# Walkaway Tm 40

> Active: United States. Medical device record on deviCERTS, compiled from public registries.

- Canonical page: <https://devicerts.com/devices/walkaway-tm-40-lrg>
- Manufacturer: [Baxter Healthcare Corp](https://devicerts.com/manufacturers/baxter-healthcare-corp) (United States, FDA 510(k) applicant)
- Name in the registry: WALKAWAY TM 40
- Device class: Class II (US FDA)
- FDA product code: LRG

## Registrations and authorizations

- **510(k)** · FDA 510(k) · United States · K911400 · decision 1991-11-19 · decision SESE · status: Active

## Recalls

- Z-1088-2022 · Class II · ongoing · initiated 2019-10-21: WalkAway Instruments may have been assembled with the incorrect AC power entry connector inlet with is rated at 1 ampere current rating rather than the correct current rating of 10 ampere
- Z-1083-2022 · Class II · ongoing · initiated 2019-10-21: WalkAway Instruments may have been assembled with the incorrect AC power entry connector inlet with is rated at 1 ampere current rating rather than the correct current rating of 10 ampere
- Z-1086-2022 · Class II · ongoing · initiated 2019-10-21: WalkAway Instruments may have been assembled with the incorrect AC power entry connector inlet with is rated at 1 ampere current rating rather than the correct current rating of 10 ampere
- Z-1084-2022 · Class II · ongoing · initiated 2019-10-21: WalkAway Instruments may have been assembled with the incorrect AC power entry connector inlet with is rated at 1 ampere current rating rather than the correct current rating of 10 ampere
- Z-1085-2022 · Class II · ongoing · initiated 2019-10-21: WalkAway Instruments may have been assembled with the incorrect AC power entry connector inlet with is rated at 1 ampere current rating rather than the correct current rating of 10 ampere
- Z-1087-2022 · Class II · ongoing · initiated 2019-10-21: WalkAway Instruments may have been assembled with the incorrect AC power entry connector inlet with is rated at 1 ampere current rating rather than the correct current rating of 10 ampere
- Z-1373-2021 · Class II · terminated · initiated 2019-09-04: Instruments were manufactured with a shortened 30mL reagent draw straw which may result in reagents not dispensing and may lead to organism misidentification and/or Quality Control (QC) failure.
- Z-1372-2021 · Class II · terminated · initiated 2019-09-04: Instruments were manufactured with a shortened 30mL reagent draw straw which may result in reagents not dispensing and may lead to organism misidentification and/or Quality Control (QC) failure.
- Z-1370-2021 · Class II · terminated · initiated 2019-09-04: Instruments were manufactured with a shortened 30mL reagent draw straw which may result in reagents not dispensing and may lead to organism misidentification and/or Quality Control (QC) failure.
- Z-1371-2021 · Class II · terminated · initiated 2019-09-04: Instruments were manufactured with a shortened 30mL reagent draw straw which may result in reagents not dispensing and may lead to organism misidentification and/or Quality Control (QC) failure.
- Z-1990-2014 · Class II · terminated · initiated 2014-06-02: Springs contained in the access door hinge assembly on the Walk Away Plus instruments are becoming detached causing premature failure of the hinge assembly and a potential safety issue for the customers.

## Identifiers

- 510(k) Number: K911400 [FDA 510(k)]
- FDA product code: LRG [FDA 510(k)]

## Classification

- Device class: Class II [FDA 510(k)]
- Regulation: 21 CFR Part 866.1640 [FDA 510(k)]
- Advisory committee: Microbiology [FDA 510(k)]
- FDA decision: SESE [FDA 510(k)]

## Source and citation

- Cite as: Walkaway Tm 40, Baxter Healthcare Corp. deviCERTS medical device catalogue. https://devicerts.com/devices/walkaway-tm-40-lrg (data as of 2026-11-18).
- Data sources: [FDA 510(k)](https://open.fda.gov/apis/device/510k/), [FDA recalls](https://open.fda.gov/apis/device/recall/)
- Registry records last synced: 2026-10-11
- Statuses computed as of: 2026-11-18
- JSON: <https://devicerts.com/devices/walkaway-tm-40-lrg.json>
- Licence: [CC BY 4.0](https://creativecommons.org/licenses/by/4.0/). The deviCERTS layout and text are licensed under CC BY 4.0. Registry data remain under their source terms: U.S. FDA data (openFDA, AccessGUDID) are U.S. public domain; EUDAMED data are subject to the European Commission reuse terms.
- Attribution: DevicErts, https://devicerts.com/devices/walkaway-tm-40-lrg
- Compiled by deviCERTS (devicerts.com) from public medical device registries (EUDAMED, European Commission; U.S. FDA via openFDA; AccessGUDID, U.S. National Library of Medicine). deviCERTS is not an official register and gives no regulatory advice: the record in the registry itself prevails.