# Visualase Cooled Laser Applicator System, Laser Diffusing Fiber, Cooling Catheter System

> Active: United States. Medical device record on deviCERTS, compiled from public registries.

- Canonical page: <https://devicerts.com/devices/visualase-cooled-laser-applicator-system-laser-diffusing-fib-gex>
- Manufacturer: [Biotex, Inc.](https://devicerts.com/manufacturers/biotex-inc) (United States, FDA 510(k) applicant)
- Name in the registry: VISUALASE COOLED LASER APPLICATOR SYSTEM, LASER DIFFUSING FIBER, COOLING CATHETER SYSTEM
- Device class: Class II (US FDA)
- FDA product code: GEX

## Registrations and authorizations

- **510(k)** · FDA 510(k) · United States · K053087 · decision 2006-03-01 · decision SESE · status: Active

## Recalls

- Z-0244-2023 · Class II · ongoing · initiated 2022-10-14 · matched through shared submission K053087, may concern other models: Due to a defect in the outer pouch sterile seal
- Z-2501-2018 · Class II · terminated · initiated 2018-03-23: Medtronic has become aware that certain lots of the VCLAS 3MM, 10MM, and 15MM laser diffusing fibers were not tested at intended laser power levels, due to an issue related to calibration of the test equipment, which can lead to the potential for low laser energy output from fiber or fiber burnout at lower wattage.
- Z-0025-2018 · Class II · terminated · initiated 2017-07-27: The VCLAS 3mm, 10mm, and 15mm tip devices were not appropriately tested for saline leakage during the manufacturing process.
- Z-2755-2017 · Class II · terminated · initiated 2017-06-07: Medtronic has become aware that the VCLAS 15MM TIP cap, also referred to as Tuohy Borst Adapter (TBA), may not be capable of creating a seal around the optical fiber to ensure adequate saline flow through the cooling system. When the TBA is unable to seal around the optical fiber, low retention force to the optical fiber may occur, causing saline to leak out of the TBA.
- Z-1672-2016 · Class II · terminated · initiated 2016-05-09: Potential for the Visualase Cooled Laser Applicator System (VCLAS) catheters to melt during procedures.

## Identifiers

- 510(k) Number: K053087 [FDA 510(k)]
- FDA product code: GEX [FDA 510(k)]

## Classification

- Device class: Class II [FDA 510(k)]
- Regulation: 21 CFR Part 878.4810 [FDA 510(k)]
- Advisory committee: General, Plastic Surgery [FDA 510(k)]
- FDA decision: SESE [FDA 510(k)]

## Source and citation

- Cite as: Visualase Cooled Laser Applicator System, Laser Diffusing Fiber, Cooling Catheter System, Biotex, Inc. deviCERTS medical device catalogue. https://devicerts.com/devices/visualase-cooled-laser-applicator-system-laser-diffusing-fib-gex (data as of 2026-11-19).
- Data sources: [FDA 510(k)](https://open.fda.gov/apis/device/510k/), [FDA recalls](https://open.fda.gov/apis/device/recall/)
- Registry records last synced: 2026-10-22
- Statuses computed as of: 2026-11-19
- JSON: <https://devicerts.com/devices/visualase-cooled-laser-applicator-system-laser-diffusing-fib-gex.json>
- Licence: [CC BY 4.0](https://creativecommons.org/licenses/by/4.0/). The deviCERTS layout and text are licensed under CC BY 4.0. Registry data remain under their source terms: U.S. FDA data (openFDA, AccessGUDID) are U.S. public domain; EUDAMED data are subject to the European Commission reuse terms.
- Attribution: DevicErts, https://devicerts.com/devices/visualase-cooled-laser-applicator-system-laser-diffusing-fib-gex
- Compiled by deviCERTS (devicerts.com) from public medical device registries (EUDAMED, European Commission; U.S. FDA via openFDA; AccessGUDID, U.S. National Library of Medicine). deviCERTS is not an official register and gives no regulatory advice: the record in the registry itself prevails.