Skip to content

Vitek® 2 Compact 27415

by Biomerieux, Inc.

Ref 27415

Vitek® 2 Compact 27415 is a device classified as Class A in the EU and Class II in the US made by Biomerieux, Inc., registered in EUDAMED on 21 Jul 2026 (Basic UDI-DI 357302BUDI000560TM), cleared under FDA 510(k) K050002 on 17 Feb 2005 and listed in FDA GUDID since 14 Sep 2016 (UDI-DI 03573026220624), active as of 25 Oct 2026. 5 FDA recalls are linked to it.

Active European Union and United States. Also listed in GUDID (US).

CE IVDR 510(k)

Valid until

No date given

CE: no date in EUDAMED. FDA: not time-limited

FDA recalls

None ongoing

6 closed

Documents

None linked in EUDAMED

A group certificate may exist

510(k) K050002

FDA submission, from GUDID

Trust

Unverified

Source EUDAMED, GUDID, synced 25 Oct 2026.

Regulatory status

as of registry sync 25 Oct 2026

RegionAuthorizationNumberDateStatus
European UnionCE/IVDR EUDAMED Number357302BUDI000560TM Basic UDI-DI 03573026220624UDI-DIDate21 Jul 2026RegisteredStatusActive
United StatesGUDID listing GUDID Number03573026220624 UDI-DI Date14 Sep 2016ListedStatusListed GUDID: in commercial distribution
United StatesFDA 510(k)SESE GUDID from GUDID Substantially Equivalent · Traditional NumberK050002 Date17 Feb 2005DecidedStatusActive

Official records

Across registries

3 registries list this device

  • EUDAMED Basic UDI-DI 357302BUDI000560TM registered 21 Jul 2026 Active
  • FDA 510(k) K050002 decided 17 Feb 2005 Active
  • FDA GUDID UDI-DI 03573026220624 listed 14 Sep 2016 Listed

Recalls

6 recalls · FDA

  • 2 recalls, same reason

    • Terminated Z-1307-2021 FDA Class II initiated 28 Jan 2021
    • Terminated Z-1308-2021 FDA Class II initiated 28 Jan 2021

    bioMrieux has identified a potential safety risk worst case of a false susceptible erroneous test result associated with this event. The problem is when HL7 Connection is used, the results in the VITEK 2 do not match the results sent to the LIS and the Laboratory Technician would need to change results in the LIS to match those in the VITEK 2. If the isolate is sent more than once to LIS, a software defect prevents the system from sending the expertised interpretation results and the Therapeutic Corrections (TC) is not sent to Laboratory Information Systems (LIS).

    • Terminated Z-0639-2019 FDA Class II initiated 13 Nov 2018

    The VITEK 2 Systems Software Version 9.01 software may display incorrect organism identification information as well as the associated Antimicrobial Susceptibility Testing (AST) results when using the VITEK 2 FLEXPREP" feature.

  • 2 recalls, same reason

    • Terminated Z-3075-2017 FDA Class II initiated 12 Jul 2017
    • Terminated Z-3076-2017 FDA Class II initiated 12 Jul 2017

    Customers have reported that some VITEK 2 cards are staying in preliminary status, not finalizing after ejection from the instrument, and not allowing cards in subsequent carousel slots to be processed. The issue was reported to occur on VITEK 2 Compact 15 and Compact 30 systems following a system software update to version 8.01.

    • Terminated Z-0991-2013 FDA Class II initiated 18 Feb 2013 on shared 510(k) K050002

    There is a potential to link an isolate result to the wrong patient and then upload the results to the LIS.

Recalls on a shared submission name another device cleared under the same FDA submission as this one.

Timeline

9 dated events

  1. Cleared FDA 510(k) K050002
  2. Recall initiated FDA recalls Z-0991-2013
  3. Listed FDA GUDID 03573026220624
  4. Recall initiated FDA recalls Z-3075-2017
  5. Recall initiated FDA recalls Z-3076-2017
  6. Recall initiated FDA recalls Z-0639-2019
  7. Recall initiated FDA recalls Z-1307-2021
  8. Recall initiated FDA recalls Z-1308-2021
  9. Registered EUDAMED 357302BUDI000560TM

Coverage

2 jurisdictions

United States of AmericaSwedenFinlandLatviaEstoniaDenmarkLithuaniaIrelandNetherlandsGermanyPolandFranceBelgiumLuxembourgCzech RepublicSlovakiaAustriaHungaryRomaniaSloveniaBulgariaPortugalSpainItalyCroatiaGreeceMaltaCyprus
  • Active
  • Expiring
  • Withdrawn
  • Listed
  • Not covered
  • European Union CE/IVDR Active
  • United States FDA 510(k) Active

Specification

EUDAMED · GUDID

Trade name
Vitek® 2 Compact
Generic name
Microorganism identification/antimicrobial-susceptibility analyser IVD
Model / reference
27415
Primary DI / UDI-DI
03573026220624
Basic UDI-DI
357302BUDI000560TM
510(k) Number
FDA product code
LON
EMDN code
W0203010180 Antibiotic susceptibility analysers - hardware accessories
GMDN term
Microorganism identification/antimicrobial-susceptibility analyser IVD
Device class
Class A EUDAMED
Class II GUDID
Regulation number
866.1645
EU registration type
Standard device
EU market of registration
France
GUDID distribution status
In Commercial Distribution

Truth per registry

2 sources · 1 field differs

FieldEUDAMEDGUDID
Device classEUDAMED
Class A
GUDID
Class II

Device class is stated on each registry's own scale: Class A in the EU (EUDAMED), Class II in the US (FDA GUDID). EU and US classes are different schemes, so this is not a disagreement.

Similar devices in this category

Other devices filed under ANTIBIOTIC SUSCEPTIBILITY ANALYSERS.

Other devices from Biomerieux, Inc.

See all devices from Biomerieux, Inc. →

Properties

  • Professional testing
  • IVD instrument
  • Prescription use
  • Serial number
Other properties (22, all answered No)