Vital Signs Monitor; Advisor Model # BCI 9200
510(k) K982279
Active
Valid until
No expiry
FDA recalls
None ongoingDocuments
Trust
Unverified
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK982279 | Date25 Sep 1998 | StatusActive |
Recalls
-
- Terminated Z-1326-2009
Recent improvements were made to the Advisor case plastic resulting in the inserts being deeply embedded into the plastic. This affects the ability of the monitor to be adequately secured on the wall mount bracket or roll stand bracket, resulting in the monitor falling off. There have been no adverse events or injuries reported.
-
- Terminated Z-1089-2007
Turn off unexpectedly/ Not turn on : monitors turned off unexpectedly sounding the critical failure alarm. Advisor Vital Signs monitors that were off, would not turn back on.
-
- Terminated Z-0229-06
- Terminated Z-0239-06
- Terminated Z-0221-06
- Terminated Z-0235-06
- Terminated Z-0223-06
- Terminated Z-0230-06
- Terminated Z-0231-06
- Terminated Z-0225-06
- Terminated Z-0227-06
- Terminated Z-0232-06
- Terminated Z-0228-06
- Terminated Z-0224-06
- Terminated Z-0238-06
- Terminated Z-0234-06
- Terminated Z-0237-06
- Terminated Z-0220-06
- Terminated Z-0233-06
- Terminated Z-0222-06
- Terminated Z-0240-06
- Terminated Z-0236-06
- Terminated Z-0226-06
Potential Safety related problem associated with the pace detect function on some Advisor monitors that have impedance respiration capabilities in that random pace markers may appear when the pace detect function is turned on and the primary lead is set to Lead II, III or any of the augmented leads, AVL, AVR or AVF on the Advisor monitor.
-
- Terminated Z-0828-05
- Terminated Z-0830-05
- Terminated Z-0824-05
- Terminated Z-0831-05
- Terminated Z-0827-05
- Terminated Z-0835-05
- Terminated Z-0834-05
- Terminated Z-0832-05
- Terminated Z-0826-05
- Terminated Z-0829-05
- Terminated Z-0833-05
- Terminated Z-0825-05
Altitude testing of the Advisor with Capnography module found the CO2 (End Tidal and Inspired) values are incorrect when the Advisor is configured to display in the relative units of percent (%) at elevations above 4,100 feet (at standard temperature pressure). The % CO values displayed will be lower than the actual CO2 value.
-
- Terminated Z-0533-05
- Terminated Z-0534-05
- Terminated Z-0531-05
- Terminated Z-0535-05
- Terminated Z-0532-05
An issue with the pace detect function was discovered on a 9200 II monitor. Could result in a falsely high or low heart rate reading.
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA Product Code
- MHX
- Models / Variants
- Device Class
- Class II
- Regulation
- 21 CFR Part 870.1025
- Advisory Committee
- Cardiovascular
- FDA Decision
- SESE Substantially Equivalent
Overview
AI-generated summary, 1 Oct 2026
The Vital Signs Monitor; Advisor Model # BCI 9200 is a Physiological Patient Monitor (with Arrhythmia Detection or Alarms) designed to continuously assess multiple vital signs. It provides real-time monitoring of electrocardiogram (ECG), respiration, non-invasive blood pressure, pulse oximetry, invasive pressures, and temperature, enabling comprehensive patient assessment in clinical settings.
This device is intended for use in hospital or clinical environments, offering a portable, tabletop design for multi-parameter monitoring. Supplied as a reusable medical device, it requires appropriate sterilization and maintenance protocols. The BCI 9200 has received FDA 510K clearance under the Cardiovascular Advisory Committee and is classified as a Class II device under regulation number 870.1025. It is not intended for MRI use and must be stored in accordance with manufacturer guidelines.
Other devices from Bci Intl., Inc.
Questions
AI-generated answers, 1 Oct 2026
What is the Vital Signs Monitor Advisor Model # BCI 9200 used for?
The Vital Signs Monitor Advisor Model # BCI 9200 is designed for continuous assessment of multiple vital signs including electrocardiogram (ECG), respiration, non-invasive blood pressure, pulse oximetry, invasive pressures, and temperature, providing comprehensive patient monitoring in hospital or clinical environments.
How is the BCI 9200 supplied and is it reusable?
The BCI 9200 is supplied as a reusable medical device with a portable, tabletop design. It requires appropriate sterilization and maintenance protocols according to manufacturer guidelines for continued safe and effective use in clinical settings.
What regulatory status does the BCI 9200 have?
The BCI 9200 has received FDA 510K clearance under the Cardiovascular Advisory Committee and is classified as a Class II device under regulation number 870.1025. It is not intended for MRI use and must be stored according to manufacturer guidelines.
What parameters does the BCI 9200 monitor?
The BCI 9200 provides real-time monitoring of electrocardiogram (ECG), respiration, non-invasive blood pressure, pulse oximetry, invasive pressures, and temperature, enabling comprehensive patient assessment in clinical settings with arrhythmia detection capabilities.