Wang
by Conmed Corporation United States EUDAMEDGUDID
Ref MW-522
Product code EOQEMDN A01020502
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| European Union | CE/MDD EUDAMED | NumberB-20653405052675 | Date3 Apr 2026 | StatusActive |
| United States | GUDID listing GUDID | Number20653405052675 | Date24 Sep 2016 | StatusIn distribution |
Coverage
- Expiring (legacy directive)
- Listed
- Not covered
- European Union
- United States
Recalls
-
- Terminated Z-0504-2013
- Terminated Z-0502-2013
- Terminated Z-0505-2013
- Terminated Z-0503-2013
ConMed received complaints from a user facility reporting that the WANG Transbronchial Aspiration Needles would not insert properly through the instrument channel of a 2.0 mm bronchoscope. If the device will not fit through the instrument channel of a 2.0 mm bronchoscope, it must be replaced with an alternate device and could result in minor inconvenience to the user.
Specification
- Generic Name
- Body aspiration needle, single-use
- Device Name
- Non-orthopedic, Non-active, Needle, Aspiration
- Model / Ref / Catalog
- MW-522
- Primary DI / UDI-DI
- 20653405052675
- Basic UDI-DI
- B-20653405052675
- 510(k) Number
- FDA Product Code
- EOQ
- EMDN Code
- A01020502 Laparoscopic and thoracoscopic biopsy needles and kits
- GMDN Term
- Body aspiration needle, single-use
- Device Class
- Class IIa EUDAMEDClass II GUDID
- Regulation Number
- 874.4680
- Market of Registration
- Belgium
- Commercial Distribution
- In Commercial Distribution
- Sterilization Method
- Ethylene Oxide
- Size
- Outer Diameter — 1.9 MillimeterLength — 130 Centimeter
Truth per registry
| Field | EUDAMED | GUDID |
|---|---|---|
| Device Class | Class IIa | Class II |
Manufacturer details
- EUDAMED SRN
- US-MF-000012663
- Authorised representative
- MDSS GmbH DE-AR-000005430
Other devices from Conmed Corporation
Properties & raw data
Show properties, registry sources and raw records
Other properties (19)
- Implantable: No
- Reusable: No
- Contains Latex: No
- Reprocessed: No
- CMR Substance: No
- Human-derived: No
- Human Tissues: No
- Animal Tissues: No
- Endocrine Disruptor: No
- Administering Medicine: No
- Measuring Function: No
- Incorporates Medicinal Product: No
- Over-the-Counter: No
- Convenience Kit: No
- Combination Product: No
- HCT/P: No
- Premarket Exempt: No
- Serial Number: No
- Manufacturing Date: No
Sources (2)
- EUDAMED 98cdf0ff-cf0e-4742-8853-2ccee025c808
- FDA UDI 683cfcce-fa6f-40e1-bfe4-085560d56a6d