Skip to content

Venclose System Power Cord VCPCB

by Venclose Inc

Venclose System Power Cord VCPCB is an electrical mains power cable (FDA Class II; FDA product code GEI, Electrosurgical, Cutting & Coagulation & Accessories) made by Venclose Inc, cleared under FDA 510(k) K160754 on 9 Sep 2016 and listed in FDA GUDID since 1 Nov 2016 (UDI-DI 00858254006046), active as of 22 Oct 2026. 2 FDA recalls are linked to it.

Active United States. Also listed in GUDID (US).

510(k)

Valid until

No expiry

510(k) clearance is not time-limited

FDA recalls

2 ongoing

This device is named in the recall

Documents

510(k) K160754

FDA submission, from GUDID

Trust

Unverified

Source GUDID, synced 22 Oct 2026.

Regulatory status

as of registry sync 22 Oct 2026

RegionAuthorizationNumberDateStatus
United StatesGUDID listing Number00858254006046 UDI-DI Date1 Nov 2016ListedStatusListed GUDID: in commercial distribution
United StatesFDA 510(k)SESE from GUDID Substantially Equivalent · Traditional NumberK160754 Date9 Sep 2016DecidedStatusActive

Official records

From the FDA record

Device description
Power Cord, US
MRI safety
Labeling does not contain MRI Safety Information
Natural rubber latex
No latex statement on the label
Sterility
Supplied non-sterile

From the GUDID record (AccessGUDID), as read on 22 Oct 2026, as the labeller entered it.

Across registries

2 registries list this device

  • FDA 510(k) K160754 decided 9 Sep 2016 Active
  • FDA GUDID UDI-DI 00858254006046 listed 1 Nov 2016 Listed

Recalls

2 recalls · 2 ongoing · FDA

    • Ongoing Z-2638-2025 FDA Class II initiated 21 Aug 2025

    Software version 3.35 of the Venclose digiRF Generator incorporates a catheter verification feature specifically designed to detect internal wiring anomalies in Venclose EVSRF Ablation Catheters prior to clinical use. This automated diagnostic check is executed immediately upon catheter connection to the generator, occurring before the procedure interface becomes accessible. When a catheter fails this verification process, the generator displays a Red X indicator on the screen without accompanying error codes, effectively disabling catheter functionality. However, BD has determined that software version 3.35 generates false positive failures due to temperature-dependent verification parameters, incorrectly flagging properly functioning Venclose EVSRF Ablation Catheters as defective when the catheter temperature is not at steady state during the check initiation.

    • Ongoing Z-0371-2025 FDA Class II initiated 4 Sep 2024

    The Venclose" RF Ablation Catheters may have been manufactured with internal wiring error. The red and yellow signal wires were soldered on swapped ports of the catheter PCB. This error in wire configuration would lead to power being applied through the proximal 3/4th of the heating coil, bypassing the thermocouple, which will then cause the generator to overdrive the catheter in an attempt to reach desired temperature set point. The wire attachment error causes the device to reach temperatures exceeding the levels intended for therapeutic efficacy, while the console is falsely indicating that temperature has not yet been reached, thus increasing the risk of harm due to excessive temperature.

Timeline

4 dated events

  1. Cleared FDA 510(k) K160754
  2. Listed FDA GUDID 00858254006046
  3. Recall initiated FDA recalls Z-0371-2025
  4. Recall initiated FDA recalls Z-2638-2025

Coverage

1 jurisdiction

United States of America
  • Active
  • Expiring
  • Withdrawn
  • Listed
  • Not covered

United States — FDA 510(k) Active

Specification

GUDID

Trade name
Venclose System Power Cord
Generic name
Electrical mains power cable
Model / reference
VCPCB
Primary DI / UDI-DI
00858254006046
510(k) Number
FDA product code
GEI
GMDN term
Electrical mains power cable
Device class
Class II
Regulation number
878.4400
GUDID distribution status
In Commercial Distribution

Similar devices in this category

Other devices filed under Electrical mains power cable.

Other devices from Venclose Inc

See all devices from Venclose Inc →

Properties

  • Prescription use
  • Lot / batch number
Other properties (13, all answered No)