VNS Therapy® Programmer, Model 3000 v1.6.2
by LivaNova USA, Inc. EUDAMEDGUDID
Active
Ref 3000v1.6.2, 3000v1.6.1Model Model 3000v1.6.2, Model 3000v1.6.1
Regulatory status
Recall ongoing. 6 ongoing FDA recalls. See recalls
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| European Union | CE/AIMDD EUDAMED | NumberB-05425025750535 | Date12 May 2026 | StatusActive |
| United States | GUDID listing GUDID | Number05425025750535 | Date1 Mar 2021 | StatusIn distribution |
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not known
- Not covered
- European Union
- United States
Certificates
EUDAMED lists no certificate for this device's Basic UDI-DI.
Recalls
-
- Ongoing Z-1067-2025
- Ongoing Z-1068-2025
Their is a potential that implanted pulse generators may stop delivering therapy due to an internal component issue.
-
- Ongoing Z-1072-2025
Vagus nerve stimulator generators may stop delivering stimulation due to an internal, mechanically activated component that may become stuck in a closed position, which may result in patients returning to baseline seizure frequency or depressive symptoms and in the longer term, patients may be required to undergo generator replacement surgery to replace their device.
-
- Ongoing Z-1563-2020
False positive warning may occur after: 1) VNS Generator interrogated at 0mA normal output current, 2) Generator programmed to non-0mA output current, 3) In-session re-interrogation performed. Users instructed to lower output current and widen pulse width. Only system diagnostic testing evaluates output current. Users may conclude device malfunction, could lead to medical/surgical intervention.
-
- Ongoing Z-0663-2020
This recall is being initiated due to reports that that the therapy programming tablet with software version 1.5 errantly performs a normal mode diagnostic test instead of the selected system diagnostic test on Model 102 and Model 102R devices, if the output current is greater than 0.5mA. This can result in false high impedance values during patient follow-up.
-
- Ongoing Z-0346-2020
Lead impedance values reported by the affected VNS generator will be higher compared to those reported by previous models. This is due to a change in the timing of when affected VNS generator takes the lead impedance measurement during diagnostic testing. As a result, normal impedance ranges for the affected VNS generator have shifted relative to the existing thresholds of 600 - 5300 Ohms defined in labeling and as present in the programming software.
-
- Terminated Z-0532-2022
A manufacturing issue may cause the silicone tubing wall thickness of vagus nerve stimulation leads to be less than the design specification. If the tubing becomes completely breached, the patient may experience discomfort or pain due to delivery of stimulation to an unintended location. If the inner lead conductor is exposed to body fluids the lead may break causing loss of therapy.
-
- Terminated Z-0462-2021
Due to an extra digit being inadvertently added to the serial number of a subset of implantable pulse generators, they become not compatible with therapy programmers with software version 11.0.4.
-
- Terminated Z-0561-2020
Sentiva generators may reset and become disabled within 60 days of enabling. Physicians can re-enable but devices will continue to be susceptible to resets. If a device experiences this issue, this may lead to additional surgery (premature replacement of generator), or patients may return to baseline seizure frequency or depressive symptoms due to cessation of intended stimulation.
-
- Terminated Z-1118-2020
Firm identified a subset of its generators that were sterilized one additional sterilization cycle, which does not meet the firm's quality specifications.
-
- Terminated Z-0869-2020
- Terminated Z-0868-2020
This recall is an expansion of Z-3019-2017 and Z-3020-2017, which was initiated to fix the devices premature battery depletion, caused by electrical leakage on the circuit board assemblies of the Models 105 and 106 generators. *Note this recall occurred in November 2018.
-
- Terminated Z-2572-2018
Unintended warning message displayed on generators programmed with a Model 3000 v.1.0.2.2 programmer.
-
- Terminated Z-0555-2018
Two Model 3000 Programmers were distributed in error by prior to FDA approval of version 1.0.2.2 software.
-
- Terminated Z-2255-2018
Certain Model 3000 programming events can result in miscalculation of parameters stored in the Models 103, 104, 105, and 106 generators. During these programming events, the miscalculations can lead to: " Delivery of more stimulation than intended, resulting in painful stimulation or other common side effects (Model 106 only); " No stimulation in the case of device disablement (Burst Watchdog Timeout), resulting in no therapy to the patient (Model 106 only); " Delivery of less stimulation than intended, resulting in therapeutic settings not being achieved within device specification (Models 103, 104, 105, or 106); and/or " Delays or absence of the 75% and 50% battery life indicators displayed by the programming software (Models 103, 104, 105, or 106).
-
- Terminated Z-0659-2016
Recall being initiated in response to three reports of "Burst Watchdog Timeout" events occurring with the Model 106 AspireSR Generator, resulting in a device reset condition where stimulation output is disabled.
-
- Terminated Z-0645-2016
Certain Model 106 Pulse Generators demonstrate delays in sensing during use of the 'Verify Heartbeat Detection' feature and exhibit the potential for decreased battery longevity.
-
- Terminated Z-0280-2016
Certain Model 106 Pulse Generators demonstrate delays in sensing during use of the 'Verify Heartbeat Detection' feature and exhibit potential for decreased battery longevity.
-
- Terminated Z-0858-2015
The recalled product was distributed with an incorrect serial number printed on the device's label.
-
- Terminated Z-0210-2012
The devices are being recalled because the output current delivered to the vagus nerve is less than the design intent and there is a potential charge imbalance at the lead cathode and generator-can during stimulation.
-
- Terminated Z-0006-2012
- Terminated Z-0005-2012
An investigation was initiated based on a report from the field in which an Intensive Follow-up Indicator (IFI) message was unexpectedly received by a medical professional when using Model 250 version 8.0 software to interrogate a patient's Model 103 Generator.
-
- Terminated Z-1591-2010
Battery life projection is inaccurate.
-
- Terminated Z-0994-2009
Reset/disabling of the VNS Therapy Demipulse Generator and Demipulse Duo Generator due to magnet interference resulting in the loss of stimulation.
-
- Terminated Z-0995-2009
Reset/disabling of the VNS Therapy Demipulse Generator and Demipulse Duo Generator due to magnet interference, resulting in the loss of stimulation.
-
- Terminated Z-0173-2008
- Terminated Z-0172-2008
- Terminated Z-0171-2008
Dissolution/Fractures to the leads of the VNS Therapy System
-
- Terminated Z-0341-2007
- Terminated Z-0342-2007
During programming, pulse generator may be inadvertently set to 8.0 mA output, regardless of the mA range selected by the clinician
Specification
- Trade Name
- VNS Therapy® EUDAMEDVNS Therapy® Programmer, Model 3000 v1.6.2 GUDIDVNS Therapy® Programmer, Model 3000 v1.6 GUDID
- Generic Name
- Vagus nerve electrical stimulation system programmer
- Device Name
- VNS Therapy® Programmer
- Model / Reference
- Model 3000v1.6.2 EUDAMED3000v1.6.2 EUDAMEDModel 3000v1.6.1 EUDAMED3000v1.6.1 EUDAMED3000 GUDID
- Primary DI / UDI-DI
- 05425025750535, 05425025750443
- Basic UDI-DI
- B-05425025750535, B-05425025750443
- PMA Number
- FDA Product Code
- LYJ, MUZ
- EMDN Code
- J020701 Neurostimulators programmers
- GMDN Term
- Vagus nerve electrical stimulation system programmer
- Device Class
- Class AIMDD EUDAMEDClass III GUDID
- Special Device Type
- -1200
- Market of Registration
- Italy
- Commercial Distribution
- In Commercial Distribution
- MRI Safety
- MR Unsafe
Truth per registry
| Field | EUDAMED | GUDID |
|---|---|---|
| Model / Reference | Model 3000v1.6.2 3000v1.6.2 Model 3000v1.6.1 3000v1.6.1 | 3000 |
| Device Class | Class AIMDD | Class III |
Other devices from LivaNova USA, Inc.
Properties
Other properties (19)
- Sterile: No
- Single Use: No
- Implantable: No
- Reusable: No
- Contains Latex: No
- Reprocessed: No
- Human-derived: No
- Human Tissues: No
- Animal Tissues: No
- Administering Medicine: No
- Measuring Function: No
- Incorporates Medicinal Product: No
- Over-the-Counter: No
- Convenience Kit: No
- Combination Product: No
- HCT/P: No
- Premarket Exempt: No
- Lot / Batch Number: No
- Expiration Date: No