Vitek Gram Positive Identification Card (Gpi)
510(k) K952095
Active
Valid until
No expiry
FDA recalls
None ongoingDocuments
Trust
Unverified
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK952095 | Date12 Mar 1996 | StatusActive |
Recalls
-
- Terminated Z-0074-2018
Customers have reported, and an internal investigation has confirmed, occurrences of atypical negative O129R reactions for Enterococcus casseliflavus ATCC)R) 700327(TM) in association with three (3) VITEK(R) 2 GP ID Lots, causing failure of the Quality Control (QC) testing for the VITEK(R) 2 GP ID card. In accordance with the VITEK(R) 2 GP ID Instructions For Use, the expected O129R reaction for ATCC(R) 700327(TM) strain is positive. As a precaution, bioMrieux is also including in the scope of this notice all GP ID card lots manufactured with the same raw materials.
-
- Terminated Z-3134-2017
Customer reports have indicated occurrences of atypical negative O129R reactions for Enterococcus casseliflavus ATCC(R) 700327(TM) in association with VITEK(R) 2 GP ID Lot 2420240403 and 2420303403, causing failure of the Quality Control (QC) testing for the VITEK(R) 2 GP ID card. The expected O129R reaction for ATCC 700327(TM) strain is positive.
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA Product Code
- LQL
- Device Class
- Class I
- Regulation
- 21 CFR Part 866.2660
- Advisory Committee
- Microbiology
- FDA Decision
- SESE Substantially Equivalent