Visera Cysto-Nephro Videoscope Olympus Cyf Type V2R
by Olympus Medical Systems Corporation United States GUDID
Product code FAJ
Regulatory status
Recall ongoing. 5 ongoing FDA recalls. See recalls
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | GUDID listing | Number04953170411212 | Date29 Jul 2019 | StatusIn distribution |
Coverage
- Listed
- Not covered
United States — GUDID listing Listed
Recalls
-
- Ongoing Z-0689-2024
- Ongoing Z-0691-2024
- Ongoing Z-0690-2024
- Ongoing Z-0692-2024
Olympus has updated the instructions for use: Reusable cleaning brushes have been removed from the manual cleaning. Single use brushes remain compatible.
-
- Ongoing Z-0331-2024
Channel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Water remaining in the channel of the endoscopes has the potential for contamination that could lead to the risk of infection
Specification
- Trade Name
- Visera Cysto-Nephro Videoscope
- Generic Name
- Flexible video cystonephroscope, reusable
- Model / Ref / Catalog
- OLYMPUS CYF TYPE V2R, N6005830
- Primary DI / UDI-DI
- 04953170411212
- 510(k) Number
- FDA Product Code
- FAJ
- GMDN Term
- Flexible video cystonephroscope, reusable
- Device Class
- Class II
- Regulation Number
- 876.1500
- Commercial Distribution
- In Commercial Distribution
Other devices from Olympus Medical Systems Corporation
Properties & raw data
Show properties, registry sources and raw records
Other properties (11)
- Sterile: No
- Single Use: No
- Implantable: No
- Contains Latex: No
- Over-the-Counter: No
- Convenience Kit: No
- Combination Product: No
- HCT/P: No
- Premarket Exempt: No
- Lot / Batch Number: No
- Expiration Date: No
Sources (1)
- FDA UDI caee2631-d3b7-45b2-96ad-5471a1b5a15d