Skip to content

Vercise Deep Brain Stimulation (DBS) System

by Boston Scientific

PMA P150031

Active United States

PMA

Valid until

No expiry

PMA approval is not time-limited

FDA recalls

3 ongoing

This device is named in the recall

Documents

PMA P150031↗

FDA record

Trust

Unverified

Registry data, not yet checked by deviCERTS. Source FDA PMA, synced 6 Oct 2026. What this means

Regulatory status

as of registry sync 6 Oct 2026

RegionAuthorizationNumberDateStatus
United StatesFDA PMAAPPR Approved NumberP150031 Date8 Dec 2017ApprovedStatusActive

Recalls

4 recalls · 3 ongoing · FDA

    • Ongoing Z-2558-2025 FDA Class II initiated 8 Jul 2025

    Wire break(s) have occurred in rechargeable deep brain stimulation implantable pulse generators, implanted in the submuscular pectoral location. Wire breaks prevent successful stimulation, thus requiring device removal/replacement and users may experience high monopolar impedances, undesired sensation, loss of therapy, return of symptoms and/or Bluetooth connectivity challenges.

    • Ongoing Z-2340-2025 FDA Class II initiated 8 Jul 2025

    Resistance and difficulties could occur while closing the retaining clip locking mechanism, which is part of a burr hole cover kit, which is part of a Deep Brain Stimulation (DBS) system, which may lead to procedural delay, or lead-migration-which may require additional surgery.

    • Ongoing Z-1890-2024 FDA Class II initiated 18 Apr 2024

    Deep Brain Stimulation (DBS) Implantable Pulse Generator (IPG) may experience routine system check during IPG charging, which may cause device reset. The device reset could lead transient loss of stimulation; patients may experience undesired sensations, transient worsening of movement disorder symptoms, which may lead patient to request surgical intervention for replacement or revision.

    • Terminated Z-2125-2020 FDA Class II initiated 24 Apr 2019

    The firm is initiating a voluntary removal of certain spinal cord stimulation leads, lead extensions, adapters, and deep brain stimulation lead extensions due to the "use by date" displayed on the label which is later than the correct "use by date". No adverse health consequence is expected to occur from this issue.

Coverage

1 jurisdiction

United States of America
  • Active
  • Expiring
  • Withdrawn
  • Listed
  • Not covered

United States — FDA PMA Active

Specification

PMA

Full name
Vercise Deep Brain Stimulation (DBS) System and Vercise Gevia Deep Brain Stimulation (DBS) System
Trade name
Vercise Deep Brain Stimulation (Dbs)System
Generic name
Stimulator, electrical, implanted, for parkinsonian symptoms
FDA applicant
Boston Scientific Corp
PMA number
FDA product code
NHL
Device class
Class III
Advisory committee
Neurology
FDA decision
APPR Approved

Overview

AI-generated summary, 20 Sep 2026

Written by a language model from registry data. Not verified by deviCERTS or the manufacturer. The registry data above is the record.

The Vercise Deep Brain Stimulation (DBS) System is an implanted electrical stimulator classified as a Stimulator, Electrical, Implanted, for Parkinsonian Symptoms. It is designed for bilateral stimulation of the subthalamic nucleus (STN) to treat moderate to advanced levodopa-responsive Parkinson’s disease (PD) that is inadequately controlled by medication alone. This system delivers targeted neural modulation to help manage motor symptoms associated with the condition.

The device is supplied as a reusable, implanted system with associated leads and pulse generators, requiring surgical implantation. It is intended for use in clinical settings by trained neurosurgeons and neurologists. The system is MRI-conditional under specific conditions, allowing full-body MRI access in compatible models. Authorized under FDA_PMA (P150031), it adheres to Class III regulatory standards and is distributed with detailed surgical and programming instructions for clinical use.

Other devices from Boston Scientific

See all devices from Boston Scientific →

Questions

AI-generated answers, 20 Sep 2026

Written by a language model from registry data. Not verified by deviCERTS or the manufacturer. The registry data above is the record.

  • What specific symptoms of Parkinson’s disease does the Vercise DBS System target, and how does it work to address them?

    The Vercise DBS System is designed to target motor symptoms associated with moderate to advanced levodopa-responsive Parkinson’s disease (PD) that are not adequately controlled by medication alone. It delivers bilateral electrical stimulation to the subthalamic nucleus (STN), a brain region involved in regulating movement. By modulating neural activity in the STN, the system helps manage tremors, rigidity, bradykinesia (slowness of movement), and other motor fluctuations commonly experienced by patients with PD.

  • Is the Vercise DBS System reusable or single-use, and what does the surgical process involve?

    The Vercise DBS System is reusable and intended for implantation, meaning it is surgically placed under the skin and connected to leads targeting the STN. The system includes pulse generators and associated leads, which require professional implantation by trained neurosurgeons and neurologists. The procedure involves precise placement of electrodes in the brain, followed by connection to the implanted stimulator, which is then programmed to deliver targeted therapy.

  • Can patients with this device undergo MRI scans, and if so, what conditions apply?

    Yes, certain models of the Vercise DBS System are MRI-conditional, allowing for full-body MRI access under specific conditions. However, compatibility depends on the exact model and settings, so it’s critical to verify with the manufacturer or clinical guidelines before scheduling an MRI. Always follow the device’s programming and safety instructions to avoid interference or damage.

  • How is the Vercise DBS System regulated, and what does this mean for clinicians or facilities using it?

    The Vercise DBS System is classified as a Class III medical device under FDA regulations and has received Premarket Approval (PMA) under the designation P150031. This means it has undergone rigorous clinical evaluation to demonstrate safety and effectiveness for its intended use—bilateral STN stimulation in Parkinson’s disease patients unresponsive to medication alone. Clinicians and facilities must adhere to FDA guidelines, including proper training, surgical protocols, and programming instructions provided by the manufacturer.