  "format": "devicerts.device.v1",
  "cite_as": "VNS Therapy®, LivaNova USA, Inc. deviCERTS medical device catalogue. https://devicerts.com/devices/vns-therapy-r-1000-d (data as of 2026-11-24).",
  "source": {
    "canonical_url": "https://devicerts.com/devices/vns-therapy-r-1000-d",
    "json_url": "https://devicerts.com/devices/vns-therapy-r-1000-d.json",
    "markdown_url": "https://devicerts.com/devices/vns-therapy-r-1000-d.md",
    "publisher": "deviCERTS (devicerts.com)",
    "data_sources": [
      {
        "registry": "EUDAMED",
        "url": "https://ec.europa.eu/tools/eudamed"
      },
      {
        "registry": "FDA GUDID (AccessGUDID)",
        "url": "https://accessgudid.nlm.nih.gov/"
      },
      {
        "registry": "FDA recalls",
        "url": "https://open.fda.gov/apis/device/recall/"
      }
    ],
    "registry_records_last_synced": "2026-10-22",
    "status_as_of": "2026-11-24",
    "license": "CC BY 4.0",
    "license_url": "https://creativecommons.org/licenses/by/4.0/",
    "license_scope": "The deviCERTS layout and text are licensed under CC BY 4.0. Registry data remain under their source terms: U.S. FDA data (openFDA, AccessGUDID) are U.S. public domain; EUDAMED data are subject to the European Commission reuse terms.",
    "attribution": "DevicErts, https://devicerts.com/devices/vns-therapy-r-1000-d",
    "disclaimer": "Compiled by deviCERTS (devicerts.com) from public medical device registries (EUDAMED, European Commission; U.S. FDA via openFDA; AccessGUDID, U.S. National Library of Medicine). deviCERTS is not an official register and gives no regulatory advice: the record in the registry itself prevails."
  },
  "device": {
    "name": "VNS Therapy®",
    "registry_name": "VNS Therapy®",
    "url": "https://devicerts.com/devices/vns-therapy-r-1000-d",
    "manufacturer": {
      "name": "LivaNova USA, Inc.",
      "url": "https://devicerts.com/manufacturers/livanova-usa-inc",
      "country": "United States",
      "country_basis": null,
      "fda_registration_country": null
    },
    "status": {
      "state": "active",
      "label": "Active",
      "detail": "European Union and United States"
    },
    "classification": [
      {
        "class": "Active implantable (AIMDD)",
        "system": "EU AIMDD"
      },
      {
        "class": "Class III",
        "system": "US FDA"
      }
    ],
    "identifiers": [
      {
        "label": "Primary DI / UDI-DI",
        "values": [
          {
            "value": "05425025750528",
            "registries": [
              "EUDAMED",
              "FDA GUDID (AccessGUDID)"
            ]
          }
        ]
      },
      {
        "label": "Basic UDI-DI",
        "values": [
          {
            "value": "B-05425025750528",
            "registries": [
              "EUDAMED"
            ]
          }
        ]
      },
      {
        "label": "PMA number",
        "values": [
          {
            "value": "P970003",
            "registries": [
              "FDA GUDID (AccessGUDID)"
            ]
          }
        ]
      }
    ],
    "product_codes": [
      "MUZ",
      "LYJ"
    ],
    "emdn": [
      {
        "code": "J020301",
        "term": "DRUG-RESISTANT EPILEPSY NON-SURGICAL THERAPY IMPLANTABLE NEUROSTIMULATORS"
      },
      {
        "code": "J020399",
        "term": "VAGAL IMPLANTABLE NEUROSTIMULATORS - OTHER"
      }
    ],
    "gmdn": [
      {
        "code": "34210",
        "term": "Antiseizure/psychiatric-therapy vagus nerve implantable electrical stimulation system"
      }
    ],
    "description": null,
    "noindex": false
  },
  "authorizations": [
    {
      "registry": "EUDAMED",
      "region": "European Union",
      "type": "CE/AIMDD",
      "number": "B-05425025750528",
      "number_label": "Basic UDI-DI",
      "device_udi_di": "05425025750528",
      "date": "2026-05-06",
      "date_kind": "registered",
      "decision_code": null,
      "expires": null,
      "withdrawn": null,
      "status": "Active",
      "status_note": null,
      "derived_from_gudid": false
    },
    {
      "registry": "GUDID",
      "region": "United States",
      "type": "GUDID listing",
      "number": "05425025750528",
      "number_label": "UDI-DI",
      "device_udi_di": null,
      "date": "2020-07-22",
      "date_kind": "listed",
      "decision_code": null,
      "expires": null,
      "withdrawn": null,
      "status": "Listed",
      "status_note": "GUDID: in commercial distribution",
      "derived_from_gudid": false
    },
    {
      "registry": "GUDID",
      "region": "United States",
      "type": "FDA PMA",
      "number": "P970003",
      "number_label": null,
      "device_udi_di": null,
      "date": "1997-07-16",
      "date_kind": "decision",
      "decision_code": "APPR",
      "expires": null,
      "withdrawn": null,
      "status": "Active",
      "status_note": null,
      "derived_from_gudid": true
    }
  ],
  "no_longer_listed_by": [],
  "certificates": [],
  "recalls": [
    {
      "number": "Z-1068-2025",
      "classification": "Class II",
      "status": "ongoing",
      "initiated": "2025-01-03",
      "reason": "Their is a potential that implanted pulse generators may stop delivering therapy due to an internal component issue.",
      "applies": "this device",
      "via": "P970003"
    },
    {
      "number": "Z-1072-2025",
      "classification": "Class II",
      "status": "ongoing",
      "initiated": "2024-12-12",
      "reason": "Vagus nerve stimulator generators may stop delivering stimulation due to an internal, mechanically activated component that may become stuck in a closed position, which may result in patients returning to baseline seizure frequency or depressive symptoms and in the longer term, patients may be required to undergo generator replacement surgery to replace their device.",
      "applies": "this device",
      "via": "P970003"
    },
    {
      "number": "Z-0535-2021",
      "classification": "Class II",
      "status": "terminated",
      "initiated": "2020-11-12",
      "reason": "During internal testing, it was found that upon a device reset, the Generator exhibits the incorrect Model Number when interrogated.",
      "applies": "catalog number listed",
      "via": null
    },
    {
      "number": "Z-1067-2025",
      "classification": "Class II",
      "status": "ongoing",
      "initiated": "2025-01-03",
      "reason": "Their is a potential that implanted pulse generators may stop delivering therapy due to an internal component issue.",
      "applies": "this device",
      "via": "P970003"
    },
    {
      "number": "Z-0532-2022",
      "classification": "Class II",
      "status": "terminated",
      "initiated": "2021-12-02",
      "reason": "A manufacturing issue may cause the silicone tubing wall thickness of vagus nerve stimulation leads to be less than the design specification. If the tubing becomes completely breached, the patient may experience discomfort or pain due to delivery of stimulation to an unintended location. If the inner lead conductor is exposed to body fluids the lead may break causing loss of therapy.",
      "applies": "this device",
      "via": "P970003"
    },
    {
      "number": "Z-0462-2021",
      "classification": "Class II",
      "status": "terminated",
      "initiated": "2020-09-28",
      "reason": "Due to an extra digit being inadvertently added to the serial number of a subset of implantable pulse generators, they become not compatible with therapy programmers with software version 11.0.4.",
      "applies": "this device",
      "via": "P970003"
    },
    {
      "number": "Z-1563-2020",
      "classification": "Class II",
      "status": "ongoing",
      "initiated": "2020-02-04",
      "reason": "False positive warning may occur after: 1) VNS Generator interrogated at 0mA normal output current, 2) Generator programmed to non-0mA output current, 3) In-session re-interrogation performed. Users instructed to lower output current and widen pulse width. Only system diagnostic testing evaluates output current. Users may conclude device malfunction, could lead to medical/surgical intervention.",
      "applies": "this device",
      "via": "P970003"
    },
    {
      "number": "Z-0561-2020",
      "classification": "Class I",
      "status": "terminated",
      "initiated": "2019-08-22",
      "reason": "Sentiva generators may reset and become disabled within 60 days of enabling. Physicians can re-enable but devices will continue to be susceptible to resets. If a device experiences this issue, this may lead to additional surgery (premature replacement of generator), or patients may return to baseline seizure frequency or depressive symptoms due to cessation of intended stimulation.",
      "applies": "this device",
      "via": "P970003"
    },
    {
      "number": "Z-1118-2020",
      "classification": "Class II",
      "status": "terminated",
      "initiated": "2019-08-13",
      "reason": "Firm identified a subset of its generators that were sterilized one additional sterilization cycle, which does not meet the firm's quality specifications.",
      "applies": "this device",
      "via": "P970003"
    },
    {
      "number": "Z-0346-2020",
      "classification": "Class II",
      "status": "ongoing",
      "initiated": "2018-11-16",
      "reason": "Lead impedance values reported by the affected VNS generator will be higher compared to those reported by previous models. This is due to a change in the timing of when affected VNS generator takes the lead impedance measurement during diagnostic testing. As a result, normal impedance ranges for the affected VNS generator have shifted relative to the existing thresholds of 600 - 5300 Ohms defined in labeling and as present in the programming software.",
      "applies": "this device",
      "via": "P970003"
    },
    {
      "number": "Z-0663-2020",
      "classification": "Class II",
      "status": "ongoing",
      "initiated": "2018-11-16",
      "reason": "This recall is being initiated due to reports that that the therapy programming tablet with software version 1.5  errantly performs a normal mode diagnostic test instead of the selected system diagnostic test on Model 102 and Model 102R devices, if the output  current is greater than 0.5mA. This can result in false high impedance values during patient follow-up.",
      "applies": "this device",
      "via": "P970003"
    },
    {
      "number": "Z-0868-2020",
      "classification": "Class II",
      "status": "terminated",
      "initiated": "2018-11-16",
      "reason": "This recall is an expansion of Z-3019-2017 and Z-3020-2017, which was initiated to fix the devices premature battery depletion, caused by electrical leakage on the circuit board assemblies of the Models 105 and 106 generators.  *Note this recall occurred in November 2018.",
      "applies": "this device",
      "via": "P970003"
    },
    {
      "number": "Z-0869-2020",
      "classification": "Class II",
      "status": "terminated",
      "initiated": "2018-11-16",
      "reason": "This recall is an expansion of Z-3019-2017 and Z-3020-2017, which was initiated to fix the devices premature battery depletion, caused by electrical leakage on the circuit board assemblies of the Models 105 and 106 generators.  *Note this recall occurred in November 2018.",
      "applies": "this device",
      "via": "P970003"
    },
    {
      "number": "Z-2572-2018",
      "classification": "Class II",
      "status": "terminated",
      "initiated": "2018-01-24",
      "reason": "Unintended warning message displayed on generators programmed with a Model 3000 v.1.0.2.2 programmer.",
      "applies": "this device",
      "via": "P970003"
    },
    {
      "number": "Z-0555-2018",
      "classification": "Class II",
      "status": "terminated",
      "initiated": "2017-12-02",
      "reason": "Two Model 3000 Programmers were distributed in error by prior to FDA approval of version 1.0.2.2 software.",
      "applies": "this device",
      "via": "P970003"
    },
    {
      "number": "Z-2255-2018",
      "classification": "Class II",
      "status": "terminated",
      "initiated": "2017-11-13",
      "reason": "Certain Model 3000 programming events can result in miscalculation of parameters stored in the Models 103, 104, 105, and 106 generators.  During these programming events, the miscalculations can lead to: \n\"  Delivery of more stimulation than intended, resulting in painful stimulation or other common side effects (Model 106 only); \n\"  No stimulation in the case of device disablement (Burst Watchdog Timeout), resulting in no therapy to the patient (Model 106 only);\n\"  Delivery of less stimulation than intended, resulting in therapeutic settings not being achieved within device specification (Models 103, 104, 105, or 106); and/or \n\"  Delays or absence of the 75% and 50% battery life indicators displayed by the programming software (Models 103, 104, 105, or 106).",
      "applies": "this device",
      "via": "P970003"
    },
    {
      "number": "Z-0659-2016",
      "classification": "Class II",
      "status": "terminated",
      "initiated": "2015-12-18",
      "reason": "Recall being initiated in response to three reports of \"Burst Watchdog Timeout\" events occurring with the Model 106 AspireSR Generator, resulting in a device reset condition where stimulation output is disabled.",
      "applies": "this device",
      "via": "P970003"
    },
    {
      "number": "Z-0645-2016",
      "classification": "Class II",
      "status": "terminated",
      "initiated": "2015-11-18",
      "reason": "Certain Model 106 Pulse Generators demonstrate delays in sensing during use of the 'Verify Heartbeat Detection' feature and exhibit the potential for decreased battery longevity.",
      "applies": "this device",
      "via": "P970003"
    },
    {
      "number": "Z-0280-2016",
      "classification": "Class II",
      "status": "terminated",
      "initiated": "2015-10-07",
      "reason": "Certain Model 106 Pulse Generators demonstrate delays in sensing during use of the 'Verify Heartbeat Detection' feature and exhibit potential for decreased battery longevity.",
      "applies": "this device",
      "via": "P970003"
    },
    {
      "number": "Z-0858-2015",
      "classification": "Class II",
      "status": "terminated",
      "initiated": "2014-11-20",
      "reason": "The recalled product was distributed with an incorrect serial number printed on the device's label.",
      "applies": "this device",
      "via": "P970003"
    },
    {
      "number": "Z-0210-2012",
      "classification": "",
      "status": "terminated",
      "initiated": "2011-08-12",
      "reason": "The devices are being recalled because the output current delivered to the vagus nerve is less than the design intent and there is a potential charge imbalance at the lead cathode and generator-can during stimulation.",
      "applies": "this device",
      "via": "P970003"
    },
    {
      "number": "Z-0005-2012",
      "classification": "",
      "status": "terminated",
      "initiated": "2011-05-06",
      "reason": "An investigation was initiated based on a report from the field in which an Intensive Follow-up Indicator (IFI) message was unexpectedly received by a medical professional when using Model 250 version 8.0 software to interrogate a patient's Model 103 Generator.",
      "applies": "this device",
      "via": "P970003"
    },
    {
      "number": "Z-0006-2012",
      "classification": "",
      "status": "terminated",
      "initiated": "2011-05-06",
      "reason": "An investigation was initiated based on a report from the field in which an Intensive Follow-up Indicator (IFI) message was unexpectedly received by a medical professional when using Model 250 version 8.0 software to interrogate a patient's Model 103 Generator.",
      "applies": "this device",
      "via": "P970003"
    },
    {
      "number": "Z-1591-2010",
      "classification": "",
      "status": "terminated",
      "initiated": "2010-04-06",
      "reason": "Battery life projection is inaccurate.",
      "applies": "this device",
      "via": "P970003"
    },
    {
      "number": "Z-0994-2009",
      "classification": "",
      "status": "terminated",
      "initiated": "2008-02-07",
      "reason": "Reset/disabling of the VNS Therapy Demipulse Generator and Demipulse Duo Generator due to magnet interference resulting in the loss of stimulation.",
      "applies": "this device",
      "via": "P970003"
    },
    {
      "number": "Z-0995-2009",
      "classification": "",
      "status": "terminated",
      "initiated": "2008-02-07",
      "reason": "Reset/disabling of the VNS Therapy Demipulse Generator and Demipulse Duo Generator due to magnet interference, resulting in the loss of stimulation.",
      "applies": "this device",
      "via": "P970003"
    },
    {
      "number": "Z-0171-2008",
      "classification": "",
      "status": "terminated",
      "initiated": "2007-08-02",
      "reason": "Dissolution/Fractures to the leads of the VNS Therapy System",
      "applies": "this device",
      "via": "P970003"
    },
    {
      "number": "Z-0172-2008",
      "classification": "",
      "status": "terminated",
      "initiated": "2007-08-02",
      "reason": "Dissolution/Fractures to the leads of the VNS Therapy System",
      "applies": "this device",
      "via": "P970003"
    },
    {
      "number": "Z-0173-2008",
      "classification": "",
      "status": "terminated",
      "initiated": "2007-08-02",
      "reason": "Dissolution/Fractures to the leads of the VNS Therapy System",
      "applies": "this device",
      "via": "P970003"
    },
    {
      "number": "Z-0341-2007",
      "classification": "",
      "status": "terminated",
      "initiated": "2006-10-17",
      "reason": "During programming, pulse generator may be inadvertently set to 8.0 mA  output, regardless of the mA range selected by the clinician",
      "applies": "this device",
      "via": "P970003"
    },
    {
      "number": "Z-0342-2007",
      "classification": "",
      "status": "terminated",
      "initiated": "2006-10-17",
      "reason": "During programming, pulse generator may be inadvertently set to 8.0 mA  output, regardless of the mA range selected by the clinician",
      "applies": "this device",
      "via": "P970003"
    }
  ],
  "eu_market_availability": null,
  "properties": [
    {
      "group": "Identity",
      "fields": [
        {
          "label": "Trade name",
          "values": [
            {
              "value": "VNS Therapy®",
              "registries": [
                "EUDAMED"
              ]
            },
            {
              "value": "VNS Therapy® SenTiva Duo™ Model 1000-D Generator",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "Generic name",
          "values": [
            {
              "value": "Antiseizure/psychiatric-therapy vagus nerve implantable electrical stimulation system",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "Device name",
          "values": [
            {
              "value": "VNS Therapy® SenTiva Duo™ Generator",
              "registries": [
                "EUDAMED"
              ]
            }
          ]
        },
        {
          "label": "Model / reference",
          "values": [
            {
              "value": "Model 1000-D",
              "registries": [
                "EUDAMED"
              ]
            },
            {
              "value": "1000-D",
              "registries": [
                "EUDAMED",
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        }
      ]
    },
    {
      "group": "Identifiers",
      "fields": [
        {
          "label": "Primary DI / UDI-DI",
          "values": [
            {
              "value": "05425025750528",
              "registries": [
                "EUDAMED",
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "Basic UDI-DI",
          "values": [
            {
              "value": "B-05425025750528",
              "registries": [
                "EUDAMED"
              ]
            }
          ]
        },
        {
          "label": "PMA number",
          "values": [
            {
              "value": "P970003",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "FDA product code",
          "values": [
            {
              "value": "MUZ",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            },
            {
              "value": "LYJ",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "EMDN code",
          "values": [
            {
              "value": "J020301",
              "registries": [
                "EUDAMED"
              ]
            },
            {
              "value": "J020399",
              "registries": [
                "EUDAMED"
              ]
            }
          ]
        },
        {
          "label": "GMDN term",
          "values": [
            {
              "value": "Antiseizure/psychiatric-therapy vagus nerve implantable electrical stimulation system",
              "code": "34210",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        }
      ]
    },
    {
      "group": "Classification",
      "fields": [
        {
          "label": "Device class",
          "values": [
            {
              "value": "Class AIMDD",
              "registries": [
                "EUDAMED"
              ]
            },
            {
              "value": "Class III",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "EU registration type",
          "values": [
            {
              "value": "Legacy device (under a directive)",
              "registries": [
                "EUDAMED"
              ]
            }
          ]
        },
        {
          "label": "EU market of registration",
          "values": [
            {
              "value": "Italy",
              "registries": [
                "EUDAMED"
              ]
            }
          ]
        },
        {
          "label": "GUDID distribution status",
          "values": [
            {
              "value": "In Commercial Distribution",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        }
      ]
    },
    {
      "group": "Clinical & Handling",
      "fields": [
        {
          "label": "MRI safety",
          "values": [
            {
              "value": "MR Conditional",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        }
      ]
    },
    {
      "group": "Properties",
      "fields": [
        {
          "label": "Supplied sterile",
          "values": [
            {
              "value": "Yes",
              "registries": [
                "EUDAMED",
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "Single use",
          "values": [
            {
              "value": "Yes",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "Implantable",
          "values": [
            {
              "value": "Yes",
              "registries": [
                "EUDAMED",
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "Prescription use",
          "values": [
            {
              "value": "Yes",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "Serial number",
          "values": [
            {
              "value": "Yes",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "Expiration date",
          "values": [
            {
              "value": "Yes",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "Manufacturing date",
          "values": [
            {
              "value": "Yes",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "Sterilise before use",
          "values": [
            {
              "value": "No",
              "registries": [
                "EUDAMED",
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "Reusable",
          "values": [
            {
              "value": "No",
              "registries": [
                "EUDAMED"
              ]
            }
          ]
        },
        {
          "label": "Contains latex",
          "values": [
            {
              "value": "No",
              "registries": [
                "EUDAMED",
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "Reprocessed single-use device",
          "values": [
            {
              "value": "No",
              "registries": [
                "EUDAMED"
              ]
            }
          ]
        },
        {
          "label": "Incorporates a substance derived from human blood or plasma",
          "values": [
            {
              "value": "No",
              "registries": [
                "EUDAMED"
              ]
            }
          ]
        },
        {
          "label": "Contains human tissues or cells",
          "values": [
            {
              "value": "No",
              "registries": [
                "EUDAMED"
              ]
            }
          ]
        },
        {
          "label": "Contains animal tissues or cells",
          "values": [
            {
              "value": "No",
              "registries": [
                "EUDAMED"
              ]
            }
          ]
        },
        {
          "label": "Administering medicine",
          "values": [
            {
              "value": "No",
              "registries": [
                "EUDAMED"
              ]
            }
          ]
        },
        {
          "label": "Measuring function",
          "values": [
            {
              "value": "No",
              "registries": [
                "EUDAMED"
              ]
            }
          ]
        },
        {
          "label": "Incorporates a medicinal substance",
          "values": [
            {
              "value": "No",
              "registries": [
                "EUDAMED"
              ]
            }
          ]
        },
        {
          "label": "Over-the-counter",
          "values": [
            {
              "value": "No",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "Convenience kit",
          "values": [
            {
              "value": "No",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "Combination product",
          "values": [
            {
              "value": "No",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "HCT/P",
          "values": [
            {
              "value": "No",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "Premarket exempt",
          "values": [
            {
              "value": "No",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "Lot / batch number",
          "values": [
            {
              "value": "No",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        }
      ]
    }
  ]
}