  "format": "devicerts.device.v1",
  "cite_as": "VARIS 1.4G, Varian Medical Systems, Inc. deviCERTS medical device catalogue. https://devicerts.com/devices/varis-1-4g-iye (data as of 2026-11-22).",
  "source": {
    "canonical_url": "https://devicerts.com/devices/varis-1-4g-iye",
    "json_url": "https://devicerts.com/devices/varis-1-4g-iye.json",
    "markdown_url": "https://devicerts.com/devices/varis-1-4g-iye.md",
    "publisher": "deviCERTS (devicerts.com)",
    "data_sources": [
      {
        "registry": "FDA 510(k)",
        "url": "https://open.fda.gov/apis/device/510k/"
      },
      {
        "registry": "FDA recalls",
        "url": "https://open.fda.gov/apis/device/recall/"
      }
    ],
    "registry_records_last_synced": "2026-10-22",
    "status_as_of": "2026-11-22",
    "license": "CC BY 4.0",
    "license_url": "https://creativecommons.org/licenses/by/4.0/",
    "license_scope": "The deviCERTS layout and text are licensed under CC BY 4.0. Registry data remain under their source terms: U.S. FDA data (openFDA, AccessGUDID) are U.S. public domain; EUDAMED data are subject to the European Commission reuse terms.",
    "attribution": "DevicErts, https://devicerts.com/devices/varis-1-4g-iye",
    "disclaimer": "Compiled by deviCERTS (devicerts.com) from public medical device registries (EUDAMED, European Commission; U.S. FDA via openFDA; AccessGUDID, U.S. National Library of Medicine). deviCERTS is not an official register and gives no regulatory advice: the record in the registry itself prevails."
  },
  "device": {
    "name": "VARIS 1.4G",
    "registry_name": "VARIS 1.4G",
    "url": "https://devicerts.com/devices/varis-1-4g-iye",
    "manufacturer": {
      "name": "Varian Medical Systems, Inc.",
      "url": "https://devicerts.com/manufacturers/varian-medical-systems-inc",
      "country": "United States",
      "country_basis": null,
      "fda_registration_country": null
    },
    "status": {
      "state": "active",
      "label": "Active",
      "detail": "United States"
    },
    "classification": [
      {
        "class": "Class II",
        "system": "US FDA"
      }
    ],
    "identifiers": [
      {
        "label": "510(k) Number",
        "values": [
          {
            "value": "K001643",
            "registries": [
              "FDA 510(k)"
            ]
          }
        ]
      }
    ],
    "product_codes": [
      "IYE"
    ],
    "emdn": [],
    "gmdn": [],
    "description": {
      "text": "The VARIS 1.4G is a Medical Linear Accelerator designed to generate high-energy charged particles, such as electrons and protons, for use in radiation therapy. As a charged-particle radiation therapy system, it accelerates particles to precise energies for targeted treatment of malignant and benign conditions, including oncology and radiosurgery applications. Its function relies on linear acceleration technology to deliver controlled radiation doses to clinical targets while minimizing exposure to surrounding tissues.\n\nThis device is intended for use in radiation oncology departments within hospital or clinic settings, where it is integrated into treatment planning systems for precise dose delivery. Supplied as a reusable, non-sterile medical device, the VARIS 1.4G adheres to Class II (Special Controls) regulatory standards under FDA 510(k) clearance (K001643). It operates within standard MRI safety guidelines for non-MRI environments and requires specialized training for operation and maintenance. Storage and handling follow manufacturer protocols to ensure calibration and performance integrity.",
      "ai_generated": true,
      "generated_on": "2026-10-01",
      "note": "Written by a language model from registry data. Not verified by deviCERTS or the manufacturer. The registry data is the record."
    },
    "noindex": false
  },
  "authorizations": [
    {
      "registry": "510(k)",
      "region": "United States",
      "type": "FDA 510(k)",
      "number": "K001643",
      "number_label": null,
      "device_udi_di": null,
      "date": "2000-06-23",
      "date_kind": "decision",
      "decision_code": "SESE",
      "expires": null,
      "withdrawn": null,
      "status": "Active",
      "status_note": null,
      "derived_from_gudid": false
    }
  ],
  "no_longer_listed_by": [],
  "certificates": [],
  "recalls": [
    {
      "number": "Z-1957-2011",
      "classification": "",
      "status": "terminated",
      "initiated": "2011-02-07",
      "reason": "An issue with the Varis Vision software where treatment can occur without the dynamic Multileaf Collimator component and may deliver a higher than expected dose of radiation.",
      "applies": "this device",
      "via": "K001643"
    },
    {
      "number": "Z-0133-2010",
      "classification": "",
      "status": "terminated",
      "initiated": "2009-09-15",
      "reason": "Software issue:  Under certain circumstances, when a therapy treatment field is transferred to VARiS Vision 6.2, 6.1 and 6.0 via LINK procedures, the rotation direction will be reversed.",
      "applies": "this device",
      "via": "K001643"
    },
    {
      "number": "Z-1809-2009",
      "classification": "",
      "status": "terminated",
      "initiated": "2009-05-08",
      "reason": "Potential mistreatment-- due to software issues, retired plans may be used and treated repeatedly.",
      "applies": "this device",
      "via": "K001643"
    },
    {
      "number": "Z-0164-2009",
      "classification": "",
      "status": "terminated",
      "initiated": "2008-06-02",
      "reason": "A malfunction within the Varis software might result in a misadministration (underdose).  The software does not correctly interpret the number, causing field parameters having decimal values to be incorrect.",
      "applies": "this device",
      "via": "K001643"
    },
    {
      "number": "Z-2194-2008",
      "classification": "",
      "status": "terminated",
      "initiated": "2008-05-21",
      "reason": "Dose delivery may be altered after import from a third party planning system during radiation therapy.",
      "applies": "this device",
      "via": "K001643"
    },
    {
      "number": "Z-0720-2008",
      "classification": "",
      "status": "terminated",
      "initiated": "2007-10-02",
      "reason": "Incorrect Software Validation- The ARC treatment plans containing a Dose Dynamic MLC for IMRT delivery that will not be recognized properly by the system. Under these circumstances, an arc plan can be administered to a patient whereby the gantry will not rotate, resulting in delivery to a single location rather than through the gantry arc.",
      "applies": "this device",
      "via": "K001643"
    },
    {
      "number": "Z-0304-2007",
      "classification": "",
      "status": "terminated",
      "initiated": "2006-09-21",
      "reason": "Selecting an empty space next to the ''Override'' or ''Acquire Actual'' buttons may result in mistreatment to the patient.",
      "applies": "this device",
      "via": "K001643"
    },
    {
      "number": "Z-0633-03",
      "classification": "",
      "status": "terminated",
      "initiated": "2003-01-31",
      "reason": "The product has a software problem. This could result in a treatment without a motorized wedge where one had been planned.",
      "applies": "this device",
      "via": "K001643"
    }
  ],
  "eu_market_availability": null,
  "properties": [
    {
      "group": "Identifiers",
      "fields": [
        {
          "label": "510(k) Number",
          "values": [
            {
              "value": "K001643",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        },
        {
          "label": "FDA product code",
          "values": [
            {
              "value": "IYE",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        }
      ]
    },
    {
      "group": "Classification",
      "fields": [
        {
          "label": "Device class",
          "values": [
            {
              "value": "Class II",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        },
        {
          "label": "Regulation",
          "values": [
            {
              "value": "21 CFR Part 892.5050",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        },
        {
          "label": "Advisory committee",
          "values": [
            {
              "value": "Radiology",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        },
        {
          "label": "FDA decision",
          "values": [
            {
              "value": "SESE",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        }
      ]
    }
  ]
}