Neurocybernetic Prosthesis (NCP) Programming Wand
PMA P970003
Active
Valid until
No expiry
FDA recalls
6 ongoingDocuments
Trust
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA PMAAPPR | NumberP970003 | Date16 Jul 1997 | StatusActive |
Official records
- FDA PMA database P970003
From the FDA record
FDA PMA P970003
- Received by the FDA
- 27 Jan 1997
- FDA decision
- 16 Jul 1997
- FDA review time
- 170 days from receipt to decision
- Expedited review
- Yes
- Docket number
- 98M-0140
Approval for the NeuroCybernetic Prosthesis (NCP(R)) System which includes the Model 100 NCP Generator, the Model 300 Series NCP Vagus Nerve Stimulation Lead, the Model 400 Tunneling Tool, and NCP System Accessories.
Recalls
-
- Ongoing Z-1067-2025
- Ongoing Z-1068-2025
Their is a potential that implanted pulse generators may stop delivering therapy due to an internal component issue.
-
- Ongoing Z-1072-2025
Vagus nerve stimulator generators may stop delivering stimulation due to an internal, mechanically activated component that may become stuck in a closed position, which may result in patients returning to baseline seizure frequency or depressive symptoms and in the longer term, patients may be required to undergo generator replacement surgery to replace their device.
-
- Ongoing Z-1563-2020
False positive warning may occur after: 1) VNS Generator interrogated at 0mA normal output current, 2) Generator programmed to non-0mA output current, 3) In-session re-interrogation performed. Users instructed to lower output current and widen pulse width. Only system diagnostic testing evaluates output current. Users may conclude device malfunction, could lead to medical/surgical intervention.
-
- Ongoing Z-0346-2020
Lead impedance values reported by the affected VNS generator will be higher compared to those reported by previous models. This is due to a change in the timing of when affected VNS generator takes the lead impedance measurement during diagnostic testing. As a result, normal impedance ranges for the affected VNS generator have shifted relative to the existing thresholds of 600 - 5300 Ohms defined in labeling and as present in the programming software.
-
- Ongoing Z-0663-2020
This recall is being initiated due to reports that that the therapy programming tablet with software version 1.5 errantly performs a normal mode diagnostic test instead of the selected system diagnostic test on Model 102 and Model 102R devices, if the output current is greater than 0.5mA. This can result in false high impedance values during patient follow-up.
-
- Terminated Z-0532-2022
A manufacturing issue may cause the silicone tubing wall thickness of vagus nerve stimulation leads to be less than the design specification. If the tubing becomes completely breached, the patient may experience discomfort or pain due to delivery of stimulation to an unintended location. If the inner lead conductor is exposed to body fluids the lead may break causing loss of therapy.
-
- Terminated Z-0462-2021
Due to an extra digit being inadvertently added to the serial number of a subset of implantable pulse generators, they become not compatible with therapy programmers with software version 11.0.4.
-
- Terminated Z-0561-2020
Sentiva generators may reset and become disabled within 60 days of enabling. Physicians can re-enable but devices will continue to be susceptible to resets. If a device experiences this issue, this may lead to additional surgery (premature replacement of generator), or patients may return to baseline seizure frequency or depressive symptoms due to cessation of intended stimulation.
-
- Terminated Z-1118-2020
Firm identified a subset of its generators that were sterilized one additional sterilization cycle, which does not meet the firm's quality specifications.
-
- Terminated Z-0868-2020
- Terminated Z-0869-2020
This recall is an expansion of Z-3019-2017 and Z-3020-2017, which was initiated to fix the devices premature battery depletion, caused by electrical leakage on the circuit board assemblies of the Models 105 and 106 generators. *Note this recall occurred in November 2018.
-
- Terminated Z-2572-2018
Unintended warning message displayed on generators programmed with a Model 3000 v.1.0.2.2 programmer.
-
- Terminated Z-0555-2018
Two Model 3000 Programmers were distributed in error by prior to FDA approval of version 1.0.2.2 software.
-
- Terminated Z-2255-2018
Certain Model 3000 programming events can result in miscalculation of parameters stored in the Models 103, 104, 105, and 106 generators. During these programming events, the miscalculations can lead to: " Delivery of more stimulation than intended, resulting in painful stimulation or other common side effects (Model 106 only); " No stimulation in the case of device disablement (Burst Watchdog Timeout), resulting in no therapy to the patient (Model 106 only); " Delivery of less stimulation than intended, resulting in therapeutic settings not being achieved within device specification (Models 103, 104, 105, or 106); and/or " Delays or absence of the 75% and 50% battery life indicators displayed by the programming software (Models 103, 104, 105, or 106).
-
- Terminated Z-0659-2016
Recall being initiated in response to three reports of "Burst Watchdog Timeout" events occurring with the Model 106 AspireSR Generator, resulting in a device reset condition where stimulation output is disabled.
-
- Terminated Z-0645-2016
Certain Model 106 Pulse Generators demonstrate delays in sensing during use of the 'Verify Heartbeat Detection' feature and exhibit the potential for decreased battery longevity.
-
- Terminated Z-0280-2016
Certain Model 106 Pulse Generators demonstrate delays in sensing during use of the 'Verify Heartbeat Detection' feature and exhibit potential for decreased battery longevity.
-
- Terminated Z-0858-2015
The recalled product was distributed with an incorrect serial number printed on the device's label.
-
- Terminated Z-0210-2012
The devices are being recalled because the output current delivered to the vagus nerve is less than the design intent and there is a potential charge imbalance at the lead cathode and generator-can during stimulation.
-
- Terminated Z-0005-2012
- Terminated Z-0006-2012
An investigation was initiated based on a report from the field in which an Intensive Follow-up Indicator (IFI) message was unexpectedly received by a medical professional when using Model 250 version 8.0 software to interrogate a patient's Model 103 Generator.
-
- Terminated Z-1591-2010
Battery life projection is inaccurate.
-
- Terminated Z-0994-2009
Reset/disabling of the VNS Therapy Demipulse Generator and Demipulse Duo Generator due to magnet interference resulting in the loss of stimulation.
-
- Terminated Z-0995-2009
Reset/disabling of the VNS Therapy Demipulse Generator and Demipulse Duo Generator due to magnet interference, resulting in the loss of stimulation.
-
- Terminated Z-0171-2008
- Terminated Z-0172-2008
- Terminated Z-0173-2008
Dissolution/Fractures to the leads of the VNS Therapy System
-
- Terminated Z-0341-2007
- Terminated Z-0342-2007
During programming, pulse generator may be inadvertently set to 8.0 mA output, regardless of the mA range selected by the clinician
Timeline
- Recall initiated FDA recalls Z-0868-2020
- Recall initiated FDA recalls Z-0869-2020
- Recall initiated FDA recalls Z-1118-2020
- Recall initiated FDA recalls Z-0561-2020
- Recall initiated FDA recalls Z-1563-2020
- Recall initiated FDA recalls Z-0462-2021
- Recall initiated FDA recalls Z-0532-2022
- Recall initiated FDA recalls Z-1072-2025
- Recall initiated FDA recalls Z-1067-2025
- Recall initiated FDA recalls Z-1068-2025
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA PMA Active
Specification
- Trade name
- Vns Therapy System
- Generic name
- Stimulator, autonomic nerve, implanted for epilepsy
- PMA number
- FDA product code
- LYJ
- Device class
- Class III
- Advisory committee
- Neurology
- FDA decision
- APPR Approved
Overview
AI-generated summary, 18 Sep 2026
The Model 201 NCP Programming Wand is an implantable programmer designed for use with the NeuroCybernetic Prosthesis (NCP) System. It functions as an autonomic nerve stimulator, specifically for vagus nerve stimulation, and is intended for programming and communication with implanted devices such as the Model 100 NCP Generator.
This device is powered by a single 9-volt battery and transmits signals to and from a computer via a cable connected to a standard DB9 plug. It is supplied as part of the NCP System and is regulated under FDA Premarket Approval (PMA) P970003. The wand is intended for clinical use in settings where vagus nerve stimulation therapy is administered, adhering to sterile procedures where applicable.
Devices cleared under P970003
Devices that FDA's GUDID lists as cleared or approved under this submission number. One submission usually covers a whole product family, so the sizes and models are listed here once.