WeRay P70A
WeRay P70A is a hand-held intraoral dental x-ray system (FDA Class II; FDA product code EHD, Unit, X-Ray, Extraoral With Timer) made by Picopack Co., Ltd., cleared under FDA 510(k) K241963 on 30 Jul 2024 and listed in FDA GUDID since 24 Oct 2024 (UDI-DI 08809840720046), active as of 23 Oct 2026.
Active
510(k)
Valid until
No expiry
FDA recalls
None found
Documents
Trust
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | GUDID listing | Number08809840720046 | Date24 Oct 2024 | StatusListed |
| United States | FDA 510(k)SESE | NumberK241963 | Date30 Jul 2024 | StatusActive |
Official records
- AccessGUDID record 08809840720046
- FDA 510(k) database K241963
From the FDA record
- Device description
- WERAY is a battery-operated, portable dental X-ray source designed for handheld operation. It is intended to produce diagnostic-quality X-ray images. The WERAY is designed for use in a dental office. It can also be used in other similar environments (orthodontic office, general practitioner’s office, hospital ward, etc.) where appropriate safeguards are implemented. The device uses a rechargeable battery to allow the WERAY to be used. Still, transportation or using other X-ray devices might be prohibitive due to the other devices’ size and/or lack of mobility. The WERAY is an X-ray device with an AC/DC adaptor. The handheld device features a main body (tube head), a Cone for the sensor, Acryl shielding, a cradle, a Hand switch for X-ray exposure, and an AC/DC adaptor. The power is supplied by a rechargeable Lithium-ion polymer battery pack built into a main body. This facilitates the portability of the device. A beam-limiting cone is incorporated within the device. The acryl shielding provides sufficient radiation protection to allow the clinician to remain in the operatory with the patient. A hand switch is available to expose X-rays in a location away from the device. WERAY uses a fixed tube voltage of 70kV and a fixed tube current of 2.0 mA to make the system as simple as possible for the operator. The only operator-adjustable parameter is the exposure time. This adjustment can be quickly accomplished through the user-friendly control panel. Control buttons, a display panel, and an exposure button provide the primary operator interface. Exposure settings can be selected and displayed. The tube voltage (70 kV) and tube current (2.0 mA) are fixed, with the exposure time varying based on patient type, detector type, and anatomical features. Exposures can be completed using the exposure button. The WERAY should be used with an X-ray detector and a digital intraoral sensor. But the X-ray detector is not included in the device package.
- MRI safety
- Labeling does not contain MRI Safety Information
- Natural rubber latex
- No latex statement on the label
- Sterility
- Supplied non-sterile
Across registries
- FDA 510(k) K241963 decided 30 Jul 2024 Active
- FDA GUDID UDI-DI 08809840720046 listed 24 Oct 2024 Listed
Timeline
- Cleared FDA 510(k) K241963
- Listed FDA GUDID 08809840720046
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- Trade name
- WeRay
- Generic name
- Hand-held intraoral dental x-ray system, digital
- Model / reference
- P70A
- Primary DI / UDI-DI
- 08809840720046
- 510(k) Number
- FDA product code
- EHD
- GMDN term
- Hand-held intraoral dental x-ray system, digital
- Device class
- Class II
- Regulation number
- 872.1800
- GUDID distribution status
- In Commercial Distribution
- Size
- Depth — 106 MillimeterHeight — 269 MillimeterWidth — 302 Millimeter