  "format": "devicerts.device.v1",
  "cite_as": "Vv Fluoro 3D, Brainlab AG. deviCERTS medical device catalogue. https://devicerts.com/devices/vv-fluoro-3d-haw (data as of 2026-11-17).",
  "source": {
    "canonical_url": "https://devicerts.com/devices/vv-fluoro-3d-haw",
    "json_url": "https://devicerts.com/devices/vv-fluoro-3d-haw.json",
    "markdown_url": "https://devicerts.com/devices/vv-fluoro-3d-haw.md",
    "publisher": "deviCERTS (devicerts.com)",
    "data_sources": [
      {
        "registry": "FDA 510(k)",
        "url": "https://open.fda.gov/apis/device/510k/"
      },
      {
        "registry": "FDA recalls",
        "url": "https://open.fda.gov/apis/device/recall/"
      }
    ],
    "registry_records_last_synced": "2026-10-22",
    "status_as_of": "2026-11-17",
    "license": "CC BY 4.0",
    "license_url": "https://creativecommons.org/licenses/by/4.0/",
    "license_scope": "The deviCERTS layout and text are licensed under CC BY 4.0. Registry data remain under their source terms: U.S. FDA data (openFDA, AccessGUDID) are U.S. public domain; EUDAMED data are subject to the European Commission reuse terms.",
    "attribution": "DevicErts, https://devicerts.com/devices/vv-fluoro-3d-haw",
    "disclaimer": "Compiled by deviCERTS (devicerts.com) from public medical device registries (EUDAMED, European Commission; U.S. FDA via openFDA; AccessGUDID, U.S. National Library of Medicine). deviCERTS is not an official register and gives no regulatory advice: the record in the registry itself prevails."
  },
  "device": {
    "name": "Vv Fluoro 3D",
    "registry_name": "VV FLUORO 3D",
    "url": "https://devicerts.com/devices/vv-fluoro-3d-haw",
    "manufacturer": {
      "name": "Brainlab AG",
      "url": "https://devicerts.com/manufacturers/brainlab-ag",
      "country": null,
      "country_basis": null,
      "fda_registration_country": null
    },
    "status": {
      "state": "active",
      "label": "Active",
      "detail": "United States"
    },
    "classification": [
      {
        "class": "Class II",
        "system": "US FDA"
      }
    ],
    "identifiers": [
      {
        "label": "510(k) Number",
        "values": [
          {
            "value": "K070106",
            "registries": [
              "FDA 510(k)"
            ]
          }
        ]
      }
    ],
    "product_codes": [
      "HAW"
    ],
    "emdn": [],
    "gmdn": [],
    "description": null,
    "noindex": false
  },
  "authorizations": [
    {
      "registry": "510(k)",
      "region": "United States",
      "type": "FDA 510(k)",
      "number": "K070106",
      "number_label": null,
      "device_udi_di": null,
      "date": "2007-04-23",
      "date_kind": "decision",
      "decision_code": "SESE",
      "expires": null,
      "withdrawn": null,
      "status": "Active",
      "status_note": null,
      "derived_from_gudid": false
    }
  ],
  "no_longer_listed_by": [],
  "certificates": [],
  "recalls": [
    {
      "number": "Z-1082-2019",
      "classification": "Class I",
      "status": "terminated",
      "initiated": "2019-02-27",
      "reason": "In certain occurrences, the affected navigation software application might unexpectedly display a navigated instrument in an axial, coronal/sagittal (ACS) view representation with fixed planes in the image reconstruction (the  not-updated ACS view ), instead of displaying the desirable view representation  Inline View , which is commonly used for navigating invasive instruments at the spine. This could lead the surgeon to be unable to determine the position of the navigated instrument. This might occur after a  crash restore  or after changing between different navigation workflows during the same patient treatment.",
      "applies": "this device",
      "via": "K070106"
    },
    {
      "number": "Z-0320-2018",
      "classification": "Class II",
      "status": "terminated",
      "initiated": "2017-11-16",
      "reason": "Due to a human error at a supplier, an unknown number of devices within one specific production lot of Brainlab DrapeLink Adapters (Left/Right) were not\nmanufactured according to defined specifications.",
      "applies": "this device",
      "via": "K070106"
    },
    {
      "number": "Z-0282-2016",
      "classification": "Class II",
      "status": "terminated",
      "initiated": "2015-05-08",
      "reason": "Instances of data sets not being accurately registered to the patient anatomy were observed.",
      "applies": "this device",
      "via": "K070106"
    },
    {
      "number": "Z-2213-2014",
      "classification": "Class II",
      "status": "terminated",
      "initiated": "2014-05-21",
      "reason": "Brainlab Navigation Software Spine & Trauma 3D 2.0/2.1 offers automatic registration of intraoperatively acquired CT image data sets. To enable automatic registration, the software requires the gantry position of the scanner. \n\nThe gantry position can either be entered manually or submitted automatically from the CT scanner. If automatic gantry communication is available, the so-called &quot",
      "applies": "this device",
      "via": "K070106"
    },
    {
      "number": "Z-1469-2014",
      "classification": "Class II",
      "status": "terminated",
      "initiated": "2013-11-18",
      "reason": "The \"fluoro match registration\" function of the Navigation Software Spine & Trauma 3D 2.0 allows the user to intraoperatively match e.g., CT data sets to the current patient anatomy visible on fluoroscopic images, used by the navigation software during spinal surgeries to display the position and orientation of instruments. For this registration function in combination with a digitally integrated",
      "applies": "this device",
      "via": "K070106"
    }
  ],
  "eu_market_availability": null,
  "properties": [
    {
      "group": "Identifiers",
      "fields": [
        {
          "label": "510(k) Number",
          "values": [
            {
              "value": "K070106",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        },
        {
          "label": "FDA product code",
          "values": [
            {
              "value": "HAW",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        }
      ]
    },
    {
      "group": "Classification",
      "fields": [
        {
          "label": "Device class",
          "values": [
            {
              "value": "Class II",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        },
        {
          "label": "Regulation",
          "values": [
            {
              "value": "21 CFR Part 882.4560",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        },
        {
          "label": "Advisory committee",
          "values": [
            {
              "value": "Neurology",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        },
        {
          "label": "FDA decision",
          "values": [
            {
              "value": "SESE",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        }
      ]
    }
  ]
}