VistaPhone 8303-01
VistaPhone 8303-01 is an electrocardiography telemetric monitoring system (FDA Class II; FDA product code DRG, Transmitters And Receivers, Physiological Signal, Radiofrequency) made by VitalConnect, Inc., cleared under FDA 510(k) K193343 on 20 Apr 2020 and listed in FDA GUDID since 2 Nov 2020 (UDI-DI 00850490007436), active as of 22 Oct 2026. One FDA recall is linked to it.
Active
Valid until
No expiry
FDA recalls
None ongoingDocuments
Trust
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | GUDID listing | Number00850490007436 | Date2 Nov 2020 | StatusListed |
| United States | FDA 510(k)SESE | NumberK193343 | Date20 Apr 2020 | StatusActive |
Official records
- AccessGUDID record 00850490007436
- FDA 510(k) database K193343
From the FDA record
- Device description
- Vista Phone relay device
- MRI safety
- Labeling does not contain MRI Safety Information
- Natural rubber latex
- No latex statement on the label
- Sterility
- Supplied non-sterile
Across registries
- FDA 510(k) K193343 decided 20 Apr 2020 Active
- FDA GUDID UDI-DI 00850490007436 listed 2 Nov 2020 Listed
Recalls
-
- Terminated Z-2537-2020
A healthcare provider (HCP) noticed a patient had a lowered historic SpO2 reading that the software should have triggered a notification to the care team, but no notification was found. A software defect was encountered causing the lack of notification. The defect could result in exposing a patient to potential hypoxia for a short window of time. The firm is correcting the software to bring the performance back into compliance with specifications.
Timeline
- Cleared FDA 510(k) K193343
- Recall initiated FDA recalls Z-2537-2020
- Listed FDA GUDID 00850490007436
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- Trade name
- VistaPhone
- Generic name
- Electrocardiography telemetric monitoring system
- Model / Ref / Catalog
- 8303-01
- Primary DI / UDI-DI
- 00850490007436
- 510(k) Number
- FDA product code
- DRG
- GMDN term
- Electrocardiography telemetric monitoring system
- Device class
- Class II
- Regulation number
- 870.2910
- GUDID distribution status
- In Commercial Distribution
Similar devices
Other devices filed under FDA product code DRG (Transmitters And Receivers, Physiological Signal, Radiofrequency).
- FAMILY HEALTHWARE by Sanitas, Inc.
- 3M LITTMANN SCOPE-TO SCOPE SOFTWARE SYSTEM by 3M COMPANY
- ECG TELEMETRY SYSTEM by Mennen Greatbatch Electonics
- AO USB RECEIVER, WI-BOX, WI-BOX PSU KIT, WI-BOX XPRESS CABLE, PW USB RECEIVER by St. Jude Medical Systems AB
- TelePatch Cardiac Monitor by MEDICOMP, INC.
- NIHON KOHDEN ZB-821 TELEMETRY TRANSMITTER by NIHON KOHDEN AMERICA, INC.
- Perin Health System (PHD80060-2) by Perin Health Devices, LLC
- MOBILECARE MONITOR by Aframe Digital, Inc.
- Ossler Exam Scope by Lr Med, Inc.
- RTX3370 by Rtx Healthcare A/S