VERIQA RT MonteCarlo 3D v3
by PTW-Freiburg Physikalisch-Technische Werkstätten Dr. Pychlau GmbH
Listed
Valid until
Not applicable
FDA recalls
2 ongoingDocuments
None
Trust
Unverified
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | GUDID listing | NumberEPTWS070031Q5200 | Date17 Apr 2026 | StatusListed |
Recalls
-
- Ongoing Z-3174-2026
When a bolus is included in a treatment plan and sent for secondary dose calculation, the software accounts for it but does not display this clearly, which may lead users to incorrectly omit the bolus during treatment and deliver the prescribed dose to the wrong tissue depth.
-
- Ongoing Z-1181-2024
Software: If the user excludes voxels from the Gamma calculation that are below a dose threshold (Suppress gamma calculation), the Gamma Passing Rate (GPR) calculated for individual ROIs is not correct. The GPR calculated by VERIQA overestimates the correct GPR. The evaluation can therefore show false positive results.
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — GUDID listing Listed
Specification
- Trade Name
- Veriqa
- Generic Name
- Radiation therapy software
- Model / Ref / Catalog
- VERIQA RT MonteCarlo 3D v3, S070031.520
- Primary DI / UDI-DI
- EPTWS070031Q5200
- 510(k) Number
- FDA Product Code
- IYE
- GMDN Term
- Radiation therapy software
- Device Class
- Class II
- Regulation Number
- 892.5050
- GUDID distribution status
- In Commercial Distribution
Other devices from PTW-Freiburg Physikalisch-Technische Werkstätten Dr. Pychlau GmbH
See all devices from PTW-Freiburg Physikalisch-Technische Werkstätten Dr. Pychlau GmbH →
Properties
Other properties (14)
- Sterile: No
- Single Use: No
- Implantable: No
- Contains Latex: No
- Prescription Use: No
- Over-the-Counter: No
- Convenience Kit: No
- Combination Product: No
- HCT/P: No
- Premarket Exempt: No
- Lot / Batch Number: No
- Serial Number: No
- Expiration Date: No
- Manufacturing Date: No