSurg SG30

> Active: United States. Medical device record on deviCERTS, compiled from public registries.

- Canonical page: <https://devicerts.com/devices/variosurg-sg30>
- Manufacturer: [Nakanishi Inc.](https://devicerts.com/manufacturers/nakanishi-inc) (Japan)
- Device class: Class II (US FDA)
- FDA product code: DZI, ELC
- GMDN: Dental ultrasonic surgical system handpiece tip (47015)

## Registrations and authorizations

- **GUDID** · GUDID listing · United States · UDI-DI 04560264563605 · listed 2016-09-16 · status: Listed (GUDID: in commercial distribution)
- **GUDID** · FDA 510(k) · United States · K073678 · decision 2008-07-10 · decision SESE · status: Active · from GUDID

## Recalls

- Z-2673-2016 · Class II · terminated · initiated 2016-07-08: Through an error, the labeling failed to include the symbol or other text indicating that the diamond coated tips were single use only. A contributing factor to this error is that in other markets such as Japan, the diamond coated tips are multiple use and are not limited to single use only. The tips are stainless steel with a coating of fine diamond powder. The tips are sold non-sterile and are to be sterilized prior to first use. The tips were cleared as single use only. This problem was discovered during an ongoing, proactive, quality system improvement plan that was initiated by Nakanishi in October 2014. Because the correct instructions for sterilization are not available, it is possible that the devices might not be sterile prior to use. The absence of the single use statement introduces a risk of cross-contamination if the tips are reprocessed and reused.

## Identity

- Trade name: VarioSurg [FDA GUDID (AccessGUDID)]
- Generic name: Dental ultrasonic surgical system handpiece tip [FDA GUDID (AccessGUDID)]
- Model / reference: SG30 [FDA GUDID (AccessGUDID)]

## Identifiers

- Primary DI / UDI-DI: 04560264563605 [FDA GUDID (AccessGUDID)]
- 510(k) Number: K073678 [FDA GUDID (AccessGUDID)]
- FDA product code: DZI [FDA GUDID (AccessGUDID)]; ELC [FDA GUDID (AccessGUDID)]
- GMDN term: Dental ultrasonic surgical system handpiece tip (47015) [FDA GUDID (AccessGUDID)]

## Classification

- Device class: Class II [FDA GUDID (AccessGUDID)]
- Regulation number: 872.4120 [FDA GUDID (AccessGUDID)]; 872.4850 [FDA GUDID (AccessGUDID)]
- GUDID distribution status: In Commercial Distribution [FDA GUDID (AccessGUDID)]

## Clinical & Handling

- Sterilisation method: Moist Heat or Steam Sterilization [FDA GUDID (AccessGUDID)]

## Properties

- Sterilise before use: Yes [FDA GUDID (AccessGUDID)]
- Prescription use: Yes [FDA GUDID (AccessGUDID)]
- Lot / batch number: Yes [FDA GUDID (AccessGUDID)]
- Manufacturing date: Yes [FDA GUDID (AccessGUDID)]
- Supplied sterile: No [FDA GUDID (AccessGUDID)]
- Single use: No [FDA GUDID (AccessGUDID)]
- Implantable: No [FDA GUDID (AccessGUDID)]
- Contains latex: No [FDA GUDID (AccessGUDID)]
- Over-the-counter: No [FDA GUDID (AccessGUDID)]
- Convenience kit: No [FDA GUDID (AccessGUDID)]
- Combination product: No [FDA GUDID (AccessGUDID)]
- HCT/P: No [FDA GUDID (AccessGUDID)]
- Premarket exempt: No [FDA GUDID (AccessGUDID)]
- Serial number: No [FDA GUDID (AccessGUDID)]
- Expiration date: No [FDA GUDID (AccessGUDID)]

## Source and citation

- Cite as: VarioSurg SG30, Nakanishi Inc. deviCERTS medical device catalogue. https://devicerts.com/devices/variosurg-sg30 (data as of 2026-11-19).
- Data sources: [FDA GUDID (AccessGUDID)](https://accessgudid.nlm.nih.gov/), [FDA recalls](https://open.fda.gov/apis/device/recall/)
- Registry records last synced: 2026-10-23
- Statuses computed as of: 2026-11-19
- JSON: <https://devicerts.com/devices/variosurg-sg30.json>
- Licence: [CC BY 4.0](https://creativecommons.org/licenses/by/4.0/). The deviCERTS layout and text are licensed under CC BY 4.0. Registry data remain under their source terms: U.S. FDA data (openFDA, AccessGUDID) are U.S. public domain; EUDAMED data are subject to the European Commission reuse terms.
- Attribution: DevicErts, https://devicerts.com/devices/variosurg-sg30
- Compiled by deviCERTS (devicerts.com) from public medical device registries (EUDAMED, European Commission; U.S. FDA via openFDA; AccessGUDID, U.S. National Library of Medicine). deviCERTS is not an official register and gives no regulatory advice: the record in the registry itself prevails.