Skip to content

Weck Hem-O-Lok Automatic Ligating Clip Applier

by Weck

510(k) K021808

Active United States

510(k)

Valid until

No expiry

510(k) clearance is not time-limited

FDA recalls

1 ongoing

This device is named in the recall

Documents

Trust

Unverified

Registry data, not yet checked by deviCERTS. Source FDA 510(k), synced 22 Oct 2026. What this means

Regulatory status

as of registry sync 22 Oct 2026

RegionAuthorizationNumberDateStatus
United StatesFDA 510(k)SESE Substantially Equivalent · Traditional NumberK021808 Date14 Aug 2002ClearedStatusActive

Recalls

7 recalls · 1 ongoing · FDA

    • Ongoing Z-0497-2020 FDA Class II initiated 4 Oct 2019

    Certain clip appliers have an increase in misloading and/or jamming related complaints that may be due to a discrepancy in the specification a component. Misloading and/or jamming may lead to clips falling out of the applier, clip breakage, or the jaws becoming locked in a partially closed position (jammed).

  • 5 recalls, same reason

    • Terminated Z-0160-2009 FDA Unclassified (pre-2012) initiated 22 Aug 2008
    • Terminated Z-0159-2009 FDA Unclassified (pre-2012) initiated 22 Aug 2008
    • Terminated Z-0158-2009 FDA Unclassified (pre-2012) initiated 22 Aug 2008
    • Terminated Z-0157-2009 FDA Unclassified (pre-2012) initiated 22 Aug 2008
    • Terminated Z-0156-2009 FDA Unclassified (pre-2012) initiated 22 Aug 2008

    A hole in the sterile unit blister pack was detected that would compromise sterility.

    • Terminated Z-1099-06 FDA Unclassified (pre-2012) initiated 12 Apr 2006

    The Hem-o-lok ligating clips may become dislodged following ligation of the renal artery after laparoscopic donor nephrectomy. Hem-o-lok ligating clips are now contraindicated for use in ligating the renal artery during laparascopic nephrectomies in living donor patients.

Coverage

1 jurisdiction

United States of America
  • Active
  • Expiring
  • Withdrawn
  • Listed
  • Not covered

United States — FDA 510(k) Active

Specification

510(k)

510(k) Number
FDA Product Code
FZP
Device Class
Class II
Regulation
21 CFR Part 878.4300
Advisory Committee
General, Plastic Surgery
FDA Decision