Skip to content

Vision Fx 40 & 80 Series Gamma Camera

by Summit Nuclear

510(k) K935273

Active United States

510(k)

Valid until

No expiry

510(k) clearance is not time-limited

FDA recalls

None ongoing

2 closed

Documents

Trust

Unverified

Registry data, not yet checked by deviCERTS. Source FDA 510(k), synced 11 Oct 2026. What this means

Regulatory status

as of registry sync 11 Oct 2026

RegionAuthorizationNumberDateStatus
United StatesFDA 510(k)SESE Substantially Equivalent · Traditional NumberK935273 Date18 Apr 1994ClearedStatusActive

Recalls

2 recalls · FDA

  • 2 recalls, same reason

    • Terminated Z-1394-04 FDA Unclassified (pre-2012) initiated 30 Jul 2004
    • Terminated Z-1393-04 FDA Unclassified (pre-2012) initiated 30 Jul 2004

    The detector head of a FX-80 Nuclear camera may fall from the gantry. Three potential causes have been determined. (1) A ball nut in the detector's radial drive assembly can back out of the bearing block that contains it. (2) Misalignment of the radial drive assembly. (3) Mounting bolts on the radial drive assembly may become loose, fail, or back out.

Coverage

1 jurisdiction

United States of America
  • Active
  • Expiring
  • Withdrawn
  • Listed
  • Not covered

United States — FDA 510(k) Active

Specification

510(k)

510(k) Number
FDA Product Code
IYX
Device Class
Class I
Regulation
21 CFR Part 892.1100
Advisory Committee
Radiology
FDA Decision
SESE Substantially Equivalent

Other devices from Summit Nuclear