VistaTablet 8203-01
VistaTablet 8203-01 is an electrocardiography telemetric monitoring system (FDA Class II; FDA product code DRG, Transmitters And Receivers, Physiological Signal, Radiofrequency) made by VitalConnect, Inc., cleared under FDA 510(k) K193343 on 20 Apr 2020 and listed in FDA GUDID since 2 Nov 2020 (UDI-DI 00850490007429), active as of 22 Oct 2026. One FDA recall is linked to it.
Active
Valid until
No expiry
FDA recalls
None ongoingDocuments
Trust
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | GUDID listing | Number00850490007429 | Date2 Nov 2020 | StatusListed |
| United States | FDA 510(k)SESE | NumberK193343 | Date20 Apr 2020 | StatusActive |
Official records
- AccessGUDID record 00850490007429
- FDA 510(k) database K193343
From the FDA record
- Device description
- VistaTablet Relay Device
- MRI safety
- Labeling does not contain MRI Safety Information
- Natural rubber latex
- No latex statement on the label
- Sterility
- Supplied non-sterile
Across registries
- FDA 510(k) K193343 decided 20 Apr 2020 Active
- FDA GUDID UDI-DI 00850490007429 listed 2 Nov 2020 Listed
Recalls
-
- Terminated Z-2537-2020
A healthcare provider (HCP) noticed a patient had a lowered historic SpO2 reading that the software should have triggered a notification to the care team, but no notification was found. A software defect was encountered causing the lack of notification. The defect could result in exposing a patient to potential hypoxia for a short window of time. The firm is correcting the software to bring the performance back into compliance with specifications.
Timeline
- Cleared FDA 510(k) K193343
- Recall initiated FDA recalls Z-2537-2020
- Listed FDA GUDID 00850490007429
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- Trade name
- VistaTablet
- Generic name
- Electrocardiography telemetric monitoring system
- Model / Ref / Catalog
- 8203-01
- Primary DI / UDI-DI
- 00850490007429
- 510(k) Number
- FDA product code
- DRG
- GMDN term
- Electrocardiography telemetric monitoring system
- Device class
- Class II
- Regulation number
- 870.2910
- GUDID distribution status
- In Commercial Distribution
Similar devices
Other devices filed under FDA product code DRG (Transmitters And Receivers, Physiological Signal, Radiofrequency).
- BL Healthcare Remote Care Management System by BL HEALTHCARE, INC.
- FFR Link, FFR Signal Processing Module by Boston Scientific
- TELEPHONE BASED BL HEALTHCARE REMOTE CARE MANAGEMENT SYSTEM (RCMS) by BL HEALTHCARE, INC.
- Intouch Remote Presence Robotic System RP-7 by Intouch Technologies
- Patient Status Engine by Isansys Lifecare, Ltd.
- FORA GW 9014 Telehealth GATEWAY/TD-9014 Telehealth Gateway by TAIDOC TECHNOLOGY CORPORATION
- PressureWire Aeris Agile Tip; PressureWire Aeris Agile Tip 300 by St. Jude Medical Systems AB
- Propaq ECG Telemetry Option by Protocol Systems, Inc.
- Care Pal, Model CPW-10X by Digio2 International Co., Ltd.
- LifeStream Management System by Honeywell Hommed, LLC