Vision Fx 40 & 80 Series Gamma Camera
510(k) K935273
Active
510(k)
Valid until
No expiry
FDA recalls
None ongoingDocuments
Trust
Unverified
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK935273 | Date18 Apr 1994 | StatusActive |
Recalls
-
- Terminated Z-1394-04
- Terminated Z-1393-04
The detector head of a FX-80 Nuclear camera may fall from the gantry. Three potential causes have been determined. (1) A ball nut in the detector's radial drive assembly can back out of the bearing block that contains it. (2) Misalignment of the radial drive assembly. (3) Mounting bolts on the radial drive assembly may become loose, fail, or back out.
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA Product Code
- IYX
- Device Class
- Class I
- Regulation
- 21 CFR Part 892.1100
- Advisory Committee
- Radiology
- FDA Decision
- SESE Substantially Equivalent